Xing Liu, Chun Gui, Weiming Wen, Yan He, Weiran Dai, Guoqiang Zhong, "Safety and Efficacy of High Power Shorter Duration Ablation Guided by Ablation Index or Lesion Size Index in Atrial Fibrillation Ablation: A Systematic Review and Meta-Analysis", Journal of Interventional Cardiology, vol. 2021, Article ID 5591590, 12 pages, 2021. https://doi.org/10.1155/2021/5591590

Safety and Efficacy of High Power Shorter Duration Ablation Guided by Ablation Index or Lesion Size Index in Atrial Fibrillation Ablation: A Systematic Review and Meta-Analysis

Academic Editor: Shenghua Zhou
Received12 Jan 2021
Revised22 Apr 2021
Accepted04 May 2021
Published03 Jun 2021

Abstract

Background. High power shorter duration (HPSD) ablation may lead to safe and rapid lesion formation. However, the optimal radio frequency power to achieve the desired ablation index (AI) or lesion size index (LSI) is insubstantial. This analysis aimed to appraise the clinical safety and efficacy of HPSD guided by AI or LSI (HPSD-AI or LSI) in patients with atrial fibrillation (AF). Methods. The Medline, PubMed, Embase, Web of Science, and the Cochrane Library databases from inception to November 2020 were searched for studies comparing HPSD-AI or LSI and low power longer duration (LPLD) ablation. Results. Seven trials with 1013 patients were included in the analysis. The analyses verified that HPSD-AI or LSI revealed benefits of first-pass pulmonary vein isolation (PVI) (RR: 1.28; 95% CI: 1.05–1.56, P = 0.01) and acute pulmonary vein reconnection (PVR) (RR: 0.65; 95% CI: 0.48–0.88, P = 0.005) compared with LPLD. HPSD-AI or LSI showed higher freedom from atrial tachyarrhythmia (AT) (RR = 1.32, 95% CI: 1.14–1.53, P = 0.0002) in the subgroup analysis of studies with PVI ± (with or without additional ablation beyond PVI). HPSD-AI or LSI could short procedural time (WMD: −22.81; 95% CI, −35.03 to −10.60, P = 0.0003), ablation time (WMD: −10.80; 95% CI: −13.14 to −8.46, P < .00001), and fluoroscopy time (WMD: −7.71; 95% CI: −13.71 to −1.71, P = 0.01). Major complications and esophageal lesion in HPSD-AI or LSI group were no more than LDLP group (RR: 0.58; 95% CI: 0.20–1.69, P = 0.32) and (RR: 0.84; 95% CI: 0.43–1.61, P = 0.59). Conclusions. HPSD-AI or LSI was efficient for treating AF with shorting procedural, ablation, and fluoroscopy time, higher first-pass PVI, and reducing acute PVR and may increase freedom from AT for patients with additional ablation beyond PVI compared with LPLD. Moreover, complications and esophageal lesion were low and no different between two groups.

1. Introduction

Compared to medical therapies alone, catheter ablation has been identified as an effective treatment for atrial fibrillation (AF), and quality of life of patients was significantly improved [1]. Pulmonary vein isolation (PVI) acted as the cornerstone for radiofrequency ablation of AF. The efficacy of radiofrequency catheter ablation (RFCA) is related to transmural, continuous, and cellular necrosis [2]. The conventional ablation therapy is mainly low power longer duration (LPLD). High power primarily increases the effect of resistive heating, while ablation duration produces conductive heating. Irreversible myocardial tissue damages with cellular necrosis are rapidly induced by resistive heating, whereas conductive heating passively stretches into deeper tissue layers, resulting in potential reversible tissue injuries. Moreover, it is quite difficult to retain catheter stability in a beating heart for a long time, and tissue edema caused by prolonged ablation hinders effective ablation [2], leading to the rate of pulmonary vein reconnection (PVR) that maintains frequently with LPLD. Simultaneously, LPLD ablation may generate damage depth excessively, thus increasing the risk of adjacent tissue damage, especially esophageal thermal injury (ETI) [3].

High power shorter duration (HPSD), as a novel ablation strategy, has been applied in AF treatment [4]. HPSD was safe and efficient for treating AF with shorting procedural and ablation time and higher first-pass pulmonary vein isolation (PVI), but it did not significantly reduce recurrence of atrial tachyarrhythmia (AT) compared with LPLD [5, 6]. Recurrent AT after PVI is generally associated with PVR, and gaps in the circumferential pulmonary veins (PVs) isolation lines are accompanied by increased recurrence of AF [7]. A weighted proprietary formula such as ablation index (AI) or lesion size index (LSI) incorporated with contact force (CF), radiofrequency (RF), application time, and power was reported to be beneficial to produce durable ablation lesion and to minimize AF recurrence following ablation [8, 9]. Recently, high-powered ablation guided by AI or LSI (HPSD-AI or LSI) was safe and procedural efficiency reduced with recurrence of AT [10, 11]. However, results of arrhythmia-related outcomes are contradictory and inconclusive [12, 13]. Therefore, we conducted systematic reviews and meta-analyses to evaluate the efficacy and safety of HPSD-AI or LSI compared with LPLD in treating AF.

2. Methods

2.1. Search Strategy

An all-round search was searched in the Medline, PubMed, Embase, Web of Science, and the Cochrane Library databases from inception up to November 2020 by two reviewers (XL and CG) independently. Articles in non-English languages were excluded. The following search strategy was applied to search PubMed, and we adapted it for the other databases: (“High-power” [Title/Abstract] OR ““HPSD” [Title/Abstract]) AND (“AF” [Title/Abstract] OR ((“atrial” [Title/Abstract] OR “atrium” [Title/Abstract] OR “auricular” [Title/Abstract]) AND (“fibrillation” [Title/Abstract] OR “arrhythmia [Title/Abstract] OR “flutter” [Title/Abstract])) OR (“Atrial Fibrillation” [MeSH Terms] OR “Atrial Flutter” [MeSH Terms])).

2.2. Inclusion and Exclusion Criteria

Two investigators (XL and WW) filtrated and identified research studies that fulfilled the following inclusion criteria: (1) full text studies of controlled experiments about HPSD-AI or LSI versus LPLD; LPLD: power ≤ 35 W, with a longer ablation duration of 10 to 30s per site; HPSD-AI or LSI: power ≥40 W, duration ≤ 10 s in ablation or less than LPLD group, with LSI ≥ 4 or AI ≥ 350 in sites on the LA posterior wall and LSI ≥ 5 or AI ≥ 400 in others; (2) patients with AF who consented radiofrequency ablation; (3) without a AF ablation history; (4) PVI applied using the contact force catheter; and (5) studies wanted to provide some dependable information with first-pass PVI, regarding procedure outcomes, acute PVR, either recurrence rates of AT including AF and atrial flutter, and complications in both groups. The exclusion criteria were as follows: (1) ablation used the noncontact force catheter; (2) studies enrolled less than 10 patients; and (3) animal studies, conference abstracts, case reports, review articles, editorials, or non-English language articles.

2.3. Quality Assessment

The study quality was evaluated by two investigators (WW and YH) using the Newcastle–Ottawa scale (NOS) for nonrandomized studies. And a star system (0–9) was used to judge studies. A research with NOS ≥7 was judged to be a study of good quality [14]. The quality of randomized controlled trials (RCTs) was evaluated by the Cochrane Collaboration tool for assessing risk of bias [15].

2.4. Data Extraction

Data were extracted using standardized protocol and reporting forms, including name of the first author, year of publication, country of origin, sample size, baseline characteristics (age, gender, left atrial diameter, and CHA2DS2-VASc), ablation strategy, ablation procedure details, AF type, ablation catheter type, the mapping system, freedom from AT, and procedure-related complications. The sample mean and standard deviation from commonly reported quantiles are estimated [16]. This data extraction process was performed independently by two investigators (XL and WD). Discrepancies between them were resolved by a third reviewer (CG).

2.5. Statistical Analysis

Dichotomous variables and outcome endpoints were reported as a risk ratio (RR) with 95% confidence intervals (CIs). The continuous variables were analyzed using weighted mean differences (WMD) or standard mean differences (SMD). The between-study heterogeneity was reflected by I2 > 50%, with deemed statistically significant. In cases of heterogeneity, random-effects models were used; otherwise (I2 ≤ 50%), fixed-effects models were preferentially used. In cases of statistical heterogeneity, subgroup analysis or sensitivity analyses were used. Sensitivity analysis was performed to determine the consistency of the overall effect estimate. When the pooled analysis still yielded significant heterogeneity, descriptive analysis was used. All values were two-tailed with a statistical significance set at 0.05. Publication bias was assessed by using the funnel plots. The statistical analysis was performed using the Revman5.4 software.

3. Results

3.1. Study and Data Selection

The results of the detailed search process are shown in Figure 1. Initially, 450 potentially relevant studies were yielded in our search strategy, of which 145 were duplicates and 248 were excluded after title and abstract review and abstracts. Of the remaining, 25 studies were excluded as topics were conducted in animals and conference, leaving a total of 32 studies for reading the full text. At this stage, further 25 studies were excluded after a detailed assessment of the full text due to the following: 5, uncontrolled trials; 3, no outcome of interests; 2, reporting duplicate date; and 15, ablation not abided by AI or LSI. No additional studies were added through manual search. Thus, 7 studies were finally selected in this meta-analysis [1013, 1719].

3.2. Study Characteristics and Quality Assessment of Included Studies

The characteristics of the included trials and ablation settings are summarized in Tables 1 and 2. A total of 1013 patients (409 patients underwent HPSD-AI or LSI strategy and 526 patients underwent LPLD strategy) were included in the analysis. There were four prospective cohort trials and two retrospective cohort trials and one RCT. There is no consensus about the power and AI or LSI for HPSD; in our study, energy levels at or above 40 W are considered as high power. The target ablation lesion index was reached: LSI ≥ 4 or AI ≥ 350 in sites on the LA posterior wall and LSI ≥ 5 or AI ≥ 400 in others [1013, 1719]. In the case of esophageal heating >38.5° or 39°, the AI target of the entire posterior ostium of that vein was lowered to ≥300 [17, 18]. Even the target AI was set at 260 on the esophagus in each ablation point [12]. The only RCT of energy difference between HPSD and LPLD was reflected only in the posterior wall ablation conducted by Leo et al. [11]. Meanwhile, we divided the study into two groups according to the difference of LSI in sites on the LA posterior wall (group 1, LSI of 4, group 2, LSI of 5). One trail by Okamatsu et al. [12] including three groups (low power, medium power, and high power) and medium-power group (≥40 W) was enrolled into HPSD-AI or LSI group according to the inclusion criteria of our study. Another trial by Castrejón-Castrejón et al. [18] containing the subgroup of power of 60 W was excluded because their ablation was not guided by LSI. In two studies [13, 17], PVI alone was performed except for cavotricuspid isthmus ablation because a typical atrial flutter was documented before or during the operation. One study [13] included only patients with paroxysmal atrial fibrillation.


StudyCountryStudy typeTreatment groupPatients (n)Follow (month)Age (years)Male (%)BMIDM (%)PAF (%)LVEF (%)LAD (mm)CHA2DS2-VASc

Leo et al. [11]United KingdomRandomized controlled trialHPSD-LSI 1202960.1 ± 9.46027.3 ± 5N/A4060.0 ± 9.241.4 ± 6.51 (IQR: 1–3)
LPLD-LSI 1202958.9 ± 9.29530.8 ± 4.6N/A4560.0 ± 11.543.0 ± 61 (IQR: 0–2)
HPSD-LSI 2202961.3 ± 9.67028.8 ± 4.9N/A4057.9 ± 6.443.7 ± 9.32 (IQR: 0–4)
LPLD-LSI 2202955.7 ± 107028 ± 4.85N/A3060.0 ± 10.242.4 ± 7.71 (IQR: 1-2)

Kaneshiro et al. [19]JapanRetrospective cohort trialHPSD-AI101N/A63 ± 1076N/AN/A66N/A40.8 ± 6.3N/A
LPLD170N/A61 ± 1081N/AN/A79N/A38.8 ± 6.5N/A

Berte et al. [17]SwitzerlandProspective cohort trialHPSD-AI80662 ± 972N/A408158 ± 8N/AN/A
LPLD-AI94663 ± 971N/A317959 ± 11N/AN/A
Okamatsu et al. [12]JapanProspective cohort trialHPSD-AI120665 ± 1065N/A256565 (IQR: 60–71)40 ± 62 (IQR: 1–3)
LPLD-AI20668 ± 875N/A58064 (IQR: 60–67)39 ± 62 (IQR: 1‐2)
HPSD-AI220664 ± 855N/A197564 (IQR: 59–71)40 ± 52 (IQR: 1–3)

Castrejon-Castrejon et al. [18]SpainProspective cohort trialHPSD50w-AI or LSI18N/AN/AN/AN/AN/AN/AN/AN/AN/A
HPSD60w30N/AN/AN/AN/AN/AN/AN/AN/AN/A
LPLD47N/AN/A6029 ± 5N/A6456 ± 11N/AN/A

Kyriakopoulou et al. [13]BelgiumRetrospective cohort trialHPSD-AI801267 (IQR: 58–73)5928 ± 5N/A100N/A43 ± 82 (IQR: 1–3)
LPLD-AI1051264 (IQR: 56–69)6227 ± 4N/A100N/A44 ± 62 (IQR: 1‐2)

Dhillon et al. [10]United KingdomProspective cohort trialHPSD-AI5012N/AN/AN/AN/AN/AN/AN/AN/A
LPLD5012N/AN/AN/AN/AN/AN/AN/AN/A

Values are mean ± SD, median (interquartile range), or n%; N/A, not available; AI, ablation index; BMI, body mass index; CHA2DS2-VASc, cardiac failure or dysfunction, hypertension, age ≥75 (doubled), diabetes, stroke (doubled)-vascular disease, age (65–74), and sex category (female); DM, diabetes mellitus; HPSD, high power and short duration; IQR, interquartile range; LAD, left atrial diameter; LPLD, low power longer duration; LSI, lesion size index; LVEF, left ventricular ejection fraction; PAF, paroxysmal atrial fibrillation.

StudyTreatment groupCF sensing catheter/Agilis sheathSTSF catheter/Agilis sheathMapping systemAnterior/Posterior wall powerLocal lesion endpointAblation strategy

Leo et al. [11]HPSD-LSI 1+/+EnSite40 WTarget LSI of 5.5–6 at the LA anterior wall and 4 at the posterior wallPVI ± line
LPLD-LSI 1+/+EnSite40 W/20 WTarget LSI of 5.5–6 at the LA anterior wall and 4 at the posterior wallPVI ± line
HPSD-LSI 2+/+EnSite40 WTarget LSI of 5.5–6 at the LA anterior wall and 5 at the posterior wallPVI ± line
LPLD-LSI 2+/+EnSite40 W/20 WTarget LSI of 5.5–6 at the LA anterior wall and 5 at the posterior wallPVI ± line

Kaneshiro et al. [19]HPSD-AI+/−CARTO45–50 WTarget AI of 400 at the LA posterior wallPVI ± line
LPLD+−/−+−/−CARTO20–30 WDuration at 10–30 s, CF 20–30 gPVI ± line

Berte et al. [17]HPSD-AI+/+CARTO45 W/35 WTarget AI of 550 at the LA anterior wall and 300–400 at the posterior wallPVI
LPLD-AI+/+CARTO35 W/25 WTarget AI of 550 at the LA anterior wall and 300–400 at the posterior wallPVI

Okamatsu et al. [12]HPSD-AI1+/+−CARTO50 W/30–40 WTarget AI of 400 at the LA anterior wall and 260–360 at the posterior wallPVI ± line ± box isolation ± CFAE
LPLD-AI+/+−CARTO30 W/20 WTarget AI of 400 at the LA anterior wall and 260–360 at the posterior wallPVI ± line ± box isolation ± CFAE
HPSD-AI2+/+−CARTO40 W/30 WTarget AI of 400 at the LA anterior wall and 260–360 at the posterior wallPVI ± line ± box isolation ± CFAE

Castrejon-Castrejon et al. [18]HPSD50w-AI or LSI+/+−CARTO/ EnSite50 WLSI≥ 5, AI≥ 350 at the LA posterior wall and ≥450 in othersPVI ± line
HPSD60w+/+−CARTO/EnSite60 W2–7 sPVI ± line
LPLD+/+−CARTO/EnSite30 W/20–30 W30–60 sPVI ± line

Kyriakopoulou et al. [13]HPSD-AI+/−CARTO40 WTarget AI of 550 at the LA anterior wall and 300–400 at the posterior wallPVI
LPLD-AI+/−CARTO35 WTarget AI of 550 at the LA anterior wall and 300–400 at the posterior wallPVI

Dhillon et al. [10]HPSD-AI+/−CARTO40 W/30 WTarget AI of 450 at the LA anterior wall and 350 at the posterior wallPVI ± line
LPLD+/−CARTO30 W/25 WCF 20–30 gPVI ± line

AI, ablation index; CF, contact force; CFAE, complex fractionated atrial electrogram; CTI, cava-tricuspid isthmus isolation; HPSD, high power shorter duration; LA, left atrial; LPLD, low power longer duration; LSI, lesion size index; PVI, pulmonary vein isolation; STSF, ThermoCool SmartTouch Surround Flow; SVCI, superior vena cava isolation.

Quality assessment of included studies is given in Table 3. None of the included studies was of poor quality.


StudyRepresentativeness of the exposed cohortSelection of the nonexposed cohortAscertainment of exposureDemonstration that outcome of interest was not present at start of the studyComparability of cohorts on the basis of the design or analysisAssessment of outcomeWas follow‐up long enough for outcomes to occurAdequacy of follow-up of cohortsTotal stars

Kaneshiro et al. [19]6
Berte et al. [17]9
Okamatsu et al. [12]9
Castrejon-Castrejon et al. [18]7
Kyriakopoulou et al. [13]8
Dhillon et al. [10]8
StudyRandom sequence generationAllocation concealmentBlinding of participants and personnelBlinding of outcome assessmentIncomplete outcome dataSelective reportingOther bias
Leo et al. [11]Low riskLow riskLow riskLow riskLow riskUnclear riskUnclear risk

3.3. First-Round Isolation Rate

6 studies [1013, 17, 18] reported the first-round isolation rate. The first-round isolation rate of PVs in the HPSD-AI or LSI group was significantly higher than in the LDLP group (RR: 1.28; 95% CI: 1.05–1.56, I2 = 92%, ) (Figure 2). Considering the high heterogeneity, the random-effects model was used for analysis. By sensitivity analysis by removing any individual study, the results did not change, indicating that the results were stable.

3.4. Acute PV Reconnection (APR) Rate

The APR rate was reported in 5 included studies [10, 11, 13, 17, 18] and the heterogeneity was low (I2 = 46%). The APR rate in the HPSD-AI or LSI group was significantly lower compared with the LDLP group (RR: 0.65; 95% CI: 0.48–0.88, ) (Figure 2).

3.5. Long-Term Freedom from AF/AT

More than 6 months follow-up outcomes were summarized from 5 studies [1013, 17], and the heterogeneity was moderate (I2 = 63%). More than 6 months success rate in the HPSD-AI or LSI group was higher than in the LDLP group (RR = 1.16, 95% CI: 1.01–1.34, ) (Figure 2). Subgroup analysis was performed according to ablation strategies to analyze the source of high heterogeneity. 3 studies [1012] PVI ± (with or without line, Box isolation or complex fractionated atrial electrogram ablation) were included, and the heterogeneity was very low (I2 = 4%). Long-term freedom from AF/AT 6 months or later after the AF ablation in the HPSD-AI or LSI group was also significantly higher than in the LDLP group (RR = 1.32, 95% CI: 1.14−1.53, ) (Figure 2). 2 studies [13, 17] with only PVI except a typical atrial flutter performed by cavotricuspid isthmus ablation was included, and the heterogeneity was very low (I2 = 0%). There was no significant difference in terms of recurrence of AF/AT in two groups (RR = 1.02, 95% CI: 0.94−1.11, ). The results of the sensitivity analysis were not altered by the deletion of any individual studies from the analysis.

3.6. Procedure Efficiency

Results including procedure, ablation, and fluoroscopy times were available in 5, 5, and 4 of the studies, respectively [10, 11, 13, 17, 18]. There was a significant reduction in the procedure time (WMD: −22.81; 95% CI: −35.03 to −10.60, I2 = 82%, ), ablation time (WMD: −10.80; 95% CI: −13.14 to −8.46, I2 = 53%, ), and fluoroscopy time (WMD: −7.71; 95% CI: −13.71 to −1.71, I2 = 95%, ) (Figure 3). Considering the high heterogeneity, the random-effects model was used for all analyses. The sensitivity analysis showed the results were not driven by any single study.

3.7. Procedural Complications

Procedural complications mainly referred to aterioesophageal fistula, pericardial effusion/cardiac tamponade, and stroke were reported in 6 studies [1013, 17, 18]. There were no significant differences in procedural complications between the two groups (RR: 0.58; 95% CI: 0.20−1.69, I2 = 0%, ) (Figure 4). Esophageal lesions were evaluated by esophagogastroduodenoscopy in two trials [18, 19]. There were no significant differences in esophageal lesions between the two groups (RR: 0.84; 95% CI: 0.43–1.61, I2 = 0%, ) (Figure 4). The fixed-effects model was used for analyses because of the very low heterogeneity. By removing any individual studies for sensitivity analysis, there was no significant change in the point estimate or CI in the results.

3.8. Publication Bias

We intended to investigate potential publication bias by funnel plots. However, since there were only as many as seven studies in our main analysis, the number was insufficient to reject the hypothesis of no funnel plot asymmetry. So we did not perform a funnel plot [20, 21].

4. Discussion

4.1. Major Findings

This study represented the first systematic review and meta-analysis on the comparison between HPSD-AI or LSI ablation and LPLD in patients with AF. The main findings were as follows: (1) HPSD-AI or LSI ablation showed higher first-round isolation rate and lower APR rate compared with LPLD, (2) The HPSD-AI or LSI group had a higher freedom from AF/AT 6 months or later after AF ablation than the LDLP group. There was a similarity between two groups in freedom from AF/AT among patients undergoing only PVI under the subgroup analysis of ablation strategy, but freedom from AF/AT rate was also significantly higher than in the LDLP group in the PVI ± subgroup, (3) HPSD-AI or LSI strategy could meaningfully shorter procedural, ablation, and fluoroscopy time compared with the LPLD, and (4) major complications and esophageal lesions were similar between two groups.

4.2. Clinical Efficacy

AI or LSI incorporated CF, RF application time, and power into a weighted proprietary formula and experimental research has revealed that lesion depth can be predicted accurately by the AI formula and power-made more contributions than CF at the initial time of ablation [8, 22]. Meanwhile, recent clinical trials targeting AI values of 550 at the anterior wall of left atrial (LA) and 400 at the LA posterior wall were related to high single operation success rate and low rate of PVR [9]. As a novel energy delivery strategy, HPSD was used to optimize LPLD. It is well known that catheter instability or poor contact may induce incomplete lesions and tissue edema during radio frequency delivery. In turn, it is difficult to achieve transmural injuries with further radiofrequency applications, resulting in conduction gaps and PV reconnections. On the contrary, HPSD can improve the stability of catheter in a short time and increase the injure area through predominant resistive heating [2]. Previous clinical studies using “uncontrolled” high power ablation for PVI showed a meaningfully shorter fluoroscopic time, procedural time, higher rate of first-pass PVI, and similar freedom from AF/AT rate [5, 23], in which results were consistent with our meta-analysis except for the last one. However, an observational study found that HPSD ablation was related to a higher risk of atrial flutter and a potential surrogate for incomplete sets/lines [24].

Recently, the study by Chen et al. [25] demonstrated that the initial 6-month follow-up showed 48 (96%) patients were free from clinical AF/AT recurrence by AI-guided 50 W ablation. Therefore, combining the superiority of the high-power ablation abided by the AI or LSI may better balance the procedural efficacy and safety. Winkle et al. [26] targeted LSI values of 5.5–6 for LA ablation at 50 W and reported a low complication rate and single procedure freedom from AF of 83% for paroxysmal AF and 72% for persistent AF at 2 years. By comparing HPSD-AI or LSI and LPLD studies, our analysis gets the same results as Chen et al. [25] who reported on their study that HPSD-AI or LSI was associated with increased more than 6-month freedom from AF/AT compared to LPLD ablation. Considering the high heterogeneity of the results, the subgroup analysis of different ablation strategies showed good homogeneity. The same conclusion was reached in the PVI ± subgroup, while freedom from AF/AT rate of HPSD-AI or LSI was not more than LDLP among patients with undergoing only PVI, indicating that HPSD-AI or LSI may increase freedom from AT for patients with additional ablation beyond PVI compared with LPLD. The possible reason is that HPSD-AI or LSI is superior than LPLD in improving the success rate of additional ablation beyond PVI and reducing the incidence of associated arrhythmias after radiofrequency ablation of AF. What merits our attention is that high power ablation can achieve the AI or LSI target in a shorter time. Importantly, clinical studies that used a significantly higher power of 70 W for 5–7 s and very higher power of 90 W for 4s have shown that therapeutic effects can be achieved, but AI or LSI has become irrelevant [27, 28]. AI is generally used as the local lesion endpoint only when ≤50 W because a reliable local lesion endpoint cannot be determined especially when very high power is used in a few seconds. This means that the operator needs to latently terminate the ablation lesion before the AI or LSI value is made visible. Otherwise, it would add the risk of overtreatment and potential complications. Therefore, large sample randomized controlled studies are needed to confirm how high the power is.

4.3. Procedural Efficiency

In terms of procedural efficiency, the pooled analysis revealed that HPSD-AI or LSI ablation can extraordinarily reduce the RF ablation time, procedure time, and fluoroscopy time compared to LDLP, which are consistent with a meta-analysis about comparison of HPSD and LDLP ablation [6], thus limiting patient exposure to intravenous fluids that could be beneficial in reducing the risk of postablation cognitive dysfunction [29]. Meanwhile, shorter radiation duration directly benefits the patient, operator, and supporting staff. In contrast, longer ablation time and procedure time in the LPLD group may increase surgical complications. Due to a reduction in RF time because of HPSD guided by AI or LSI formula, the procedure time obviously shortens. And shorter ablation time was because of the shorter time required for lesion creation, higher first-pass PVI, and fewer acute PV reconnections than those of LPLD, which are consistent with our analysis results.

4.4. Safety

The most concern was about the safety issue of HPSD-AI or LSI ablation. Under the premise of achieving fulfilling procedural efficiency and efficacy outcomes, HPSD-AI or LSI of major complications and esophageal lesions were similar with the LPLD group in our meta-analysis, which were consistent across all included study that reported this result [1013, 1719]. But, in the 5 included studies [10, 12, 13, 17, 18], cases of aterioesophageal fistula or cardiac tamponade were not observed in the HPSD-AI group or the LSI group. One case of cardiac tamponade occurring in the HPSD-LSI group reported by Leo et al. [11] was presumably due to inadvertent transseptal puncture via the transverse sinus instead of ablation. However, three (7%) patients in the LPLD group developed cardiac tamponade and required urgent pericardiocentesis, and one of them occurred following an audible steam pop during cavotricuspid isthmus ablation [18]. Importantly, whether high power can reduce esophageal damage is our concerned question. It is all known that left atrial-esophagus fistula is a fatal complication associated with PVI. HPSD approach can adjust the relationship between resistive and conductive heating, avoiding potential colleterial damage to adjacent structures such the esophagus [2]. Recent clinical studies [19] have shown that the incidence of ETI was significantly higher in the HPSD group compared to the LPLD group (37% vs. 22%, ), but the prevalence of esophageal lesions did not differ between the groups (7% vs. 8%). The use of the HPSD setting could avoid deeper thermal injuries that reach the esophageal mucosal layer because it was a strong predictor of ETI. All esophageal lesions inspected by gastroscopy in the HPSD group were mild erythema, and the esophageal lesions in the LPLD group showed ulceration, which also suggests that thermal injury could not reach the esophageal mucosal layer deeply when using the HPSD setting. Consistently, Wolf M et al. also reported low rate of esophageal lesions (1.2%, more than 7 days) following the AI-guided PVI [30]. Animal experiment on the pig model found that HPSD ablation can significantly reduce the lesion volume and cause less damage to the esophagus when AI is taken as a predefined target for different power settings [31]. Thus, a large sample randomized controlled study may conclude that HPSD-AI or LSI may cause less esophageal damage than LPLD.

5. Limitation

This meta-analysis has some limitations. First, there were variations in the high power definition and AI or LSI setting between the included studies as we analyzed. Second, there were different operator experiences, types of catheters, irrigation fluid delivery rate, and ablation strategy, all of which led to otherness in lesion formation. Data, which were extracted from the included studies, were not adjusted for these. Third, the included studies did not compare HPSD-AI or LSI with LPLD in patients with paroxysmal and persistent AF separately. Fourth, only seven studies with small sample size were included in our meta-analysis, and only one of them was RCT. Thus, more well-designed and large-scale RCTs with large sample size and longer term follow-up are demanded to validate the safety and efficiency of HPSD-AI or LSI strategy. Fifth, most included studies did not monitor esophageal temperature and perform gastroscopy, resulting in limitations in assessing esophageal damage.

6. Conclusions

Our systematic review and meta-analysis showed that HPSD-AI or LSI was effective method for AF ablation. Compared with the LPLD approach, it had some obvious advantages, including shorter procedure time, ablation time, and fluoroscopy time. In addition, HPSD-AI or LSI approach had higher first-pass PVI and lower acute PV reconnection and may increase freedom from AT for patients with additional ablation beyond PVI. Moreover, complications and esophageal lesion were low and there was no difference between the two groups. Further randomized multicenter studies with larger sample sizes and longer term follow-up are necessary to confirm the safety of HPSD-AI or LSI.

Data Availability

The data used to support the findings of this study are included within the article.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

References

  1. D. B. Mark, K. J. Anstrom, S. Sheng et al., “Effect of catheter ablation vs. medical therapy on quality of life among patients with atrial fibrillation,” JAMA, vol. 321, no. 13, pp. 1275–1285, 2019. View at: Publisher Site | Google Scholar
  2. E. Leshem, I. Zilberman, C. M. Tschabrunn et al., “High-power and short-duration ablation for pulmonary vein isolation,” JACC: Clinical Electrophysiology, vol. 4, no. 4, pp. 467–479, 2018. View at: Publisher Site | Google Scholar
  3. A. Thiyagarajah, K. Kadhim, D. H. Lau et al., “Feasibility, safety, and efficacy of posterior wall isolation during atrial fibrillation ablation: a systematic review and meta-analysis,” Circulation: Arrhythmia and Electrophysiology, vol. 12, no. 8, p. e007005, 2019. View at: Publisher Site | Google Scholar
  4. A. Bhaskaran, W. Chik, J. Pouliopoulos et al., “Five seconds of 50-60 W radio frequency atrial ablations were transmural and safe: an in vitro mechanistic assessment and force-controlled in vivo validation,” Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology, vol. 19, no. 5, pp. 874–880, 2017. View at: Publisher Site | Google Scholar
  5. D. G. Shin, J. Ahn, S.-J. Han, and H. E. Lim, “Efficacy of high-power and short-duration ablation in patients with atrial fibrillation: a prospective randomized controlled trial,” EP Europace, vol. 22, no. 10, pp. 1495–1501, 2020. View at: Publisher Site | Google Scholar
  6. X. Sun, P. Qi, B. Yang, Z. Li, Z. Bie, and X. Li, “The procedural efficiency, efficacy and safety of high power and short duration ablation in patients with atrial fibrillation: a systemic review and meta-analysis,” International Journal of Cardiology, vol. 325, pp. 76–81, 2020. View at: Publisher Site | Google Scholar
  7. K. H. Kuck, B. A. Hoffmann, S. Ernst et al., “Impact of complete versus incomplete circumferential lines around the pulmonary veins during catheter ablation of paroxysmal atrial fibrillation: results from the gap-atrial fibrillation-German atrial fibrillation competence network 1 trial,” Circulation: Arrhythmia and Electrophysiology, vol. 9, no. 1, p. e003337, 2016. View at: Publisher Site | Google Scholar
  8. V. Calzolari, L. De Mattia, S. Indiani et al., “In vitro validation of the lesion size index to predict lesion width and depth after irrigated radiofrequency ablation in a porcine model,” JACC: Clinical Electrophysiology, vol. 3, no. 10, pp. 1126–1135, 2017. View at: Publisher Site | Google Scholar
  9. P. Taghji, M. El Haddad, T. Phlips et al., “Evaluation of a strategy aiming to enclose the pulmonary veins with contiguous and optimized radiofrequency lesions in paroxysmal atrial fibrillation,” JACC: Clinical Electrophysiology, vol. 4, no. 1, pp. 99–108, 2018. View at: Publisher Site | Google Scholar
  10. G. Dhillon, S. Ahsan, S. Honarbakhsh et al., “A multicentered evaluation of ablation at higher power guided by ablation index: establishing ablation targets for pulmonary vein isolation,” Journal of Cardiovascular Electrophysiology, vol. 30, no. 3, pp. 357–365, 2019. View at: Publisher Site | Google Scholar
  11. M. Leo, M. Pedersen, K. Rajappan et al., “Power, lesion size index and oesophageal temperature alerts during atrial fibrillation ablation: a randomized study,” Circulation: Arrhythmia and Electrophysiology, vol. 13, no. 10, 2020. View at: Publisher Site | Google Scholar
  12. H. Okamatsu, J. Koyama, Y. Sakai et al., “High‐power application is associated with shorter procedure time and higher rate of first‐pass pulmonary vein isolation in ablation index‐guided atrial fibrillation ablation,” Journal of Cardiovascular Electrophysiology, vol. 30, no. 12, pp. 2751–2758, 2019. View at: Publisher Site | Google Scholar
  13. M. Kyriakopoulou, J. Y. Wielandts, T. Strisciuglio et al., “Evaluation of higher power delivery during RF pulmonary vein isolation using optimized and contiguous lesions,” Journal of Cardiovascular Electrophysiology, vol. 31, no. 5, pp. 1091–1098, 2020. View at: Publisher Site | Google Scholar
  14. A. Stang, “Critical evaluation of the Newcastle-Ottawa scale for the assessment of the quality of nonrandomized studies in meta-analyses,” European Journal of Epidemiology, vol. 25, no. 9, pp. 603–605, 2010. View at: Publisher Site | Google Scholar
  15. J. P. T. Higgins, D. G. Altman, P. C. Gotzsche et al., “The Cochrane Collaboration’s tool for assessing risk of bias in randomised trials,” BMJ, vol. 343, p. d5928, 2011. View at: Publisher Site | Google Scholar
  16. S. McGrath, X. Zhao, R. Steele et al., “Estimating the sample mean and standard deviation from commonly reported quantiles in meta-analysis,” Statistical Methods in Medical Research, 2020. View at: Publisher Site | Google Scholar
  17. B. Berte, G. Hilfiker, I. Russi et al., “Pulmonary vein isolation using a higher power shorter duration CLOSE protocol with a surround flow ablation catheter,” Journal of Cardiovascular Electrophysiology, vol. 30, no. 11, pp. 2199–2204, 2019. View at: Publisher Site | Google Scholar
  18. S. Castrejón-Castrejón, M. Martínez Cossiani, M. Ortega Molina et al., “Feasibility and safety of pulmonary vein isolation by high-power short-duration radiofrequency application: short-term results of the POWER-FAST PILOT study,” Journal of Interventional Cardiac Electrophysiology, vol. 57, no. 1, pp. 57–65, 2020. View at: Publisher Site | Google Scholar
  19. T. Kaneshiro, M. Kamioka, N. Hijioka et al., “Characteristics of esophageal injury in ablation of atrial fibrillation using a high-power short-duration setting,” Circulation: Arrhythmia and Electrophysiology, vol. 13, no. 10, p. e008602, 2020. View at: Publisher Site | Google Scholar
  20. M. Simmonds, “Quantifying the risk of error when interpreting funnel plots,” Systematic Reviews, vol. 4, p. 24, 2015. View at: Publisher Site | Google Scholar
  21. T. P. A. Debray, K. G. M. Moons, and R. D. Riley, “Detecting small-study effects and funnel plot asymmetry in meta-analysis of survival data: a comparison of new and existing tests,” Research Synthesis Methods, vol. 9, no. 1, pp. 41–50, 2018. View at: Publisher Site | Google Scholar
  22. A. Ikeda, H. Nakagawa, H. Lambert et al., “Relationship between catheter contact force and radiofrequency lesion size and incidence of steam pop in the beating canine heart,” Circulation: Arrhythmia and Electrophysiology, vol. 7, no. 6, pp. 1174–1180, 2014. View at: Publisher Site | Google Scholar
  23. T. Pambrun, C. Durand, M. Constantin et al., “High-power (40-50 W) radiofrequency ablation guided by unipolar signal modification for pulmonary vein isolation: experimental findings and clinical results,” Circulation: Arrhythmia and Electrophysiology, vol. 12, no. 6, p. e007304, 2019. View at: Publisher Site | Google Scholar
  24. T. J. Bunch, H. T. May, T. L. Bair et al., “Long-term outcomes after low power, slower movement versus high power, faster movement irrigated-tip catheter ablation for atrial fibrillation,” Heart Rhythm, vol. 17, no. 2, pp. 184–189, 2020. View at: Publisher Site | Google Scholar
  25. S. Chen, B. Schmidt, S. Bordignon et al., “Ablation index‐guided 50 W ablation for pulmonary vein isolation in patients with atrial fibrillation: procedural data, lesion analysis, and initial results from the FAFA AI High Power Study,” Journal of Cardiovascular Electrophysiology, vol. 30, no. 12, pp. 2724–2731, 2019. View at: Publisher Site | Google Scholar
  26. R. A. Winkle, R. Moskovitz, R. Hardwin Mead et al., “Atrial fibrillation ablation using very short duration 50 W ablations and contact force sensing catheters,” Journal of Interventional Cardiac Electrophysiology, vol. 52, no. 1, pp. 1–8, 2018. View at: Publisher Site | Google Scholar
  27. V. Y. Reddy, M. Grimaldi, T. De Potter et al., “Pulmonary vein isolation with very high power, short duration, temperature-controlled lesions,” JACC: Clinical Electrophysiology, vol. 5, no. 7, pp. 778–786, 2019. View at: Publisher Site | Google Scholar
  28. M. Kottmaier, M. Popa, F. Bourier et al., “Safety and outcome of very high-power short-duration ablation using 70 W for pulmonary vein isolation in patients with paroxysmal atrial fibrillation,” EP Europace, vol. 22, no. 3, pp. 388–393, 2020. View at: Publisher Site | Google Scholar
  29. C. Medi, L. Evered, B. Silbert et al., “Subtle post-procedural cognitive dysfunction after atrial fibrillation ablation,” Journal of the American College of Cardiology, vol. 62, no. 6, pp. 531–539, 2013. View at: Publisher Site | Google Scholar
  30. M. Wolf, M. El Haddad, V. De Wilde et al., “Endoscopic evaluation of the esophagus after catheter ablation of atrial fibrillation using contiguous and optimized radiofrequency applications,” Heart Rhythm, vol. 16, no. 7, pp. 1013–1020, 2019. View at: Publisher Site | Google Scholar
  31. T. Kawaji, S. Hojo, A. Kushiyama et al., “Limitations of lesion quality estimated by ablation index: an in vitro study,” Journal of Cardiovascular Electrophysiology, vol. 30, no. 6, pp. 926–933, 2019. View at: Publisher Site | Google Scholar

Copyright © 2021 Xing Liu et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Related articles

No related content is available yet for this article.
 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder
Views455
Downloads427
Citations

Related articles

No related content is available yet for this article.

Article of the Year Award: Outstanding research contributions of 2021, as selected by our Chief Editors. Read the winning articles.