Abstract

External cardiac loop recorder (ELR) is a kind of ECG monitoring system that records cardiac activities of a subject continuously for a long time. When the heart palpitations are not the frequent and nonspecific character, it is difficult to diagnose the disease. In such a case, ELR is used for long-term monitoring of heart signal of the patient. But the cost of ELR is very high. Therefore, it is not prominently available in developing countries like India. Since the design of ELR includes the ECG electrodes, instrumentation amplifier, analog to digital converter, and signal processing unit, a comparative review of each part of the ELR is presented in this paper in order to design a cost effective, low power, and compact kind of ELR. This review will also give different choices available for selecting and designing each part of the ELR system. Finally, the review will suggest the better choice for designing a cost effective external cardiac loop recorder that helps to make it available even for rural people in India.

1. Introduction

Norman J. Holter (1914–1983), the famous American biophysicist, introduced a remote cardiac telemetry first time in the 1940s [1]. The Holter system was developed for home ECG monitoring of patients with suspected cardiac arrhythmias. The original Holter monitor had analog patient interface electronics, a 75 lb backpack with a reel-to-reel FM tape recorder, and large batteries. It was the first monitoring system that could record single ECG lead 24–48 hours [2] and analyse ambulatory ECG data outside a standard hospital or outpatient care setting. At present, the Holter monitors are available in the market with the cost of about $369–$2490 [3] depending on their features and the cost of Holter test is around $175–$250 [4] if it is interpreted by a cardiologist. The clinical need to monitor ambulatory ECG has resulted in advances in technology that now allow us to monitor heart rhythms remotely through a wide variety of devices, including ambulatory external monitors and implantable event recorders.

Implantable/insertable loop recorder (ILR) was developed first time by Medtronic’s Reveal [5] (the world’s first implantable diagnostic device). The Reveal ILR detects ECGs during the actual episode, which may allow physicians to take decisions or confirm an abnormal heart rhythm more definitively. Because it could be worn continuously for 14 months, the likelihood of capturing heart rhythm during an infrequent episode was probable. The cost per diagnosis using ILR is around $6,158 [6]. The cause of seizure-like symptoms or related symptoms was diagnosed with the Reveal ILR that may also result in fewer physician and emergency room visits and reduce the number of tests involved when trying to diagnose their cause. Most importantly, diagnosing the cause helps in early treatment effectively. Even though ILR was useful in monitoring of ECG for the detection of abnormal episodes, it had some disadvantages that include the following: (1) a minor surgical procedure is needed, (2) there is always difficulty in differentiating supraventricular from ventricular arrhythmias, (3) under- or oversensing may exhaust the memory of the ILR, and (4) cost of the device is more.

To overcome the limitations of both the Holter and ILR, an intermittent patient- or event-activated recorder was developed. This is also referred to as event monitor or external loop recorder (ELR). The ELR is smaller than Holter in size and is attached to the patient through chest electrodes and records the data when it is activated by the patient or by an automatic trigger that detects irregular heart rates. It is used for monitoring up to 14–30 days. The cost of ELR is $627 and cost per diagnosis using ELR is around $265,9 [7]. The use of ELR avoids the surgical implantation of electrodes. But the activation of the device every time by the patient is difficult unless the autotrigger is used. The autotrigger activates the device as it is programmed which is built into the monitor. Therefore during infrequent symptoms, there is a more chance of missing the activation of the device. This may not give enough information for effective diagnosis. After recording using any of the above-mentioned systems, the data is sent to the central monitoring station where the data is loaded in the computer and analysed. Finally, the reports are sent to the doctor for a final decision or for further tests to detect and confirm the disease. As mentioned above, ELR is providing the noninvasive diagnosis by long-term monitoring. Even the cost of design is less, the ELR test cost is high. Further, the recorded data has to be sent to the specialist to analyse the data which increase the cost further. And there is no option for autosending the recorded data or analysed data to the doctor. The comparison among Holter, ILR, and ELR is given [8] in Table 1.

In Table 2 some of the available ILR and ELR products are given [5, 916].

The mentioned products in Table 2 are very expensive and most of them are not significantly available in India. Distributors are there all over India, but they are providing only a few products like Medtronic SpiderView, SEEQ MCT, Piix NUVANT MCT, GE Healthcare MARS, SEER 1000, SEER Light, Omron HCG801, and BPL cardiac loop recorder monitor. Therefore, a systematic review on internal parts of ECG monitoring system is required to design a cost effective ambulatory ECG monitoring system with an accurate measurement, portable and wearable one as explained in the following sections.

2. Designing of External Cardiac Loop Recorder

The design of external cardiac loop recorder consists of ECG electrodes, instrumentation amplifier, filtering, analog to digital converter, and signal processing unit. PC/laptop/mobile is also used to analyse the data. The major blocks and connection of them are shown in Figure 1.

The ECG signal is acquired from the chest electrodes and is amplified by the instrumentation amplifier. The amplified signal is filtered by the suitable filter to remove the noise. Mostly band pass filter is used for noise removal. Later, an analog to digital converter converts the filtered signal into a digital form which is suitable to process signal by the processor. A signal processing unit is used for processing and feature extraction of the signal to find the normal and abnormal conditions of the patient. For the effective detection of the abnormal conditions during daily activities accelerometer and/or gyroscope is also used along with the chest electrodes. By correlating the signals from chest electrodes and accelerometer/gyroscope, the abnormality of the patient can be defined. The signal processing unit is connected to the PC/laptop/mobile or system on chip (SoC) where the open source software is installed and used for displaying, processing, and saving the data. Further, communication with the doctor can be provided using wireless technology which helps to develop the smart city. The comparative study of each block is explained in following sections.

2.1. ECG Electrodes

Basically, disposal electrodes that may be Ag/AgCl gel type wet sensors or dry sensors are used for acquiring biopotentials from heart. The gel type disposable electrodes have a circular contact. The close electrode placement is allowed by small vinyl backing where necessary and a slightly less firm adhesive allows ouchless removal. The electrodes incorporate liquid electrolyte gel and moderately high chloride salt concentration for quick and accurate readings. These disposable electrodes shown in Figure 2 provide the same signal transmission as reusable electrodes, with added convenience. Each peel and stick electrode is pregelled and designed for one use only. It is very cost effective compared to other electrodes. It is easily attachable to the subject himself/herself and there will not be any assistance needed. The placement of electrodes is also simple and only three electrodes at a time are required for two lead ECG acquisition systems as one electrode is the reference. These electrodes can be used for longer periods depending on the comfort level of the subject.

SKINTACT electrodes [73] shown in Figure 3 are available in the market with three different gels: AQUA-TAC electrode with solid adhesive gel provides 100% contact with skin surface, AQUA-WET electrode with liquid gel provides fast pickup of ECG signal which is preferred for short term monitoring, and AQUA-SET electrode with solid wet gel is used for long-term monitoring.

North Carolina State University researchers [74] had developed a new dry sensor shown in Figure 4 for long-term ECG and EMG monitoring. This device has relied on elastic conductors made from silver nanowires embedded in a pliable polymer.

Imec and Holst Centre introduced the polymer dry electrodes [75] shown in Figure 5 fabricated from ethylene propylene diene monomer (EPDM) rubber which offers a high user comfort and high conductivity.

PDMS (polydimethylsiloxane) based surface electrode shown in Figure 6 was designed [76] for the long-term and unsupervised monitoring. This electrode did not show negative influence on skin even it was worn for one week.

Apart from wet and dry electrodes, there are noncontact electrodes called capacitive electrodes. These were fabricated on silicon with a thermally grown silicon dioxide as the dielectric layer. Dry capacitive electrodes were used for short-term ECG monitoring [77]. A new class of bioelectric sensors was developed by quantum applied science and research (QUASAR) in 2002. These electrodes were capacitively coupled with the body by incorporating the sensors into shirts, elastic belts, and glasses. The QUASAR two-generation electrodes are shown in Figure 7(a). The first-generation electrode IBEv1 is a larger, square sensor (1′′  ×  1′′) used to measure bioelectric potentials through T-shirt [78]. The second-generation electrode IBEv2 was developed as a small circular shape sensor shown in Figure 7(b).

2.2. Accelerometers and Gyroscopes

Accelerometers and gyroscopes are also used along with dry or wet sensors for cancelling muscle contraction interferences, to measure heart rate under different activities like stress, movements, and so forth. The accelerometer is a 3-axis one. It is used in tilt-sensing applications, as well as dynamic acceleration resulting from motion or shock to measure the static acceleration of gravity. In previous work done the people used the accelerometer for different purposes. In previous work done, the people used the accelerometer for different purposes. ADXL335 triaxial accelerometer [9, 79] and triaxial accelerometer MotionPodTM by MOVEA were used for removal of motion artefacts. SDI1221, a low cost, integrated 1-axis accelerometer, was used in zero to medium frequency instrumentation applications to provide extremely low noise (5 μg/√Hz) [80]. A triple axis accelerometer [8183] and MMA7260QT [84] were used in telehealth monitoring. ADXL330 was used in deciding of the cardiac disease [85, 86]. Bosch BMA180 accelerometer was used in human behaviour tracing [87]. A triple axis accelerometer [81, 83, 8890], ADXL345 [91], and ADXL330 [92] were used in activity recognition. MC301 made by Wacoh was used in ambulatory monitoring to find human posture and walking velocity [93]. MMA8451Q (Austin, TX, USA), a triple axis, low power, capacitive digital accelerometer (freescale semiconductor) [94], a triaxial accelerometer (patch sensor device designed by Vital Connect, Inc. (Campbell, CA)) [95], and inbuilt on-board 3-axis accelerometer SCA3000 [96] were used in extraction of respiratory rate. And also a triaxial accelerometer was used to measure the body movements [90] or daily stress [97] and for left ventricular functions monitoring [98]. A triaxial gait accelerometer MMA7260Q (freescale semiconductor, Austin, TX, USA) [99], piezoelectric foils [100], and Pegasus activity monitors developed by ETB, UK, were used for time-frequency analysis of heart rate. Triaxial accelerometer ADXL335 [9, 79] and MotionPodTM by MOVEA [98] were used as the reference for removing motion artefact by adaptive filtering algorithm (LMS or ANC) in acquiring of ECG during treadmill exercise. Apart from these, a triaxial accelerometer (LlS344ALH, ST Microelectronics) was used for seismocardiography.

Among all the accelerometers mentioned in Table 3, the model ADXL345 shown in Figure 8 can be selected because of less power consumption and better full scale range with 2–3.6 V supply voltage. In ECG monitoring, the accelerometer is used to get the change in acceleration due to body movements during daily activities. This is helpful in detecting the arrhythmias. Finally, the heart rate measured by disposal electrodes and the accelerometer readings will be correlated. Using this information alerts or notifications are sent.

The gyroscope is used to find the tilt in position when there is motion in the body. This is required for monitoring of ECG during daily activities. In previous work, gyroscopes were used in different applications like L3G4200D gyroscope used for head movement tracking along with accelerometer and magnetometer [17]. Gyroscope and accelerometer inbuilt MEMS chip [101] were used in robotic arm control by detecting the motion of arm [102] and vehicle speed control [103]. Ring laser gyroscope [104] and microgyroscope [105] are advanced gyros used for various applications. In cardiac applications, gyroscope was used for monitoring electric and mechanical functioning of heart [106] (gyro developed by Zimpher Technology and Shimmer Research was used in [107]) means, heart rate [108], rotational velocity of foot [108], emotional eating (2-axis gyro was used), human posture and walking velocity (ENC03J developed by Murata Manufacturing Co. Ltd., Kyoto, Japan, was used in [18]), stride strength and walking velocity (ENV05S developed by Murata Manufacturing Co. Ltd., Kyoto, Japan, was used), muscle contractions (vibrating disc piezoelectric gyroscope was used in [19]), and motion processing in handsets (InvenSense MPU-3000 3-axis MEMS gyroscope was used).

The differences between gyroscope and accelerometer are given in Table 5 that help in the selection of gyroscope or accelerometer or both for ECG monitoring systems.

In order to differentiate the ECG signal due to heart activity from the patient’s daily life activities, accelerometer and gyroscope alone are not sufficient. Therefore, it is suggested to use both accelerometer and gyroscope to find daily activities of patient effectively.

2.3. Placement of Electrodes

The placement of electrodes on the body varies based on type of wearable design. For different wearable types, placement of electrodes according to the previously proposed designs is given in Table 6.

2.4. Instrumentation Amplifier (IA)
2.4.1. Mostly Used IA ICs

There are a number of instrumentation amplifier ICs available in the market suitable for ECG signal amplification. The use of IA in IC form is very easy and more convenient in ECG signal acquisition because of its small size and high noise immunity. The most widely used IA ICs were developed by Texas Instruments and Analog Devices. Texas Instruments ICs INA116 [109, 110], INA121 [111], and INA128 [112] were most widely used in ECG signal acquisition systems. INA116 provided high input impedance (1015 Ω) and the bandwidth of 0.38–44 Hz (±5%) with a single supply of 2 V; it was used for long time ECG monitoring of athletes [109]. It was also used in the designing of low noise EEG/ECG sensor circuit [110]. INA121 with a two-input voltage buffer as driving Right Leg (RL) circuit provided differential gain = 1000 from 0.05 Hz–100 Hz and common-mode gain = 0.06 at power-line frequency (50 Hz) that results in CMRR = 86 dB [111]. Analog devices ICs AD620 [113] and AD623 [114] were used for ECG signal acquisition and monitoring.

2.4.2. Circuit Designs of IA

Basically the instrumentation amplifier is designed using operational amplifier which acts as voltage amplifier [115] that provided gain = 54.83 dB, CMRR = 141.61 dB, and bandwidth = 223 Hz [55]. A simple unity-gain buffer stage and differential amplifier stage with high input impedance [116] were used to design IA to have optimised low-frequency response, low power, and CMRR. The minimum input resistance of the amplifier required was obtained as 1.3 MΩ [117]. A composite stabilised amplifier with active current feedback at its input stage was used to reduce amplifier saturation problems and baseline drift [118] in off-the-shelf ECG amplifier for a continuous long duration. But the amplitude is not matched with that of standard (3 electrodes) voltage ECG amplifier. If RE < 50 kΩ, the bandwidth of the circuit will decrease below the bandwidth of the acceptable limit. DDA (differential difference amplifier) was used to lower the power consumption and keep the open loop gain to enough value. The AC coupled technique was used to reduce offset noise. DDA with AC coupled technique [61] provided power supply rejection ratio = 62 dB & CMRR = 150 dB at 10 Hz and with the preferred input noise at 5 μV/Hz power consumption = 3.99 μW at 1 Hz. To remove offset voltage and reduce noise, the low-frequency signal was to be eliminated. This was done by differential AC coupling network and the HP difference amplifier [119]. A design for remote electrocardiogram system, which consists of five stages ECG input, isolated amplifier, main amplifier, active BRF, and high order LPF with bandwidth 1 Hz–200 Hz [120], was used for ECG signal amplification and power supply (60 Hz) noise reduction.

Two-stage IA using operational transconductance amplifier (OTA) and common-mode feedback amplifier topology was used for common-mode amplifier noise reduction. This provided power consumption = 1.47 μW and CMRR = 82 dB [56]. An IA with series combination of two OTAs (one is preamplifier and second is variable-gain amplifier) provided power consumption = 233 nW, bandwidth = 21 Hz, gain = 44.2 dB, and CMRR = 80 dB [60]. Flicker noise was removed by both chopper stabilised front end amplifier [121] and chopped capacitively coupled IA (CCIA) [122, 123]. Chopper technique which was implemented using folded cascode structure provided 36.44 dB of SNR in [54].

The instrumentation amplifier using the opamp for ECG signal acquisition cannot reduce noise much effectively. Therefore in order to solve this problem ECG amplifiers were designed using CMOS technology [5663, 124127] which also provide less power consumption and small area. The noise reduction in terms of CMRR obtained in different papers is mentioned in Table 7.

From Table 7, one can observe that most of the work reported was based on usage of same processing technology with different battery voltage. The work done in [62] was given better common-mode rejection ratio with a Monolithic CMOS current-mode instrumentation amplifier.

2.5. Filter

Filtering was required to remove the noise in ECG signal acquisition from electrodes followed by IA. The noise interferences were involved in many ways in ECG acquisition as its amplitude is less (in the order of mV) and variability of ECG segments durations. Muscle contractions, electrode movements during acquisition, base line wandering, and 60 Hz power supply noise were some of the significant noise interferences. And also, filtering was required to separate the segment of interest from the acquired ECG signal like P wave, R-peak, QRS complex, T wave, and ST segment. Here removal of noise interference was not considered in this paper. Different filters and their frequency range for different parameters acquisition used by previously proposed authors are given in Table 8 for selecting and deigning of required filter.

From Table 8, it is shown that the most of the authors used LPF and HPF or BPF for measuring almost any parameter. But the frequency range is not the same for all. It is different for different parameters. Therefore, the designer has to select the frequency range based on his/her segment of interest.

2.6. ADC

The ADC ICs such as 16-bit, 100-kSPS SAR ADC ADS83212 [33], 10-bit SAR ADC [30, 38, 128], and 24-bit ADS1292 [129] were used for analog to digital conversion of signal. But nowadays the signal processing development boards like Texas products ADS1298, ADS1191, ADS1192, ADS1194, ADS1196, ADS1198, ADS1291, ADS1299, ADS1298R, ADS1296R, ADS1296, ADS1294R, ADS1294, ADS1293, and ADS1291 that provide analog voltage 2.7 V–5.25 V and digital voltage 1.65 V to 3.6 V [130] and analog devices ADAS1000 (low power, 5-electrode ECG analog front end) and AD8232 (single-lead heart rate monitor analog front end) [131] are available with ADC inbuilt at significantly reduced size, power, and overall cost. Therefore, there is no need for external ADC to place.

2.7. Signal Processing Unit

Generally microcontroller board is used as signal processing unit to process the digital signal. This unit is further connected to PC/laptop to display the signals and measurements. It can also be used to communicate with other systems using transmitter and receiver. In previous designs proposed, for short term monitoring of ECG for 10 sec or 1-2 minutes MSP430 microcontroller was used [34, 71] and for long-term monitoring TI CC2530 system [37], CC2431 [132], DSP [128], DSP chip TMS320VC5509A [133], TMS320F2812 [134], TMDX5505eZDsp/VC5505eZdsp [33], MSP430 (monitoring for 45 days) [68], MSP430F5515 [129], MSP430F1232 [43], MSP430FG439 [135], MSP430F2418 [136], MSP430F5529 [66] (monitoring for 88 h) [39], MSP430F5419A (monitoring for 48 h) [137], ATmega8 [41], ATmega328 [42], Arduino UNO (ATmega328) [47], ATmega8L [28, 31, 32], Concerto MCU [65], Revitive Device [27], PIC18LF4620 [69], Altera EP2C35 Nios II soft-core CPU based FPGA [138], ARM9 [139], ADuC842 [140], C8051F021 [141], 32-bit ARM Cortex M0 CPU (monitoring for 24 h) [30], and STM32 chip as the system controller with ARM Cortex-M3 core (monitoring for 44 h) [67] were used.

2.8. Communication to PC/Laptop/Mobile Phone
2.8.1. Need of Communication to PC/Laptop/Mobile Phone/Soc Network

After acquiring ECG signal, to display process and report the results of analysis to physician or doctor for diagnosis of the disease, there is a need for connecting to PC or Laptop. Mobile also can be used with specially designed apps.

2.8.2. Available Communication Techniques

(1)USB-SPI is generally used to connect the MC development board to PC. To display the signals and measurements MATLAB Simulink GUI or specially designed GUI is used.(2)Bluetooth is used for connecting to PC or mobile phone. To display the signals and measurements specially designed GUI is used in PC and for mobile phone (and also tablet) an android app is used. Motorola cell phone is providing an app developed with Java 2 Micro edition (J2ME).(3)IEEE 80215.4/ZigBee is used for PC connection. The signals can be displayed by using LabView or MATLAB GUI.(4)WiFi is used for connecting to PC or mobile phone. With a specially designed GUI developed in Java, the signals can be displayed on PC and mobile phone. Open source software (the app is written in X-code using object C) developed by E.P. Ltd. is available in Apple’s iPhone 4S smartphone.(5)GSM/GPRS/GPS is used for long distance monitoring of ECG by connecting with PC or mobile phone.(6)Some of the system on chip (SoC) products like AT86RF212B, AT86RF233, AT86RF215, AT86RF215M, and AT86RF215IQ [142] will provide wireless communication network through ZigBee technology. TMS37157, TRF796X, TRF7970, AT86RF212, MCRF200, ADF7021, and ADF7025 [143] will provide communication using RFID technology. CC1101, CC1110, CC430, CC1190, CC11XL, CC112X, and CC120X [143] will provide communication using WPAN technology. CC2520, CC2530, CC2530ZNP, CC2531, CC2533, ADF4242, and AT86RF231 [143] will provide communication using ZigBee technology. CC2560, CC2540, CC2570, EM250, EM260, BCM4329, and BCM2045 [143] will provide communication using Bluetooth technology. WL1271, WL1281, BCM43241, BCM25/29, BCM4318, BCM4330, BCM4752, and AR6102 [143] will provide communication using WiFi technology. And WL1281, NL5500, UBX-G6010, BCM4750, and SiR starV [143] will provide communication using GPS technology.

2.8.3. Selection of Effective Communication Technique

The communication mode is selected based on the distance of monitoring the signals. For short distance wired connection like USB SPI or wireless communication (1 or up to 100 m, depending on radio class) through Bluetooth or ZigBee technology (up to 75 m) or WiFi (indoors about 150 feet (46 m) and outdoors about 300 feet (92 m)) is generally preferred. For long distances GSM/GPRS (35 kilometres) or GPS (up to 25000 Km) is preferable. At present, all the communication technologies are inbuilt in the hardware and available as SoC (some of available SoC products are given in Section 2.8.2). When the SoC is selected for signal processing, it is better to select the suitable SoC product which is having preferred communication network technology. By providing long distance communication with the doctor, there is a scope for online monitoring of the patient condition and online diagnosis. This will not only save money and time, it will save lives of poor people. And also it helps to develop a smart city in the area of medical engineering.

3. Discussion

The death rate is increasing every year due to heart diseases from past few decades in India. This can be reduced by early detection of symptoms of abnormalities. A few years back, the ECG systems for detecting abnormalities were only available in the hospital and used only in the presence of specialists. It was very difficult to go every time to the hospital and take the ECG, which was also very expensive, especially for rural people. But present situation is slowly changing by using the health monitoring systems. Therefore, everything is going to change within few years in India like developed countries in the field of biomedicine by developing the smart and wearable health monitoring systems. So much of work is done by many people, but there is a lack of validation and communication provision with the doctor. There are options for recording and sending the data to the service centre where the data is analysed. But there is no accessibility of data to the user or patient. And also, they used commercial software which is licensed and very costly. Therefore, the net cost is very high.

In order to overcome these limitations and to add the missing features in existing systems, a new framework is proposed in this paper. In this review, quantitative information for designing of external cardiac loop recorder (ELR) is presented as a study of real-time ECG monitoring from remote area continuously. With the proper selection of the devices such as electrodes/sensors, instrumentation amplifier, filters, processor, and communication mode, an advanced external cardiac loop recorder is going to be designed to achieve better performance with less cost. New framework includes an option for saving the recorded ASCII data in text or excel form, and then it is easy to access and process the data. Further, the data can be processed and extract the features for detecting the normal or abnormal condition of the patient by using open source software called Scilab that reduces the cost of the system by avoiding commercial software used for analysis. And also by using open source software like TeraTerm, CoolTerm, and Processing with Arduino, data can be sent to a doctor via Bluetooth or Wi-Fi. Using Gobetwino open source software with Arduino data can be sent via the internet. Therefore, the doctor can receive and analyse the data using open source software and further he can send the suggestions or precautions to the patient at an early stage. If such a system is designed, it would become a milestone in the field of biomedical engineering and would help to develop the smart city towards the biomedical field in India. And also it will reach the rural people effectively so that the death rate due to heart diseases can be reduced.

It is evident from Tables 19 that one can design and configure appropriate internal circuitry components for the development of the cost effective external cardiac loop recorder system. The appropriate selection of open source software along with suitable internal circuitry will give way for new ELR suitable for implementation with less cost. Under a pilot process, a working prototype is under investigation by duly considering all the design parameters and software requirements. This expected design system will ensure the required diagnostic precision suitable for detecting the cardiac episodes.

4. Conclusion

This research study provided an insight into the systematic review on external cardiac loop recorders. It gives the quantitative information which helps in the selection of internal parts of the external cardiac loop recorder. Although several techniques for monitoring cardiac episodes were available, the scope for a new cardiac device is still in demand. This is due to the fact that the real-time cardiac episodes monitoring and its corresponding alert mechanism can help in saving the life of the patient. Such mechanism through the advent of cost effecting wearable external cardiac loop recorder will provide a major healthcare revolution in the developing countries.