Clinical Study

Randomized Comparison of Terumo® Coated Slender™ versus Terumo® Noncoated Traditional Sheath during Radial Angiography or Percutaneous Coronary Intervention

1Department of Cardiology, Aarhus University Hospital, DK-8200 Aarhus N, Denmark
2Viborg Regional Hospital, Heibergs Alle 4, 8800 Viborg, Denmark

Correspondence should be addressed to Christian Juhl Terkelsen; kd.mr.ybjeks@neslekret.naitsirhc

Received 26 November 2018; Accepted 13 February 2019; Published 4 March 2019

Academic Editor: Matteo Tebaldi

Copyright © 2019 Birthe Sindberg et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Background. The transradial approach is generally associated with few complications. However, periprocedural pain is still a common issue, potentially related to sheath insertion and/or arterial spasm, and may result in conversion to femoral access. Radial artery occlusion (RAO) following the procedure is also a potential risk. We evaluate whether the design of the sheath has any impact on these variables. Methods. A total of 1,000 patients scheduled for radial CAG or PCI were randomized (1:1) to the use of a Slender or a Standard sheath during the procedure. Randomization was stratified according to chosen sheath size (5, 6, 7 French) and gender. A radial band was used to obtain hemostasis after the procedure, employing a rapid deflation technique. A reverse Barbeau test was performed to evaluate radial artery patency after removal of the radial band, and level of pain was assessed using a numeric rating scale (NRS). Results. Use of the Slender sheath was associated with less pain during sheath insertion (median NRS 1 versus 2, p=0.02), whereas no difference was observed in pain during the procedure, radial procedural success rates, use of analgesics and sedatives during the procedure, and radial artery patency following the procedure. Rate of RAO was 1.5% with no difference between groups. Conclusion. The use of the hydrophilic coated Slender sheath during radial CAG or PCI was associated with less pain during sheath insertion, whereas no difference in other endpoints was observed. A rapid deflation technique was associated with RAO of only 1.5%.