Abstract

Bioprosthetic surgical aortic valve failure requiring reintervention is a frequent clinical problem with event rates up to 20% at 10 years after surgery. Transcatheter aortic valve-in-valve implantation (ViV-TAVI) has become a valuable treatment option for these patients, although it requires careful procedural planning. We here describe and illustrate a stepwise approach to plan and perform ViV-TAVI and discuss preprocedural computerized tomography planning, transcatheter heart valve selection, and implantation techniques. Particular attention is paid to coronary artery protection and the possible need for bioprosthetic valve fracture since patients with small surgical aortic bioprostheses are at a risk of high residual gradients after ViV-TAVI. Considering updated clinical data on long-term outcomes following ViV-TAVI, this approach may become the default treatment strategy for patients with a failing surgical aortic bioprosthesis.

1. Introduction

Surgical implantation of a bioprosthetic aortic valve (aortic valve replacement, AVR) has been the treatment of choice for many patients with aortic valve stenosis (AS) or regurgitation. Current guidelines of the ACC/AHA (American College of Cardiology/American Heart Association) and ESC/EACTS (European Society of Cardiology/European Association for Cardio-Thoracic Surgery) recommend surgical AVR for patients below, respectively, 65 or 75 years of age, who have symptomatic AS or severe asymptomatic AS, provided they have a long life expectancy [1, 2]. However, bioprosthetic aortic valve failure is frequent, with event rates ranging from 5% up to 20% at 10-year follow-up, depending on the valve type [3, 4]. Bioprosthetic valve failure is defined as prosthetic dysfunction which leads to valve-related death, repeat intervention, or severe hemodynamic structural valve degeneration (a mean gradient of 40 mmHg or more or a 20 mmHg increase compared to immediately after implantation) [5].

In the recently updated ESC guidelines (2021), redo cardiac surgery is a class I, level of evidence C indication for symptomatic patients with bioprosthetic valve failure (after excluding thrombosis and endocarditis) and class IIa (C) indication for asymptomatic patients with low surgical risk [1]. However, transcatheter aortic valve-in-valve implantation (ViV-TAVI) has gained much attention in recent years because of the high procedural success rates of more than 90%. ViV-TAVI has been upgraded from a class IIa (C) in the 2017 ESC guidelines to class IIa (B) recommendation in the 2021 ESC guidelines and should be considered based on anatomical characteristics and features of the surgical prosthesis and in patients at high surgical risk. These recommendations are based on registry data and propensity-matched registry studies, showing better short- and long-term outcomes with ViV-TAVI vs. redo surgery [68]. Considering these data and new insights, ViV-TAVI may become the preferred treatment for bioprosthetic valve failure, irrespective of the surgical risk category of the patient.

However, ViV-TAVI requires meticulous planning of the procedure and selection of the approach to minimize the risk of coronary obstruction, device malposition, and high residual gradients [9]. While multiple studies have shown that patients with bioprosthetic valves with small dimensions (21 mm or lower) or with high gradients (patient-prosthesis mismatch, PPM) are at an increased risk of early degeneration, these patients are also at risk for high residual gradients after ViV-TAVI [10].

We here systematically describe the approach to plan and perform a ViV-TAVI (Figure 1), carefully addressing the issues mentioned above, with particular attention to coronary artery protection strategies and bioprosthetic valve fracture (BVF) to avoid high residual gradients in case of small aortic bioprostheses.

2. Surgical Aortic Bioprosthetic Valve Types

When assessing a potential candidate for ViV-TAVI, the first step is to identify the type and size of the surgical bioprosthesis as this may identify procedural risks (e.g., coronary occlusion because of externally mounted leaflets) and influence the approach (e.g., bioprosthetic valve fracture). Surgical bioprostheses can be classified into stented, stentless, or sutureless valves (Table 1). Stented valves may have internally mounted leaflets, resulting in a true inner diameter (ID) that is smaller than the labeled valve size. To maximize effective orifice area (EOA), surgical bioprostheses with externally mounted leaflets (e.g., St. Jude Trifecta and Sorin Mitroflow) and stentless valves have been designed.

In the Valve-in-Valve International Data (VIVID) registry, reporting on the outcomes in 1600 ViV-TAVI procedures, the main challenge in stented valves was reported to be a high residual gradient or PPM [12]. In contrast, stentless valves are more challenging because of the lack of fluoroscopic markers, an increased risk of device malposition (10.3% versus 6.2% in stented valves, ), coronary obstruction (6.0% vs. 1.5%, ), and paravalvular leak (PVL, 11% vs. 4.5%, ). Coronary obstruction following ViV-TAVI is also a major concern in patients with stented valves with externally mounted leaflets [13].

Once the surgical valve has been identified, the “valve-in-valve aortic” app (by UBQO and Dr. Vinayak Bapat) can be used to identify which THV types can be used for ViV-TAVI. The aortic ViV app also provides information on the true internal diameter, fluoroscopic landmarks to correctly position the THV, and the possibility of valve fracturing. Ex vivo images of the transcatheter heart valves (THVs) mounted inside surgical bioprostheses, obtained from the aortic ViV app, are shown in Figure 2 (internally mounted leaflets in Figures 2(a), 2(b), and 2(d) and externally mounted leaflets in Figure 2(c)).

3. Preprocedural Computerized Tomography Planning

Preprocedural computerized tomography (CT) planning is the key when performing ViV-TAVI (Figure 3). The stent diameter of the valve and true ID can be measured and should be in line with the expected values based on the surgical valve type. In case of internally mounted leaflets, the true ID can be smaller than the stent diameter. The upper stent posts of the surgical valve can be marked with three generic markers which can then be used to determine the optimal fluoroscopic views (i.e., three-cusp coplanar view and left/right coronary (LCC/RCC) cusp-overlap).

Next, it is important to identify anatomical factors which may lead to coronary obstruction. When implanting a THV in a surgical bioprosthesis, the leaflets are displaced outward and may occlude the coronary ostia. The implanted THV is not necessarily restricted by the surgical valve, especially not at the level of the coronary ostia since the surgical valve commissures are most often aligned with the native valve [13, 14]. This is also the reason why surgical bioprostheses with externally mounted leaflets or stentless valves have a higher risk of coronary occlusion. The sinuses can even be completely sealed off when the displaced leaflets extend to the sinotubular junction (sinus sequestration), which is more likely in the case of a narrow sinotubular junction or TAVI-in-TAVI [15]. The risk for coronary obstruction can be assessed by measuring the virtual transcatheter valve-to-coronary (VTC) ostium distance and the valve-to-sinotubular junction distance. The VTC distance is measured from the ostium of the coronaries to a virtual cylinder, aligned at the base of the surgical valve, extending up to the coronary ostia and with a diameter equal to the planned THV or its waist at that level (Figure 3). It has been shown that patients with a VTC distance ≤4 mm are at an increased risk of coronary obstruction, and a cutoff of ≤3 mm is considered high risk [13, 14].

4. Choice of the Transcatheter Heart Valve

The US Food and Drug Administration (FDA) has approved the balloon-expandable (BEV) Sapien XT and Sapien 3 (Ultra) valve (Edwards Lifesciences) with intra-annular leaflet position and the self‐expanding (SEV) Evolut platform (Medtronic) with supra-annular leaflet position for ViV-TAVI. Other platforms can be used off-label (Figure 2). Differences in valve stent frame design, leaflet position, and THV expansion/implantation may potentially be beneficial in specific situations.

To prevent high gradients after ViV-TAVI, a THV with supra-annular leaflet position may theoretically be preferred to maximize the EOA. Especially, patients with small annuli, stented bioprostheses, and a small EOA are at risk for high transvalvular gradients after ViV-TAVI. In the VIVID registry, the use of the Sapien device was an independent predictor for increased gradient after ViV-TAVI (mean gradient >20 mmHg, odds ratio: 2.3) [9]. These higher postprocedural gradients with Sapien were especially observed in patients with small annuli (<20 mm) (43% for Sapien versus 24% for the CoreValve/Evolut platform) [16]. No significant difference in postprocedural gradients was reported for patients with larger annuli (23 mm or more), with rates of 21% in both groups. Importantly, at longer-term follow-up, the use of Sapien was associated with higher reintervention rates, caused by higher postprocedural gradients [17].

While a SEV with supra-annular leaflet position may be preferred to maximize the EOA, device malposition is more common when using the SEV as compared to BEV [12]. However, with increasing operator experience and the introduction of repositionable SEV, the rates of device malposition have markedly dropped from 15% in 2012 to 6.5%, and this number can be expected to further decrease [9, 17]. The reported rates of PVL have also been higher for the SEV as compared to BEV, although again, the rate of PVL was reported to be reduced with newer generation devices [12]. No difference in all-cause mortality has been reported for BEV versus SEV in ViV-TAVI procedures [16].

Finally, the need for future coronary access and the need for a possible future re-ViV-TAVI can also guide the THV type selection [18]. A lower stent frame or intra-annular design (Sapien platform) or bigger stent struts are preferred when the possibility of future coronary access needs to be optimized. An intra-annular valve design with low stent height (Sapien platform) may also facilitate future re-ViV-TAVI.

5. Implantation Techniques

In contrast to native valve TAVI, the THV in ViV-TAVI is positioned relative to the fluoroscopic landmarks of the surgical aortic bioprosthesis (usually 2–4 mm below the surgical valve) and not to the annular plane (Figure 4(a)). The optimal implantation depth is also provided by the aortic ViV app. A radiopaque ring at the inflow part of the surgical aortic bioprosthesis may facilitate this, while it is more difficult when there are only radiopaque markers of the surgical valve posts (e.g., Mosaic) or no radiopaque markers at all (e.g., stentless valves). When severe calcification is observed on the preprocedural CT scan, predilatation should be considered to avoid severe underexpansion of a self-expanding THV. However, there is no indication for routine predilatation. Patient-specific commissural alignment can be obtained by aligning the THV with the surgical valve, using the predefined fluoroscopic views and generic markers outlined above [19]. For Evolut, the hat marker should be at the center front during implantation, and the C-tab should be on the inner curve after valve deployment in the R/L cusp-overlap view. No implantation techniques are currently available to provide patient-specific commissural alignment when using Sapien, although this is less critical because of the lower profile of the valve. Standard pacing approaches are used during implantation of the BEV, i.e., 180/min or up to 200–220/min if slower rates do not provide enough pressure decrease. In case of implantation of the SEV in surgical aortic bioprostheses, it is typically best to use intermediate rate pacing, starting at 120–130/min but often transiently increasing the pacing rate to 160/min for a few seconds during valve expansion. Often, postdilatation is performed to optimize the hemodynamic result. The risk of inducing conduction disorders and permanent pacemaker implantation is low (typically <5%) [16]. For Evolut valves, it is recommended to size the postdilatation balloon to the true ID of the surgical valve or 1 mm smaller. For the Sapien platform, the valve is usually oversized minimally 1 mm based on the true ID [20].

6. Bioprosthetic Valve Fracture

Routine postdilatation of the SEV should be considered as high residual transvalvular gradients have consistently been shown to be associated with increased mortality [17, 21]. When residual invasive peak-to-peak gradients of 20 mmHg or more are measured after the ViV-TAVI procedure, bioprosthetic valve fracture (BVF) can be considered (Figures 4(b)4(f)). BVF will be more likely necessary if the immediate postoperative gradients after SAVR were already high (PPM). BVF also reduces pinwheeling, especially in the case of Sapien valves. While this maneuver may be considered to be relatively aggressive, BVF was shown to be safe in a multicenter registry [22]. Whether BVF should be performed just before or after the THV implantation is still a matter of debate among experts; solid real-world data on this topic are still missing.

An indication on which surgical valves can be fractured is shown in Table 1. In general, sutureless and stentless valves cannot be fractured, but they can potentially be overexpanded. To perform BVF, a 50 cc syringe is used, connected with a 3-way stopcock to a noncompliant (NC) high-pressure balloon (True or Atlas Gold balloon (Becton, Dickinson and Company, New Jersey, US)) and an indeflator. It is recommended to fill the 50 cc syringe and indeflator with a 20% contrast-80% saline mixture. After positioning the balloon at the level of the valve, rapid pacing (180–220/min) is initiated. The balloon is quickly inflated with the 50 cc syringe (volume phase, which takes 3–5 seconds) after which the 3-way stopcock is opened to the indeflator to allow pressures up to 14 atm (pressure phase, which can take up to 15–20 seconds). Fracturing of the frame is confirmed visually on fluoroscopy or when a pressure drop is observed on the indeflator (Figures 4(e) and 4(f) and video 1). The indeflator is then released, the balloon is deflated, and the invasive gradients can be measured again.

Typically, the size of the balloon is 1 mm larger than the labeled valve size or 1–3 mm larger than the true ID [11, 23, 24]. Postdilatation balloon sizes have to be carefully chosen in case of the Evolut valve to avoid trauma to the Evolut leaflets. Medtronic recommends that the NC postdilatation balloon does not exceed 1 mm more than the THV waist diameter (i.e., a 20 mm waist in a 23 mm Evolut valve). Intraventricular balloon positioning can be performed if postdilatation with a larger NC balloon is needed. BVF can increase the surgical frame diameter by 2–4 mm. As a consequence, it is also reasonable to use a safety margin when measuring the VTC distance (e.g., 5 mm instead of 3-4 mm).

Achieving a good hemodynamic result after BVF is important since a high residual gradient results in a poor clinical outcome, as illustrated in the two examples in Figure 4. Nevertheless, it can often be difficult to observe geometric changes in the stent frame. Moreover, there can be a discrepancy between invasively measured gradients after BVF and the final result measured noninvasively days to weeks after the implantation [25].

7. Coronary Protection Strategies

Coronary obstruction is a rare but devastating complication of ViV-TAVI, with rates of approximately 3% [13]. Meticulous preprocedural cardiac CT analysis can identify risk factors for coronary obstruction, as previously discussed. In a multivariable model, VTC distance and the use of stented bioprostheses with externally mounted leaflets or stentless bioprostheses were independent risk factors for coronary obstruction [13]. BVF may also increase the risk of coronary obstruction. To mitigate this risk, a THV with a lower stent frame height may sometimes be preferred.

In elderly patients at the risk of coronary obstruction, protection with a coronary guidewire, with or without undeployed stent standby in the coronary artery, is a reasonable strategy. In case of obstruction or anticipated very high risk of obstruction (e.g., Figures 5(a)5(c), bail-out TAVI-in-TAVI in a patient already at a high risk of coronary obstruction), chimney stenting can be performed. In younger patients, in whom the need for future coronary access is important, a preemptive intervention to safeguard coronary access should be considered. In the recently described BASILICA technique (Bioprosthetic Aortic Scallop Intentional Laceration to prevent Iatrogenic Coronary Artery obstruction), the bioprosthetic leaflet is punctured from the side of the sinus, snaring a wire in the left ventricular outflow tract, externalizing it, and then lacerating the leaflet using electrocauterization (Figure 5(d)) [2629]. By lacerating the leaflet and creating a V-shaped surgical valve leaflet, the risk of coronary occlusion by the bioprosthetic leaflet is reduced when it is pushed to the side by the THV. Although there is a substantial learning curve, the safety and feasibility of this technique have been documented in a prospective cohort study [30]. To assess whether the coronary ostium is patent at the end of the ViV-TAVI procedure, an aortogram in an isolated RCC or LCC cusp view can be performed. In case of doubt, IVUS can also be helpful (Figures 5(e) and 5(f)).

8. Cerebral Protection

Ischemic stroke risk after ViV-TAVI has been reported to be approximately 2%, but no information on the use of cerebral protection is available in these reports [12, 17]. As the risk for embolic stroke or debris may theoretically be higher in ViV-TAVI, it may be advisable to use cerebral protection during these procedures, especially when performing BASILICA-assisted ViV-TAVI (Figure 6).

9. Conclusion

ViV-TAVI is a safe and valuable treatment option to treat failed surgical aortic bioprostheses, provided that the procedure is carefully planned and performed. Besides newer THV generations with better implantation results, also newer techniques for coronary protection and cerebral protection are now available to mitigate the major risks of ViV-TAVI. Achieving good hemodynamic results with low gradients after implantation is the key to ascertain good long-term outcome.