Xinyang Ge, Youjun Liu, Zhaofang Yin, Shengxian Tu, Yuqi Fan, Yuri Vassilevski, Sergey Simakov, Fuyou Liang, "Comparison of Instantaneous Wave-Free Ratio (iFR) and Fractional Flow Reserve (FFR) with respect to Their Sensitivities to Cardiovascular Factors: A Computational Model-Based Study", Journal of Interventional Cardiology, vol. 2020, Article ID 4094121, 12 pages, 2020. https://doi.org/10.1155/2020/4094121

Comparison of Instantaneous Wave-Free Ratio (iFR) and Fractional Flow Reserve (FFR) with respect to Their Sensitivities to Cardiovascular Factors: A Computational Model-Based Study

Academic Editor: Piotr Musiałek
Received08 Nov 2019
Revised18 Jan 2020
Accepted27 Feb 2020
Published12 May 2020

Abstract

While coronary revascularization strategies guided by instantaneous wave-free ratio (iFR) are, in general, noninferior to those guided by fractional flow reserve (FFR) with respect to the rate of major adverse cardiac events at one-year follow-up in patients with stable angina or an acute coronary syndrome, the overall accuracy of diagnosis with iFR in large patient cohorts is about 80% compared with the diagnosis with FFR. So far, it remains incompletely understood what factors contribute to the discordant diagnosis between iFR and FFR. In this study, a computational method was used to systemically investigate the respective effects of various cardiovascular factors on FFR and iFR. The results showed that deterioration in aortic valve disease (e.g., regurgitation or stenosis) led to a marked decrease in iFR and a mild increase in FFR given fixed severity of coronary artery stenosis and that increasing coronary microvascular resistance caused a considerable increase in both iFR and FFR, but the degree of increase in iFR was lower than that in FFR. These findings suggest that there is a high probability of discordant diagnosis between iFR and FFR in patients with severe aortic valve disease or coronary microcirculation dysfunction.

1. Introduction

Fractional flow reserve (FFR), defined as the ratio between mean poststenosis coronary arterial and aortic blood pressures under a vasodilator-induced hyperemic condition [1], has been used as a gold standard for assessing the functional severity of epicardial coronary artery lesions in the past decades [2]. Recently, the instantaneous wave-free ratio (iFR), which can be measured without the need for vasodilator administration, has emerged as an alternative index of stenosis severity [3]. The concept of iFR is based on the hypothesis that there is a diastolic “wave-free” period (WFP) during the heartbeat period when coronary microvascular resistance is inherently stable and minimized [3]. In comparison with the measurement of FFR, measuring iFR is quicker and cheaper, and more importantly, it can avoid the potential side effects (e.g., breathlessness and chest tightness) associated with vasodilator infusion [3]. Clinical studies have shown that iFR is comparable to FFR with respect to diagnostic categorization [4] and that revascularization strategies guided by iFR are noninferior to those guided by FFR with respect to the risk of major adverse cardiac events at 12 months in patients with stable angina or acute coronary syndrome [5, 6]. A study comparing FFR and iFR with a third ischemic test (e.g., positron emission tomography myocardial perfusion imaging) as the arbiter showed that iFR did not perform differently from FFR in identifying hemodynamically significant ischemic coronary lesions [7]. On the other hand, some studies have revealed that iFR only correlated weakly with FFR in patients whose FFRs were in the clinically important range for decision making of 0.60 to 0.90 [8] and that the overall diagnostic accuracy of iFR (using a ROC-determined cutoff value of 0.90) was about 80% when FFR was used as the reference index for diagnosis (i.e., discordant diagnosis with iFR and FFR occurred in over 20% patients) [9]. So far, reasons underlying the discordant diagnosis between iFR and FFR remain incompletely elucidated. The study by Lee et al. [10] found that patients with discordant iFR and FFR (i.e., negative iFR while positive FFR) usually had higher hyperemic myocardial blood flow and CFR (coronary flow reserve) and higher resting microvascular resistance while there was greater reduction of coronary microvascular resistance at hyperemia compared to patients with concordant iFR and FFR, which implies that the resting state and hyperemic response of coronary microvasculature may be important factors related to the diagnostic agreement between iFR and FFR [11]. In addition, the functional status of the aortic valve has also been demonstrated to affect the diagnostic performance of iFR compared with FFR [12, 13]. For instance, it was found that the diagnostic accuracy of iFR in predicting an FFR of ≤0.8 was poor (65%) for coronary lesions in patients with severe aortic valve obstruction and tended to improve after transcatheter aortic valve implantation (TAVI) [13]. Despite the valuable insights from these clinical studies, a systemic analysis of the relationship between iFR and FFR under various pathophysiological conditions with identical severity of coronary artery disease remains absent due to the difficulties in completely removing the effects of interpatient variability and measuring all cardiovascular and hemodynamic parameters necessary for analysis in general clinical settings.

In comparison with in vivo measurements, computational modeling methods provide a more convenient approach to quantifying the impacts of any pathophysiological factors of interest on iFR and FFR at fixed severities of coronary artery disease and thereby establishing a basis for exploring mechanisms underlying the discordant diagnosis between iFR and FFR. Computational modeling methods have been widely applied in conjunction with medical image-based model reconstruction techniques to predict iFR and/or FFR [1420]. While most model-based studies have demonstrated the ability of computational models to predict iFR and/or FFR with good accuracy in comparison with their in vivo counterparts, few studies have been dedicated to addressing the relationship between iFR and FFR over a wide range of pathophysiological conditions.

In the present study, a computational modeling method was employed to quantitatively investigate the respective sensitivities of iFR and FFR to various cardiovascular factors whose pathophysiological states are expected to differ among patients and have considerable influence on systemic and/or coronary hemodynamics and, based on this, identify the specific conditions under which iFR and FFR are most likely to give discordant diagnostic results.

2. Materials and Methods

2.1. Configuration of the Computational Model

The computational model was adapted from the models developed in our previous studies [21, 22] where the modeling methods and associated numerical schemes have been described in detail. In brief, a zero-one-dimensional (0-1-D) multiscale modeling method was employed to represent the coronary circulation coupled to the global cardiovascular system. 1-D modeling was applied to large epicardial coronary arteries and systemic arteries to describe pulse wave propagation and pressure/flow waveforms, while 0-D modeling was applied to intramyocardial vessels and the rest of the global cardiovascular system to describe intramyocardial and systemic hemodynamics. Coupling of the 1-D and 0-D models yielded a closed-loop model capable of describing both coronary and systemic hemodynamics as well as their interaction (see Figure 1). More importantly, the model provided a flexible platform for simulating iFR and FFR under various pathophysiological conditions through modifying the values of model parameters that represent various cardiovascular properties.

2.2. Parameter Assignment and Model Calibration

Model parameters were initially assigned based on population-averaged data reported in the literature [23, 24] to let model predictions fall in the ranges of in vivo hemodynamic data acquired from healthy subjects. It is noted that parameter assignment was implemented for nonhyperemic resting and hyperemic conditions, respectively, in order to meet the requirement for simulating iFR and FFR. Comparisons of model predictions and in vivo measurements under resting and hyperemic conditions are summarized in Table 1. In cases when aortic valve diseases (e.g., stenosis and regurgitation) were present, model parameters such as resting intramyocardial vascular resistance and systemic vascular resistance were further adjusted so that model-predicted coronary arterial flow and systemic arterial pressure were comparable to those measured in patients with aortic valve disease [25, 26] (refer to [21] for more details).


RestingHyperemic
In vivo measurementSimulationIn vivo measurementSimulation

QLAD (mL/min)76.15 ± 33.41 [24]86.60256.15 ± 110.84 [24]264.91
QLCx (mL/min)54.62 ± 24.59 [24]64.40163.85 ± 67.18 [24]171.26
QRCA (mL/min)68.46 ± 31.87 [24]72.00217.69 ± 76.70 [24]232.54
Pas (mmHg)113.0 ± 5.0 [23]121.30113.00 ± 6.0 [23]111.68
Pad (mmHg)74.0 ± 8.0 [23]79.7070.00 ± 5.0 [23]74.93
CO (L/min)5.19 ± 0.83 [23]5.147.6 ± 1.19 [23]7.49

Q, mean flow rate over a cardiac cycle; , aortic systolic/diastolic pressure; CO, cardiac output.
2.3. Definitions of iFR and FFR

iFR was defined as the ratio between the mean values of poststenosis coronary arterial and aortic blood pressures (herein denoted by and , respectively) during the diastolic wave-free period (WFP) under the nonhyperemic resting condition [3]. Herein, WFP was set to begin in 25% of the way into diastole and end 5 ms before the end of diastole in accordance with the general definition of iFR in clinical practice [3]:

FFR was defined as the ratio between the mean poststenosis coronary arterial and aortic blood pressures (herein denoted by and , respectively) during the entire cardiac cycle under the hyperemic condition [1]:

2.4. Baseline Computation Conditions

A stenosis was introduced in the middle segment of the left anterior descending coronary artery (LAD) (see Figure 1 for the location), with its length being fixed at 10 mm while the diameter stenosis rate (SR) varied from 0% (i.e., no stenosis) to 70% (i.e., severe stenosis). Heart rate (HR) was set to 66 beats/minute and 90 beats/minute for normal resting and hyperemic conditions, respectively.

2.5. Sensitivity Analyses of iFR and FFR with respect to Cardiovascular Factors

Physiologically, iFR and FFR could be affected by any cardiovascular factors involved in the regulation of coronary and/or systemic hemodynamics irrespective of whether they are related to the severity of coronary artery disease or not. In the present study, we considered six representative factors and categorized them into three groups: (1) cardiac factors, which include aortic valve function, the systolic and diastolic functions of the left ventricle, and heart rate (HR) that affects the magnitude and shape of aortic/cardiac blood pressure wave, as well as the extravascular tissue pressure of intramyocardial coronary vessels; (2) systemic vascular factors, which include the stiffness of the aorta and total systemic vascular resistance that affects the amplitude and mean value of aortic pressure wave, respectively; and (3) coronary vascular factors, which mainly include coronary microvascular resistance, a major determinant of trans-stenosis blood flow rate and pressure drop given coronary perfusion pressure and severity of stenosis.

2.5.1. Parametric Representations of Cardiovascular Factors in the Model

All the aforementioned cardiovascular factors were represented in the model with parameters that can be quantitatively modified to reflect the variations in the pathophysiological states of the factors.

The status of aortic valve function was controlled by the effective orifice areas of the aortic valve during diastole and systole (herein denoted by EOAdia and EOAsys, respectively). Assigning a value of >0 cm2 to EOAdia represents the presence of aortic valve regurgitation (AR), whereas assigning a value lower than 4 cm2 (i.e., the normal value of EOAsys) to EOAsys represents the presence of aortic valve stenosis (AS). Accordingly, progressively increasing EOAdia (from 0 to 0.3 cm2) and reducing EOAsys (from 4 to 1.0 cm2) represent the increasing severities of AR and AS, respectively. The systolic and diastolic functions of the left ventricle were parametrically represented by the peak systolic elastance (Elva) and baseline diastolic elastance (Elvb), respectively. Increasing Elva represents the enhancement in myocardial contractility during systole, whereas increasing Elvb represents the stiffening of the ventricular chamber (or impairment in myocardial relaxation) during diastole. HR was assigned directly in the model.

The stiffness of the aorta was controlled by the value assigned to the elastic modulus of the aortic wall in the model. Since the elastic modulus of the aortic wall is the main determinant of the aortic pulse wave velocity (aPWV), we herein took aPWV as a measure of aortic stiffness. An increase in aPWV corresponds to an increase in aortic stiffness. The total systemic vascular resistance (Rsys) and coronary microvascular resistance (Rcmv) are holistic descriptions of vascular resistances distributed in systemic tissues/organs and myocardium, respectively, and were modified by simultaneously varying all the corresponding vascular resistances.

2.5.2. Quantification of the Sensitivities of iFR and FFR to Cardiovascular Factors

In order to investigate how iFR/FFR is affected by varying the pathophysiological state of each aforementioned cardiovascular factor, we incrementally changed the value (values) of the model parameter (parameters) corresponding to the factor while fixing other model parameters at their reference states. In other words, we performed a one-at-a-time parametric study using the computational model to evaluate the sensitivity of iFR/FFR with respect to each individual cardiovascular factor. The range of variations in each model parameter was estimated based on clinical data measured under the nonhyperemic resting condition [22, 2740] and is listed along with its reference value in Table 2. It is noted that for the purpose of simplicity, we assumed that the ranges of parameter variations relative to their reference values under the hyperemic condition were the same as those assigned for the resting condition. In all the sensitivity analyses, the severity of the mid-LAD stenosis was fixed at 50% or 70%.


Model parameterReference value (resting/hyperemic)Range of variation (resting)

EOAdia (cm2)0.0/0.0(0.0∼0.3) [27]
EOAsys (cm2)4.0/4.0(4.0∼1.0) [28, 29]
Elva (mmHg/ml)2.87/2.87(1.435∼6.601) [30]
Elvb (mmHg/ml)0.056/0.056(0.028∼0.112) [30]
HR (beats/min)66/90(48∼111) [31, 32]
aPWV (m/s)4.7/4.7(3.478∼10.011) [3335]
Rsys (mmHg·s/ml)1.14/0.98(0.456∼1.824) [36, 37]
Rcmv (mmHg·s/ml)196.97/45.94(157.58∼433.33) [3840]

Note that the ranges of parameter variations under resting condition were estimated based on available clinical data reported in the literature.

The percentage difference of computed iFR/FFR relative to its reference value (computed with all parameters being held at their reference states) was then calculated to evaluate the impact of varying each model parameter on iFR/FFR. It is noted that due to the differential physiological conditions under which iFR and FFR are measured, there were two sets of reference values of model parameters: (1) one set corresponding to the intact resting condition, and (2) the other set corresponding to the hyperemic condition.

3. Results

3.1. Changes in iFR and FFR with the Severity of Coronary Artery Stenosis and Typical Hemodynamic Characteristics during iFR Measurement

Numerical simulations were firstly carried out to simulate iFR and FFR, respectively, with the severity of the mid-LAD stenosis being increased incrementally from 0% (no stenosis) to 70% (severe stenosis) while other cardiovascular factors fixed at their reference resting or hyperemic states. The simulated values of iFR and FFR both decreased monotonously with the severity of stenosis (see Figure 2). If a FFR of 0.8 was taken as the threshold for identifying a physiologically significant lesion [41], the corresponding iFR was 0.913, a value close to the cutoff value (0.89–0.93) established in previous clinical studies [12, 42, 43]. These results indicate that our model can reasonably predict the general relationship between iFR and FFR in the context of various severities of coronary artery stenosis.

Figure 3 shows the model-simulated pressure waves in the ascending aorta and those immediately distal to a 50% stenosis in mid-LAD under the control condition (i.e., all model parameters were at the reference state) and under two altered physiological conditions characterized by a 67% increase in HR and a 200% elevation in aPWV, respectively. The wave-free pressure portions used to calculate iFR are highlighted by the gray shadows. Figure 3 also shows the corresponding time histories of wave intensity (WI) in the LAD (Figures 3(d)–3(f)) and total resistance of coronary vessels distal to the stenosis (Figures 3(g)–3(i)). As expected, the variations in HR and aPWV both led to considerable changes in pressure waveform and time history of WI via their influence on pressure wave propagation and reflection in the systemic arterial system, but they had little influence on iFR. In the wave-free period (WFP), WI was close to zero, proving that the “wave-free” assumption in the definition of iFR is reasonable; however, the poststenosis coronary vascular resistance was not constant during WFP. Nevertheless, the relatively low value of poststenosis vascular resistance during WFP compared with that in systole can still partly support the clinical hypothesis that iFR is an index derived under the condition of low coronary vascular resistance.

3.2. Sensitivities of iFR and FFR to Variations in the State of Each Cardiovascular Factor

The sensitivities of iFR and FFR to variations in each of the eight model parameters that represent various cardiac or vascular factors are presented in the form of percentage changes relative to the reference values of iFR and FFR in Figure 4.

As for the sensitivities of iFR and FFR to cardiac factors (represented by EOAdia, EOAsys, Elva, Elvb, and HR in the model) (see Figures 4(a)4(e)), iFR was observed to be highly sensitive to both EOAdia and EOAsys that represent the status of the aortic valve function, whereas FFR was only mildly affected by the variations in EOAdia or EOAsys. Moreover, varying EOAdia or EOAsys induced opposite changes in iFR and FFR. For instance, increasing EOAdia (representing a progressive deterioration in AR) or decreasing EOAsys (representing a progressive deterioration in AS) remarkably reduced iFR whilst it elevated FFR mildly. When the results of the sensitivity analyses were further investigated with respect to the severity of coronary artery stenosis, an increase in stenosis rate (i.e., from 50% to 70%) was observed to considerably augment the sensitivities of iFR and FFR to EOAdia and EOAsys. Relatively, both iFR and FFR were insensitive to the systolic and diastolic functions of the left ventricle (represented by Elva and Elvb) and HR.

Varying the systemic vascular factors (i.e., aortic stiffness represented by aPWV and total systemic vascular resistance (Rsys)) induced detectable while only mild changes in iFR and FFR (see Figures 4(f) and 4(g)). As is different from systemic vascular factors, increasing coronary microvascular resistance (Rcmv) under the resting or hyperemic condition tended to significantly elevate iFR and FFR, although the degree of elevation in FFR was larger than that in iFR (see Figure 4(h)).

In summary, if a maximal percentage change in iFR or FFR of >10% in response to the variations in a model parameter was set as the threshold for judging high sensitivity, iFR was observed to be highly sensitive to EOAdia, EOAsys (aortic valve function), and Rcmv (state of coronary microvasculature), whereas FFR was solely sensitive to Rcmv.

3.3. Hemodynamic Characteristics Underlying the Differential Sensitivities of iFR and FFR to Aortic Valve Function and Coronary Microvascular Resistance

In order to explore hemodynamic characteristics underlying the differential sensitivities of iFR and FFR to aortic valve function and coronary microvascular resistance, taking the 50% mid-LAD stenosis as an example, we plotted the model-simulated aortic pressure wave and poststenosis coronary arterial pressure/flow waves and poststenosis coronary microvascular resistance under the control condition (i.e., all model parameters were fixed at their reference states) against those under the condition characterized by the presence of severe aortic valve stenosis (AS) (represented by setting EOAsys = 1.0 cm2) or increased coronary microvascular resistance (represented by increasing Rcmv by 120%) in Figure 5. It is noted that the numerical simulations were performed under the resting and hyperemic conditions, respectively, in consideration of the differential physiological conditions corresponding to iFR and FFR measurements.

Under the resting condition, although the presence of severe AS induced a marked decrease in both aortic and poststenosis coronary pressures, the degree of decrease in poststenosis pressure was larger than that of aortic pressure, resulting in an evident decrease in iFR. The enhanced decrease in poststenosis pressure was caused mainly by the increased resting coronary blood flow (which augments the pressure drop across the stenosis) as a consequence of coronary microvascular adaptive responses to increased myocardial stress and oxygen demand in the presence of AS [21]. Under the hyperemic condition, the simulated coronary blood flow rate in the presence of AS was however comparable to or even slightly lower than that under the control condition (which is consistent with previous clinical observations [44]), leading to a mild increase in FFR.

In contrast to AS, increasing coronary microvascular resistance under the resting or hyperemic condition had an overall small influence on the aortic pressure, but significantly elevated the poststenosis coronary pressure primarily due to its role in reducing trans-stenosis flow rate. Such effects were particularly pronounced under the hyperemic condition when the flow rate was higher and more sensitive to the variation in poststenosis coronary vascular resistance (see Figure 5(h)) compared with the resting condition, thereby leading to a larger increase in FFR than in iFR.

4. Discussion

In the present study, we employed a computational model to simulate the processes of iFR and FFR measurements and quantitatively investigated the respective sensitivities of iFR and FFR to various cardiovascular factors involved in the regulation of systemic and/or coronary hemodynamics. The results revealed that iFR and FFR differed considerably with respect to the cardiovascular factors to which they are sensitive and the degree and/or pattern of changes in response to the variations in the state of each cardiovascular factor.

The model-predicted marked decrease in iFR while mild increase in FFR following increasing severity of AS (simulated by reducing the value of EOAsys in the model) implies that in patients with severe AS, the measured iFRs may be much lower than those in patients with equivalent severity of coronary artery disease while normal aortic valve function, although the measured FFRs in the two patient cohorts might be comparable, which may consequently lead to increased probability of discordant diagnosis between iFR and FFR in the former patient cohort if cutoff values of iFR and FFR established based on clinical data acquired from the latter patient cohort were used. These theoretical findings are consistent with relevant clinical observations reported in the literature. For instance, it was found that in patients with severe AS, the conventional iFR cutoff value had lower diagnostic agreement with FFR in the classification of coronary lesions and that a lower iFR cutoff value (e.g., shifting the cutoff value from 0.89 to 0.83) should be used in order to better predict a positive FFR [12, 13, 45, 46]. In the case of increasing severity of AR (simulated by increasing the value assigned to EOAdia in the model), our study revealed similar patterns of differential changes in iFR and FFR to those found in the case of increasing severity of AS and would cause a similar trend of discordant diagnosis between iFR and FFR, although relevant clinical evidence from studies focused on patients with AR is rare, probably due to the low prevalence of AR in patients with coronary artery disease [47].

Unlike aortic valve disease which affects iFR and FFR in opposite ways, increasing coronary microvascular resistance led to a considerable increase in both iFR and FFR, although the degree of increase in iFR was lower than that in FFR. The differential effects of coronary microvascular resistance on iFR and FFR would become more evident when the resting coronary microvascular resistance is preserved while the hyperemic counterpart is higher than the normal value due to impaired vasodilation function, which may explain why low iFR and high FFR (i.e., iFR+/FFR−) were more frequently observed in patients with diabetes mellitus who usually have increased coronary microvascular resistance and low coronary flow at hyperemia due to microcirculation dysfunction [48, 49].

Relatively, varying left ventricular systolic and diastolic functions and HR and systemic vascular factors (i.e., aortic stiffness and systemic vascular resistance) over large ranges only had mild influences on iFR and FFR, which indicates that iFR and FFR would both perform well in assessing the functional severity of coronary artery lesions irrespective of potential high interpatient variability in these cardiac or vascular properties.

In summary, the present study demonstrates the general trend that iFR and FFR are more likely to give discordant diagnostic results in the presence of severe aortic valve disease (stenosis or regurgitation) or increased coronary microvascular resistance. Therefore, special caution should be taken in the interpretation of measured iFR and FFR or the use of general cutoff values for diagnosis in patients with these specific cardiovascular conditions. Furthermore, given the differential effects on iFR and FFR of aortic valve disease and increased coronary microvascular resistance, the changes in iFR and FFR would become more complex in the presence of aortic valve disease combined with increased coronary microvascular resistance. Our additional numerical tests revealed that increasing coronary microvascular resistance could counteract or even reverse the decrease in iFR whilst augmenting the increase in FFR caused by aortic valve stenosis (see Figure 6). In this sense, in patients suffering from concomitant aortic valve disease and coronary microcirculation dysfunction, the diagnostic agreement between iFR and FFR could be highly complex and should be carefully interpreted in the context of patient-specific conditions.

5. Limitations

While our study, through quantifying the respective sensitivities of iFR and FFR to the variations in the pathophysiological state of each individual cardiovascular factor, provided useful insights for exploring mechanisms underlying the clinically observed discordant diagnosis between iFR and FFR in some patient cohorts, the study is limited by its theoretical nature and the focus on single-factor sensitivity analyses that render the findings unable to be applied directly to explain the measurements in individual patients whose cardiovascular conditions are highly complex and may deviate significantly from those represented by the model. In addition, the numerical simulations tailored to single-factor sensitivity analyses were not sufficient to generate a large database for statistical determination of the cutoff values of iFR or FFR under specific pathological conditions (e.g., various types and severities of aortic valve disease combined with other cardiovascular abnormalities). For this purpose, large-scale stochastic numerical simulations (similar to those reported in [50]) that cover a wide range of various pathophysiological conditions would be needed. The 0-1-D multiscale model employed in the present study is however computationally costly and therefore not well suited to such a study. The problem might be solved by developing a lumped-parameter model that contains the main components of the present model whilst is computationally much cheaper, which would be addressed in our future studies.

6. Conclusion

A computational model-based numerical study has been carried out to compare the sensitivities of iFR and FFR to variations in the pathophysiological states of various cardiovascular factors. It was found that aortic valve disease and increased coronary microvascular resistance had considerable while differential influences on iFR and FFR, which provides theoretical evidence for explaining the increased risk of discordant diagnosis between iFR and FFR in patients with aortic valve disease or coronary microcirculation dysfunction.

Data Availability

All clinical data used in the present study have been reported in the corresponding references and the values of most model parameters not provided in the manuscript have been reported in a previous study of ours [21] (https://onlinelibrary.wiley.com/doi/full/10.1002/cnm.3257).

Conflicts of Interest

The authors have no conflicts of interest to declare.

Authors’ Contributions

Xinyang Ge built the computational model, performed numerical simulations, and drafted the manuscript. Youjun Liu and Shengxian Tu analyzed the data and revised the manuscript. Zhaofang Yin and Yuqi Fan interpreted the data from the clinical point of view. Sergey Simakov and Yuri Vassilevski optimized the computational model and revised the manuscript. Fuyou Liang designed the study, interpreted the data, and revised the manuscript. All authors have read and approved the final manuscript.

Acknowledgments

The study was supported in part by the National Natural Science Foundation of China (Grant nos. 11832003, 11972231, and 81611530715) and the SJTU Medical-Engineering Cross-cutting Research Project (Grant no. YG2016MS09).

References

  1. N. H. J. Pijls, B. De Bruyne, K. Peels et al., “Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses,” New England Journal of Medicine, vol. 334, no. 26, pp. 1703–1708, 1996. View at: Publisher Site | Google Scholar
  2. T. Force, S. Windecker, P. Kolh et al., “ESC/EACTS guidelines on myocardial revascularization: the task force on myocardial revascularization of the European society of cardiology (ESC) and the European association for cardio-thoracic surgery (EACTS) developed with the special contribution of the European association of percutaneous cardiovascular interventions (EAPCI),” European Heart Journal, vol. 35, no. 37, pp. 2541–2619, 2014. View at: Publisher Site | Google Scholar
  3. S. Sen, J. Escaned, I. S. Malik et al., “Development and validation of a new adenosine-independent index of stenosis severity from coronary wave-intensity analysis,” Journal of the American College of Cardiology, vol. 59, no. 15, pp. 1392–1402, 2012. View at: Publisher Site | Google Scholar
  4. S. Sen, K. N. Asrress, S. Nijjer et al., “Diagnostic classification of the instantaneous wave-free ratio is equivalent to fractional flow reserve and is not improved with adenosine administration,” Journal of the American College of Cardiology, vol. 61, no. 13, pp. 1409–1420, 2013. View at: Publisher Site | Google Scholar
  5. J. E. Davies, S. Sen, H. M. Dehbi et al., “Use of the instantaneous wave-free ratio or fractional flow reserve in PCI,” The New England Journal of Medicine, vol. 376, no. 376, pp. 1824–1834, 2017. View at: Publisher Site | Google Scholar
  6. M. Götberg, E. H. Christiansen, I. J. Gudmundsdottir et al., “Instantaneous wave-free ratio versus fractional flow reserve to guide PCI,” New England Journal of Medicine, vol. 376, no. 19, pp. 1813–1823, 2017. View at: Publisher Site | Google Scholar
  7. G. A. de Waard, I. Danad, R. Petraco et al., “Fractional flow reserve, instantaneous wave-free ratio, and resting Pd/Pa compared with [15O]H2O positron emission tomography myocardial perfusion imaging: a PACIFIC trial sub-study,” European Heart Journal, vol. 39, no. 46, pp. 4072–4081, 2018. View at: Publisher Site | Google Scholar
  8. C. Berry, M. van’t Veer, N. Witt et al., “VERIFY (VERification of instantaneous wave-free ratio and fractional flow reserve for the assessment of coronary artery stenosis severity in EverydaY practice),” Journal of the American College of Cardiology, vol. 61, no. 13, pp. 1421–1427, 2013. View at: Publisher Site | Google Scholar
  9. A. Jeremias, A. Maehara, P. Généreux et al., “Multicenter core laboratory comparison of the instantaneous wave-free ratio and resting Pd/Pa with fractional flow reserve,” Journal of the American College of Cardiology, vol. 63, no. 13, pp. 1253–1261, 2014. View at: Publisher Site | Google Scholar
  10. J. M. Lee, D. Hwang, J. Park, Y. Tong, and B.-K. Koo, “Physiologic mechanism of discordance between instantaneous wave-free ratio and fractional flow reserve: insight from 13 N-ammonium positron emission tomography,” International Journal of Cardiology, vol. 243, pp. 91–94, 2017. View at: Publisher Site | Google Scholar
  11. H. Matsuo, Y. Kawase, and I. Kawamura, “FFR and iFR,” Annals of Nuclear Cardiology, vol. 3, no. 1, pp. 53–60, 2017. View at: Publisher Site | Google Scholar
  12. R. Scarsini, G. Pesarini, C. Zivelonghi et al., “Coronary physiology in patients with severe aortic stenosis: comparison between fractional flow reserve and instantaneous wave-free ratio,” International Journal of Cardiology, vol. 243, pp. 40–46, 2017. View at: Publisher Site | Google Scholar
  13. R. Scarsini, G. Pesarini, C. Zivelonghi et al., “Physiologic evaluation of coronary lesions using instantaneous wave-free ratio (iFR) in patients with severe aortic stenosis undergoing transcatheter aortic valve implantation,” EuroIntervention, vol. 13, no. 13, pp. 1512–1519, 2018. View at: Publisher Site | Google Scholar
  14. C. A. Taylor, T. A. Fonte, and J. K. Min, “Computational fluid dynamics applied to cardiac computed tomography for noninvasive quantification of fractional flow reserve,” Journal of the American College of Cardiology, vol. 61, no. 22, pp. 2233–2241, 2013. View at: Publisher Site | Google Scholar
  15. Y. Ma, H. Liu, Y. Hou et al., “Instantaneous wave-free ratio derived from coronary computed tomography angiography in evaluation of ischemia-causing coronary stenosis: feasibility and initial clinical research,” Medicine, vol. 96, no. 4, p. e5979, 2017. View at: Publisher Site | Google Scholar
  16. L. Calmac, R. Niculescu, E. Badila et al., “TCT-40 image-based computation of instantaneous wave-free ratio from routine coronary angiography-initial validation by invasively measured coronary pressures,” Journal of the American College of Cardiology, vol. 66, no. 15, pp. B17–B18, 2015. View at: Publisher Site | Google Scholar
  17. W. Wang, B. Mao, B. Li et al., “Numerical simulation of instantaneous wave-free ratio of stenosed coronary artery,” International Journal of Computational Methods, vol. 16, no. 3, Article ID 1842009, 2019. View at: Publisher Site | Google Scholar
  18. J. M. Carson, S. Pant, C. Roobottom et al., “Non‐invasive coronary CT angiography‐derived fractional flow reserve: a benchmark study comparing the diagnostic performance of four different computational methodologies,” International Journal for Numerical Methods in Biomedical Engineering, vol. 35, no. 10, p. e3235, 2019. View at: Publisher Site | Google Scholar
  19. L. O. Müller, F. E. Fossan, A. T. Bråten, A. Jørgensen, R. Wiseth, and L. R. Hellevik, “Impact of baseline coronary flow and its distribution on Fractional Flow Reserve prediction,” International Journal for Numerical Methods in Biomedical Engineering, p. e3246, 2019. View at: Publisher Site | Google Scholar
  20. J. Carson, C. Roobottom, R. Alcock, and P. Nithiarasu, “Computational instantaneous wave‐free ratio (IFR) for patient‐specific coronary artery stenoses using 1D network models,” International Journal for Numerical Methods in Biomedical Engineering, vol. 35, no. 11, p. e3255, 2019. View at: Publisher Site | Google Scholar
  21. X. Ge, Y. Liu, S. Tu, S. Simakov, Y. Vassilevski, and F. Liang, “Model‐based analysis of the sensitivities and diagnostic implications of FFR and CFR under various pathological conditions,” International Journal for Numerical Methods in Biomedical Engineering, p. e3257, 2019. View at: Publisher Site | Google Scholar
  22. X. Ge, Z. Yin, Y. Fan, Y. Vassilevski, and F. Liang, “A multi-scale model of the coronary circulation applied to investigate transmural myocardial flow,” International Journal for Numerical Methods in Biomedical Engineering, vol. 34, no. 10, p. e3123, 2018. View at: Publisher Site | Google Scholar
  23. N. Watzinger, G. K. Lund, M. Saeed et al., “Myocardial blood flow in patients with dilated cardiomyopathy: quantitative assessment with velocity-encoded cine magnetic resonance imaging of the coronary sinus,” Journal of Magnetic Resonance Imaging, vol. 21, no. 4, pp. 347–353, 2005. View at: Publisher Site | Google Scholar
  24. H. Wieneke, C. Von Birgelen, M. Haude et al., “Determinants of coronary blood flow in humans: quantification by intracoronary Doppler and ultrasound,” Journal of Applied Physiology, vol. 98, no. 3, pp. 1076–1082, 2005. View at: Publisher Site | Google Scholar
  25. A. Nitenberg, J.-M. Foult, I. Antony, F. Blanchet, and M. Rahali, “Coronary flow and resistance reserve in patients with chronic aortic regurgitation, angina pectoris and normal coronary arteries,” Journal of the American College of Cardiology, vol. 11, no. 3, pp. 478–486, 1988. View at: Publisher Site | Google Scholar
  26. P. Meimoun, A. L. Germain, F. Elmkies et al., “Factors associated with noninvasive coronary flow reserve in severe aortic stenosis,” Journal of the American Society of Echocardiography, vol. 25, no. 8, pp. 835–841, 2012. View at: Publisher Site | Google Scholar
  27. M. Enriquez-Sarano and A. J. Tajik, “Aortic regurgitation,” New England Journal of Medicine, vol. 351, no. 15, pp. 1539–1546, 2004. View at: Publisher Site | Google Scholar
  28. D. Garcia, P. G. Camici, L.-G. Durand et al., “Impairment of coronary flow reserve in aortic stenosis,” Journal of Applied Physiology, vol. 106, no. 1, pp. 113–121, 2009. View at: Publisher Site | Google Scholar
  29. P. Pibarot and J. G. Dumesnil, “Improving assessment of aortic stenosis,” Journal of the American College of Cardiology, vol. 60, no. 3, pp. 169–180, 2012. View at: Publisher Site | Google Scholar
  30. F. Liang, H. Senzaki, C. Kurishima, K. Sughimoto, R. Inuzuka, and H. Liu, “Hemodynamic performance of the Fontan circulation compared with a normal biventricular circulation: a computational model study,” American Journal of Physiology-Heart and Circulatory Physiology, vol. 307, no. 7, pp. H1056–H1072, 2014. View at: Publisher Site | Google Scholar
  31. J. M. Mangrum and J. P. Dimarco, “The evaluation and management of bradycardia,” New England Journal of Medicine, vol. 342, no. 10, pp. 703–709, 2000. View at: Publisher Site | Google Scholar
  32. A. F. Chu, E. Zado, and F. E. Marchlinski, “Atrial arrhythmias in patients with arrhythmogenic right ventricular cardiomyopathy/dysplasia and ventricular tachycardia,” The American Journal of Cardiology, vol. 106, no. 5, pp. 720–722, 2010. View at: Publisher Site | Google Scholar
  33. P. V. Vaitkevicius, J. L. Fleg, J. H. Engel et al., “Effects of age and aerobic capacity on arterial stiffness in healthy adults,” Circulation, vol. 88, no. 4, pp. 1456–1462, 1993. View at: Publisher Site | Google Scholar
  34. R. P. Wildman, G. N. Farhat, A. S. Patel et al., “Weight change is associated with change in arterial stiffness among healthy young adults,” Hypertension, vol. 45, no. 2, pp. 187–192, 2005. View at: Publisher Site | Google Scholar
  35. Z. Dhakam, Yasmin, C. M. McEniery, T. Burton, M. J. Brown, and I. B. Wilkinson, “A comparison of atenolol and nebivolol in isolated systolic hypertension,” Journal of Hypertension, vol. 26, no. 2, pp. 351–356, 2008. View at: Publisher Site | Google Scholar
  36. W. G. Haynes, C. J. Ferro, K. P. J. O’Kane, D. Somerville, C. C. Lomax, and D. J. Webb, “Systemic endothelin receptor blockade decreases peripheral vascular resistance and blood pressure in humans,” Circulation, vol. 93, no. 10, pp. 1860–1870, 1996. View at: Publisher Site | Google Scholar
  37. T. Linde, B. Sandhagen, A. Hägg, C. Mörlin, B. Wikström, and B. G. Danielson, “Blood viscosity and peripheral vascular resistance in patients with untreated essential hypertension,” Journal of Hypertension, vol. 11, no. 7, pp. 731–736, 1993. View at: Publisher Site | Google Scholar
  38. S. E. Reis, R. Holubkov, J. S. Lee et al., “Coronary flow velocity response to adenosine characterizes coronary microvascular function in women with chest pain and no obstructive coronary disease,” Journal of the American College of Cardiology, vol. 33, no. 6, pp. 1469–1475, 1999. View at: Publisher Site | Google Scholar
  39. M. Fineschi, A. Bravi, and T. Gori, “The “slow coronary flow” phenomenon: evidence of preserved coronary flow reserve despite increased resting microvascular resistances,” International Journal of Cardiology, vol. 127, no. 3, pp. 358–361, 2008. View at: Publisher Site | Google Scholar
  40. R. F. Wilson, K. Wyche, B. V. Christensen, S. Zimmer, and D. D. Laxson, “Effects of adenosine on human coronary arterial circulation,” Circulation, vol. 82, no. 5, pp. 1595–1606, 1990. View at: Publisher Site | Google Scholar
  41. D.-Y. Kang, J.-M. Ahn, Y. W. Kim et al., “Impact of coronary lesion geometry on fractional flow reserve: data from interventional cardiology research in-cooperation society-fractional flow reserve and intravascular ultrasound registry,” Circulation: Cardiovascular Imaging, vol. 11, no. 6, Article ID e007087, 2018. View at: Publisher Site | Google Scholar
  42. L. Itu, T. Passerini, E. Badila et al., “Image-based computation of instantaneous wave-free ratio from routine coronary angiography: evaluation of a hybrid decision making strategy with ffr,” Journal of the American College of Cardiology, vol. 67, no. 13, p. 328, 2016. View at: Publisher Site | Google Scholar
  43. R. Scarsini, G. Pesarini, M. Lunardi et al., “Observations from a real-time, iFR-FFR “hybrid approach” in patients with severe aortic stenosis and coronary artery disease undergoing TAVI,” Cardiovascular Revascularization Medicine, vol. 19, no. 3, pp. 355–359, 2018. View at: Publisher Site | Google Scholar
  44. F. R. Eberli, M. Ritter, J. Schwitter et al., “Coronary reserve in patients with aortic valve disease before and after successful aortic valve replacement,” European Heart Journal, vol. 12, no. 2, pp. 127–138, 1991. View at: Publisher Site | Google Scholar
  45. F. Yamanaka, K. Shishido, T. Ochiai et al., “Instantaneous wave-free ratio for the assessment of intermediate coronary artery stenosis in patients with severe aortic valve stenosis: comparison with myocardial perfusion scintigraphy,” JACC: Cardiovascular Interventions, vol. 11, no. 20, pp. 2032–2040, 2018. View at: Publisher Site | Google Scholar
  46. R. Scarsini, R. Cantone, G. Venturi et al., “Correlation between intracoronary physiology and myocardial perfusion imaging in patients with severe aortic stenosis,” International Journal of Cardiology, vol. 292, pp. 162–165, 2019. View at: Publisher Site | Google Scholar
  47. A.-H. Hakki, D. Kimbiris, A. S. Iskandrian, B. L. Segal, G. S. Mintz, and C. E. Bemis, “Angina pectoris and coronary artery disease in patients with severe aortic valvular disease,” American Heart Journal, vol. 100, no. 4, pp. 441–449, 1980. View at: Publisher Site | Google Scholar
  48. C. M. Cook, A. Jeremias, R. Petraco et al., “Fractional flow reserve/instantaneous wave-free ratio discordance in angiographically intermediate coronary stenoses,” JACC: Cardiovascular Interventions, vol. 10, no. 24, pp. 2514–2524, 2017. View at: Publisher Site | Google Scholar
  49. J. M. Lee, E.-S. Shin, C.-W. Nam et al., “Discrepancy between fractional flow reserve and instantaneous wave-free ratio: clinical and angiographic characteristics,” International Journal of Cardiology, vol. 245, pp. 63–68, 2017. View at: Publisher Site | Google Scholar
  50. T. Wang, F. Liang, Z. Zhou, and X. Qi, “Global sensitivity analysis of hepatic venous pressure gradient (HVPG) measurement with a stochastic computational model of the hepatic circulation,” Computers in Biology and Medicine, vol. 97, pp. 124–136, 2018. View at: Publisher Site | Google Scholar

Copyright © 2020 Xinyang Ge et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.