Research Article

Scanning Electron Microscopic Assessment of Stent Coating Integrity in Jailed Wire Technique for Bifurcation Treatment

Table 2

Impact of the three types of wire on coating integrity in the procedure of JWT.

Total (n = 27)Runthrough NS wire (n = 9)HT BMW universal II wire (n = 9)#HT whisper wire (n = 9)# value

No defect3 (11.1)0 (0.0)1 (11.1)2 (22.2)<0.001
Slight defects11 (40.7)0 (0.0)4 (44.4)7 (77.8)
Moderate defects4 (14.8)0 (0.0)4 (44.4)0 (0.0)
Severe defects9 (33.3)9 (100.0)0 (0.0)0 (0.0)

compared with the Runthrough NS wire group. Comparison among three groups.