Abstract

In the present paper, graphene oxide was directly electrodeposited by means of cyclic voltammetric techniques on the glassy-carbon electrode (GCE) to obtain a reduced graphene-oxide-modified electrode (ErGO/GCE). Cyclic voltammetry (CV) and differential pulse anodic stripping voltammetry (DP-ASV) had been utilized to study the electrochemical behavior of ErGO/GCE toward ascorbic acid (AA), paracetamol (PA), and caffeine (CA). Differential pulse voltammetry results show that AA, PA, and CA could be detected selectively and sensitively on ErGO/GCE with peak-to-peak separation of 312 mV and 756 mV for AA–PA and PA–CA, respectively. The factors affecting the voltammetric signals such as pH, scan rate, and interferents were addressed. The results reveal that the ErGO/GCE-modified electrode exhibits excellent electrochemical activity in the oxidation of PA, CA, and AA. The detection limits are 0.36 μM, 0.25 μM, and 0.23 μM for AA, PA and CA, respectively, suggesting that the ErGO/GCE can be utilized with high sensitivity and selectivity for the simultaneous determination of these compounds. Finally, the proposed method was successfully used to determine AA, PA, and CA in pharmaceutical preparations.

1. Introduction

Ascorbic acid (AA), a vitamin commonly present in many biological systems and in multivitamin formulations, is widely employed to provide an adequate dietary intake and as an antioxidant [1, 2]. Its overdoses may cause gastrointestinal discomfort, headache, trouble sleeping, and flushing of the skin [3]. These drugs are often combined in two or three types to increase the possibility of therapy. Paracetamol or acetaminophen (N-acetyl-p-aminophenol or paracetamol (PA)) is an antipyretic and analgesic drug commonly used against arthritis, muscle aches, and fevers [4]. A high amount of PA can cause the accumulation of toxic metabolites, leading to severe and sometime fatal hepatotoxicity and nephrotoxicity [5]. Caffeine (1,3,7-trimethylxanthine (CA)) is an alkaloid component present in natural products such as coca cola, coffee, and tea leaves [6]. It is known to have many pharmacological effects including as a diuretic, cardiac stimulant, and stimulant of the central nervous system [6, 7]. Caffeine is employed in combination with ergotamine in migraine treatment [8]. However, large doses of caffeine can cause nausea, abdominal pain, nervousness and seizures [9], and mutation effects such as inhibition of DNA [10]. Therefore, it is very important to determine the amount of these compounds sensitively, simultaneously, accurately, and fast.

Several analytical methods for individual or simultaneous determination of AA, PA, and CA have been reported in the literature such as spectrofluorometry [11, 12], chromatography [13, 14], spectrophotometry [15, 16], capillary zone electrophoresis [17, 18], and electrochemistry [1922], among which electrochemical methods based on modified electrodes have fascinated many researchers due to their simplicity, high sensitivity, and low cost. However, one major obstacle is that the oxidation peaks of AA and PA are almost overlapping on traditional electrodes, which make their simultaneous determination highly difficult [23, 24]. For this reason, it is necessary to prepare modified electrodes that can be employed for simultaneous detection of these compounds.

Graphene consists of a flat monolayer of sp2-carbon atoms bonded and arranged in a honeycomb lattice. It has attracted extensive interest because of its unique properties including its ideal two-dimensional structure [25], electronic properties [26], mechanical properties [27], and thermal properties [28]. Furthermore, graphene oxide (GO) and reduced graphene oxide (rGO) have also been employed in modified electrodes. rGO has better conductivity than GO and restores the unique properties found in the pristine graphene. Some reports are related to the application of rGO-based materials in electrochemistry [2729]. Various methods including chemical reduction [30, 31], thermal reduction [32, 33], photocatalytic reduction [34, 35], and electrochemical reduction [36] have been used to prepare rGO. The electrochemical reduction of GO is a relatively simple, economical, fast, and green method. This method could restore some of the original properties of pristine graphene and exploit new functionalities of rGO, along with other nanoparticles or compounds [36]. rGO produced using this method is known as electrochemically reduced graphene oxide (ErGO). To the best of our knowledge, the simultaneous voltammetric determination of AA, PA, and CA using an ErGO-modified electrode has not been reported.

In the present paper, an electrode modified with electrochemically reduced graphene oxide is demonstrated. The obtained electrode was used for the simultaneous determination of AA, PA, and CA by means of the DP-ASV method.

2. Experimental

2.1. Materials

Graphite powder, potassium permanganate (KMnO4), sodium acetate (NaCH3COO), sodium citrate (Na3C6H5O7), sodium dihydrogen phosphate (NaH2PO4), sodium hydrogen phosphate (Na2HPO4), acetic acid (CH3COOH), ascorbic acid (C6H8O6), citric acid (C6H8O7), boric acid (H3BO3), ammonia solution (NH4OH, 25%), paracetamol (C8H9NO2, 99%), caffeine (C18H10N4O2, 98.5%), and ascorbic acid (C8H6O6, 99.7%) were supplied from Merck company (Germany). Sodium nitrate (NaNO3), ethanol (C2H5OH), hydroperoxide (H2O2, 30%), and potassium hydroxide (KOH) were supplied by Daejung company (Korea). Phosphate-buffered solution (PBS) with pH 6 was prepared from 1 M NaH2PO4 and 1 M Na2HPO4. Britton–Robinson buffer solutions (B–RBS) in the range of pH from 2 to 6 were prepared from 0.5 M H3BO3, 0.5 M H3PO4, and 0.5 M CH3COOH. The desired pH 6 buffer was adjusted using 1 M KOH or 1 M H3PO4 solutions.

2.2. Methods

X-ray diffraction measurements were performed on a D8 Advance (Bruker, Germany) with CuKα radiation (). Fourier transformation infrared analyses were recorded on a Shimadzu IRPrestige-21 (Japan). Raman spectra were obtained using XploRA, HORIBA (Japan), 532 nm YAG Laser. A CPA-HH5 Computerized Polarography Analyzer (Vietnam) was used for voltammetry measurements. All measurements were performed in a cell with three electrodes: a glassy-carbon electrode (GCE) with a diameter of 2.8 ± 0.1 mm used for formatting the modified electrode as the working electrode, an Ag/AgCl/3 M KCl as a reference electrode, and a platinum wire as an auxiliary electrode. All measurements were performed at ambient temperature.

The HPLC (high-performance liquid chromatography) method was carried out to analyse AA, PA, and CA for the sake of comparison. Chromatographic determinations were performed in a Shimadzu 2030 HPLC system. The chromatographic conditions are as follows: UV detector (), flow rate set at 2.0 mL·min−1 with 10 μL as the injection volume, and column temperature set at 45°C ± 1°C. All the calculations concerning the quantitative analysis were performed with external standardization by measuring the peak areas of the chromatograms.

2.3. Preparation of ErGO/GCE

Graphene oxide was prepared using the modified Hummers process [36, 37]. Reduced graphene oxide was directly prepared on GCE by means of electrochemical reduction. Graphene oxide was applied by cyclic voltammetry with a potential range from 0 V to −1.5 V in the 0.2 M phosphate buffer of pH 7.

In this study, the effect of the amount of GO (z1), number of cycles (z2), and scanning rate (z3) on the peak current of ErGO/GCE was studied using the Box–Behnken design [3840]. The number of experiments (N) required for the development of this design is defined as , where is the number of factors () and is the number of replicates at the center point () [41]. Thus, a total of 15 runs were carried out for optimizing these three variables. Each independent variable was considered at three levels: low, medium, and high, coded as −1, 0, and +1, respectively (Table 1). The center points were used for the determination of error. Based on the experimental data, a second-order polynomial model was obtained, which correlates the relationship between the response and the studied variables. The relationship could be expressed in where is the predicted response value (peak current ()); , , and are independent variables; is the intercept term; , , and are linear coefficients; , , and are cross-product coefficients; and , , and are quadratic-term coefficients. All the coefficients of the second polynomial model and the response obtained from the experimental design were subjected to multiple nonlinear regression analyses. The empirical relationships between the response () and the tested variables were obtained using Minitab version 16.

2.4. Pharmaceutical Preparation

Panadol Extra (Sanofi-Synthelabo company, Vietnam), Hapacol Extra (DHG Pharma company, Vietnam), Tatanol (Pymepharco company, Vietnam), Effe-Paracetamol (Pymepharco company, Vietnam), Ameflu C (OPV company, Vietnam), and Efferalgan Vitamin C (Laboratoires UPSA company, France) were employed in this study. Ten tablets of each analysed pharmaceutical preparation were weighed and finely ground in a mortar. An exact amount of the powder was weighed and transferred to a 100 mL calibrated flask, which was filled to the volume with the 0.2 M B–R buffer solution of pH 3.2. The standard addition method was used for determining the pharmaceutical formulations.

3. Result and Discussion

3.1. The Preparation of ErGO/GCE Based on the Box–Behnken Design

In the present study, the optimum levels of the variables including the amount of GO (z1), number of cycles (z2), and scan rate (z3) were studied with the Box–Behnken design.

Based on the experimental data (Table S1, Table S2, and Table S3), three equations of the experimental design were established using Minitab 16 ((2), (3), and (4)). They correlate the relationship between the peak current of AA, PA, and CA and the three variables.

The coefficients corresponding to each factor in the linear and quadratic terms and their interactions in (2), (3), and (4) have the same sign. A minus sign suggests an antagonistic effect, whereas a plus sign indicates a synergistic effect. The absolute value of the coefficients determines the significance of the effect. Thus, the signs of the coefficients and their values indicate that those amounts of GO, number of cycles, and scan rate act in the same way on the peak current for AA, PA, and CA oxidation, but the level of importance is different. As seen in the equations, the amount of GO () and number of cycles () have a positive effect. This indicates that the intensity of the peak current can be improved when the value of these factors increase, whereas the increasing scan rate () reduces the peak current because the sign of is minus.

Since the peak current for AA oxidation is the lowest (Figure 1(b)), it was chosen for the optimization of the variables in the preparation of ErGO/GCE.

The obtained results through ANOVA (Table S4) show that most of the factors (, , , , , , , and ) have the value of less than 0.05, indicating that they are significant (this means that the amount of GO, number of cycles, scan rate, the interactions between the amount of GO and the number of cycles, and the amount of GO and scan rate significantly affect the peak current of AA). Only has the value of more than 0.05, and it will be removed from (2). Then, the reduced form of (2) is as follows:

The analysis of variance is used to estimate the statistical significance of the model. The summary of the analysis of variance is shown in Table 2. If the value is lower than 0.05, it means that the model is statistically significant. The fitness of the model equation was estimated according to the determination coefficient (). The results show that the value of is 0.988. This means that the calculated model is able to explain 98.8% of the results (or of the variability of the response). This result also indicates a good agreement between the experimental and predicted values of the response. The value for the “lack of fit” test is 0.856. This finding suggests that the model employed to fit the response variables is significant and adequate to describe the relationship between the response and the independent variables.

The profile for the predicted values and desirability option in the Minitab 16 is applied for the optimization process. Figure 1(a) presents the optimization plots of AA. The optimization shows the maximum (2.49) obtained under the following conditions: amount of GO being 7.0 μg, number of cycles being 10, and scan rate being 0.04 V·s−1. The reliability of this prediction was checked by carrying out three replicates under these optimal conditions. The experimental for AA are 2.47, 2.47, and 2.49 (Figure 1(b)). The comparison by means of one-sample t-test shows that there is no significant difference in the respective value presented by the model (; ). This demonstrates that the optimization is acceptable. These conditions were selected for the next experiments.

Figure 2(a) illustrates the electrochemical reduction of the GO by means of the CV technique. GO was drop cast on GCE under the optimal conditions (7.0 μg, 10 cycles, and 0.04 mV·s−1). In aqueous solution, the film was reduced by potential cycling between 0 V and −1.5 V at pH 7. As seen in Figure 2(a), the first cycle shows a large cathodic peak current starting at around −1.23 V, and the reduction current observed could be related to the reduction of oxygen functional groups on GO. During the second cycle, the reduction current decreases significantly and vanishes after the third cycle. This observation is consistent with that of the previous papers [42, 43]. XRD measurements were used to study the phase of the obtained samples. Figure 2(b) shows the XRD patterns of graphite, GO, and ErGO. The characteristic diffraction peak of GO at 2θ of around 11.3° indicates that oxygen-containing functional groups on graphite sheets are formed, that is, the formation of GO [44, 45]. Peaks at around 26° could be attributed to the initial graphite indicating that the oxidation of graphite is incomplete. After being electrochemically reduced, the material does not show the peak at 11.3°, and the weak and broad reflection peak at 25.8° corresponds to the relative short-range-order structures in disordered stacked rGO [44], which indicates the successful reduction of GO.

The existence of oxygen-containing functional groups was confirmed by FT–IR (Figure 2(c)). The strong absorption band at 3417 cm−1 is attributed to as the graphite was oxidized. The vibration bands at 1720 cm−1 and 1620 cm−1 can be responsible for of carbonyl and of aromatic rings, respectively. The absorption bands at 1051 cm−1 can be assigned to of alkoxy [36, 46]. The vibrational band at 1395 cm−1 can belong to . The presence of the oxygen-containing functional groups again indicates the successful oxidation of graphite to graphite oxide.

Raman spectroscopy is a useful technique for the characterization of the physicochemical properties of graphene. In general Raman spectroscopy, the G band contributes to the first-order scattering of the E2g phonon of sp2 C atoms (∼1570 cm−1) and the D band is attributed to a breathing mode of k-point photons of A1g symmetry (∼1350 cm−1). At the same time, an increased / ratio arises from a lower degree of crystallinity in the graphitic materials. Two bands of the Raman spectra of GO and ErGO were observed at approximately 1352 and 1570 cm−1, contributing to the D band and G band, respectively (Figure 2(d)). The intensity ratio of the D and G bands increased from 0.90 to 1.14 due to the electrochemical reduction, suggesting a decrease in the size of the in-plane sp2 domains and the partially ordered crystal structure of graphene. These results provided more evidence for removing oxygen-containing functional groups from the surface of GO by means of electrochemical reduction [36, 46].

3.2. Electrochemical Behavior of ErGO-Based GCE

The CV experiments were conducted on bare GCE and ErGO/GCE by using 1.0 mM K4Fe (CN)6 in 0.1 M KCl as a standard probe [47]. For the reversible process, the Randles–Sevcik formula in a simplified form (6) was employed to calculate the effective surface area of the GCE and modified GCE. where refers to the anodic peak current, is the number of electrons transferred, is the effective surface area, is the diffusion coefficient, is the concentration of K4Fe (CN)6, and is the scan rate.

For 1.0 mM K4Fe (CN)6 in 0.1 M KCl electrolyte, is equal to 1 and is 7.6 × 10−6 cm2·s−1. Then, the effective surface area calculated from the slope of the plot of versus is 0.043 cm2 and 0.050 cm2 for GCE and ErGO/GCE, respectively (see Figure S1). This means that ErGO/GCE provided a larger effective surface than the bare GCE did.

The values of the heterogeneous electron transfer rate constant, , were determined by means of the Nicholson method combined with the following equation: , where is the kinetic parameter, is the diffusion coefficient, is the number of electrons involved in the process (), is the scan rate, is the Faraday constant (), is the universal gas constant (), and is the temperature () [47, 48]. The kinetic parameter, , is calculated as a function of (peak-to-peak separation) at the set temperature (298 K) for a one-step, one-electron process with the transfer coefficient, , equal to 0.5. The function of (), which fits Nicholson’s data for a practical usage, is given by where is used to determine as a function of from the experimentally recorded cyclic voltammetry. The values of were obtained from the slope of the plot of against . The heterogeneous electron transfer rate constants for 1.0 mM K4Fe (CN)6 in 0.1 M KCl electrolyte were 16 × 10−6 for GCE and 3.7 × 10−6 cm·s−1 for ErGO/GRE (see Figure S2). The for ErGO/GGE is lower than that for the bare GGE. This may be due to the electrode modification which increases the resistance of electrode.

The CV experiments were also performed on a modified GCE with and without ErGO in order to confirm its electrochemical behavior for the detection of AA, PA, and CA (Figure 3). As can be seen in the figure, the current response on bare GCE and GO/GCE exhibits broad peaks, while the current response on ErGO/GCE provides defined peaks. The intensity of the anodic peak current (henceforth ) of AA, PA, and CA on ErGO/GCE was 59.17-, 72.78-, and 12.49-fold, respectively, compared with that on the bare GCE. The peaks of AA and PA are often overlapped, leading to difficulties in their simultaneous determination. It is worth noting that the peak-to-peak separation of 312 mV and 756 mV for AA–PA and CA–AA, respectively, was obtained on the ErGO, while it is not observed on bare GO or ErGO/GCE in this study. The favorable signal-promoting effect of the ErGO suggests that it could accelerate the electron transfer of the analytes and had a good electrocatalytic activity for the redox reaction of AA, PA, and CA. ErGO contains carboxyl groups and aromatic rings. The π–π stacking interaction between the aromatic rings of the analytes and ErGO is possibly responsible for the accumulation of AA, PA, and CA on the modified electrode. The process is illustrated in Scheme 1.

3.3. Effect of the pH

The effect of the pH on the voltammetric signals of AA, PA, and CA was studied in the pH range of 2.3 to 5.8 using the B–RBS (Figure 4(a)) with the CV method in order to estimate the relevant pH and also the ratio of electrons and protons participating in the voltammetric process on ErGO/GCE. As can be seen from Figure 4, the highest values of for CA are at pH 3.2, while for AA and PA decreases slightly as pH increases from 2.3 to 5.8. Therefore, , which is suitable for the analytes, was chosen for subsequent analytical experiments.

It was found that the oxidation peak potential () is dependent on pH. The potentials for AA, PA, and CA shift towards less positive values as the pH gradually increases, suggesting the involvement of protons in the oxidation reaction. Within the pH range 2.3–5.8, the linear regression equations for the oxidation peak potentials (, , and ) and pH are expressed as follows:

The values of 0.054 V/pH for AA, 0.061 V/pH for PA, and 0.049 V/pH for CA are close to the Nernstian slope of 0.059 V/pH, indicating that the oxidation of AA, PA, and CA involves an equal number of electrons and protons.

The mechanisms of PA [49] and AA [5] oxidation with two electrons and two protons are illustrated in Schemes 2(a) and 2(b). The CA oxidation on the electrode is believed to take place in two steps [50]. The first low step involves the oxidation of the C–8 to N–9 bonds and produces 1,3,7-trimethyluric acid with the involvement of two electrons and two protons. The second fast step is followed by 2e, 2H+ oxidation to 4,5-dihydroxy-1,3,7-trimethyltetrahydro-1-H-purine-2,6,8-trione and 4,5-dihydroxy-1,7,9-trimethyltetrahydro-1-H-purine-2,6,8-trione (Scheme 2(c)). Since the first step is slow, it limits the rate of the reaction. Hence, the CA oxidation involves two electrons and two protons.

(a)
(b)
(c)
3.4. Effect of Scan Rate

The relationship between the voltammetric signal and the scan rate (denoted as ) could give important information about the voltammetric mechanism. Hence, the relationships between on , , and were performed by cyclic voltammetry, as shown in Figure 5(a).

If the electrooxidation reaction is irreversible, the peak potential is independent of the scan rate. As can be seen from Figure 5(b), seems to be independent of , and the peak potential shifts only slightly to a higher potential as the scan rate increases. It may, therefore, be concluded that electron transfers in the AA, PA, and CA electrooxidation are quasi-reversible. increases with the increase in from 0.1 V·s−1 to 0.4 V·s−1 (Figure 5(c)), indicating that the electron transfer reaction involves a surface-confined process [46].

The plots of versus were established to determine whether the electrooxidation reactions are controlled by adsorption or diffusion processes. If the plot of versus is linear and the slope is close to 1, the electrooxidation reaction is an adsorption-controlled process. Otherwise, the slope is close to 0.5, then the process is controlled by the diffusion [51]. Natural logarithm for AA, PA, and CA as a function of is expressed as follows:

The slope of plots for , , and against (Figure 5(d)) are highly linear (, ) and 0.70, 0.863, and 0.64, respectively. The values of the slope are around 0.5 and 1. This indicates that AA, PA, and CA electrooxidations are diffusion-adsorption-controlled processes.

3.5. Effect of Pulse Amplitude, Accumulation Potential, and Time

The influence of accumulation potential and time on the of the analytes on ErGO/GCE was also investigated using the DP-ASV method in the range of −0.2 V to 0.3 V (Figures 6(a) and 6(b)).

As can be seen in Figure 6(a), increases as the potential shifts positively from −0.2 V to 0.3 V but of AA and CA decreases. The appropriate accumulation potential of −0.1 V was selected for AA, PA, and CA. The accumulation time also has an effect on the peak current. From Figure 6(b), it can be seen that increases slightly with time and peaks at 45 s and then decreases afterwards. Therefore, 45 s was chosen as the accumulation time.

Pulse amplitudes significantly affect the voltammetric signal of the analytes. The pulse amplitude in the range of 0.04 V to 0.10 V was used in the DP-ASV method (Figure 6(c)). As seen in Figure 6(c), increases linearly and shifts negatively as the pulse amplitude increases. is larger than 0.06 V. The width of the current peak tends to be large, the peak resolution reduces, and the selectivity also reduces. The pulse amplitude of 0.06 V that provides a symmetric peak with low RSD (0.6%) was selected for the next experiments (Figure S3).

3.6. Effect of Interferents

The effect of the interferents on the anodic peak currents of AA, PA, and CA was estimated based on the relative error of the AA, PA, and CA when there is an interferent in the solution as opposed to its absence. Some possible interferents were investigated by adding them to a solution containing an equal concentration of AA, PA, and CA in the B–R buffer solution of pH 3.2 (5 μM). Inorganic cations including K+, Na+, Ca2+, NH4+, CO32−, NO3, SO42−, Cl with concentration ratios (interference compound : standard solution) of 1000 : 1 ÷ 3100 : 1 (mM/mM), and glutamic acid, benzoic acid, and glucose with concentration ratios in the range of 10 : 1 to 310 : 1 were tested as interferents (Tables S511). The analysis of the obtained responses made it possible to conclude that these compounds do not significantly interfere (<5%) in the determination of PA and CA under employed working conditions. However, dopamine and uric acid significantly affect the voltammetric signals of the analytes even when the molar ratio interferent : standard solution is equal to 1 : 10.

3.7. Repeatability, Linear Range, and Limit of Detection (LOD)

The repeatability of ErGO/GCE for DP-ASV was estimated with different analyte concentrations. The DP-ASV curves were drawn to detect the mixtures of , and , and , and , in which each signal was obtained by successive measurements for nine times (Figure S2). The obtained RSD is 0.76%, 1.00%, and 1.83% for , , and , respectively; 1.28%, 3.03%, and 1.08% for , , and , respectively; and 2.21%, 2.14%, and 2.35% for , , and , respectively. These values are lower than 1/2 RSDH [52]. Such reasonable RSD of the successive measurements indicates that ErGO/GCE can be repeatedly utilized for the detection of AA, PA, and CA in either low or high concentration range (Figure S4).

The individual determination of AA, PA, and CA was performed using the DP-ASV method, in which the concentration of one compound was raised while that of the other two was kept constant in the 0.2 M B–R buffer solution of pH 3.2 (Figure 7). As can be seen in the figure, the peak current of the two compounds with fixed concentration was fairly unchangeable while that of the third analyte rises with the increase of its concentration. This suggests that the oxidation reactions take place independently. The peak current increases in the concentration range from 0.2 μM to 4.4 μM. The linear regression equations are as follows:

In the range of 0.2 μM to 4.4 μM, the limit of detections (LOD) calculated using for AA, PA, and CA are 0.36, 0.25, and 0.23 μM, respectively.

The simultaneous determination of AA, PA, and CA was also conducted using the DP-ASV method. AA, PA, and CA were estimated simultaneously by increasing their concentrations in the 0.2 M B–R buffer solution of pH 3.2. Figure 8 presents the DP-ASV curves obtained for solutions containing an equal concentration of AA, PA, and CA in the range of 0.2 μM to 4.4 μM. The DP-ASV curves show oxidation peaks at around 0.16 V for AA, 1.25 V for PA, and 0.54 V for CA. The larger separation peak of the analyte favors simultaneous determination. The linear regression equations are expressed as follows:

In the range of 0.2 μM to 4.4 μM, the values of LOD were 0.30, 0.24, and 0.26 μM for AA, PA, and CA, respectively. It is noted that LOD of AA, PA, or CA in the simultaneous and individual determination is very close to each other. This, again, reaffirms that the oxidation of these compounds on the ErGO/electrode is independent. Hence, the simultaneous voltammetric determination of AA, PA, and CA is possible on ErGO/GCE in the mixed samples without any cross interference. This linearity is also acceptable for the simultaneous determination of AA, PA, and CA in pharmaceutical preparations, in which their amount is in the range of 100–200 mg, 200–500 mg, and 20–65 mg, respectively.

A comparison of the proposed method with other voltammetric methods for AA, PA, and CA determination is listed in Table 3. It could be noticed that LOD of AA, PA, and CA from the proposed method is lower or comparable with those results based on modified electrodes in previous papers [47, 49, 50, 5258]. Overall, ErGO is proved to be an effective electrode modifier for the determination of AA, PA, and CA.

3.8. Real Sample Analysis

The proposed method was employed to determine the content of PA and CA in Panadol Extra, Hapacol Extra, and Tatanol (all of them are produced in Vietnam) and AA and PA in Effe Paracetamol, Ameflu C, and Efferalgan Vitamin C. Table 4 presents the amounts of AA, PA, and CA in the preparations with those determined using the HPLC method. The paired t-test was applied to these results. With the significant level , the results show that the amounts of AA, PA, and CA determined with the proposed method found in pharmaceutical preparations are fairly close to the labeled amounts (AA: , ; PA: , ; and CA: , ). The results obtained with the proposed and HPLC methods are not statistically different (AA: , ; PA: , ; and CA: , ). Therefore, the precision of the determination was also found to be good. In these experiments, the concentrations of AA, PA, and CA were determined by means of standard addition method. The values of recovery of the spiked samples were conducted to evaluate matrix effects after the standard solution additions. The obtained results yielded good average recoveries, ranging from 96% to 108% for AA, 95% to 106% for PA, and 98% to 104% for CA suggesting that the present method does not suffer from any significant effects of matrix interference.

4. Conclusions

The present paper shows that the electrodeposition and electroreduction of graphene oxide using cyclic voltammetry have been successfully carried out on the glassy-carbon electrode. The modified electrode thereof can be used for the simultaneous determination of ascorbic acid, paracetamol, and caffeine. The results also reveal that the modified electrode exhibits high electrocatalytic activity towards the electrooxidation of ascorbic acid, paracetamol, and caffeine by enhancing the peak currents. Large peak separation between ascorbic acid and paracetamol could be obtained using DP-ASV, and this can facilitate their simultaneous and individual determination on the ErGO-modified glassy-carbon electrode. The excellent analytical performance and simple fabrication of ErGO make it a promising material for highly sensitive and selective electrochemical biosensors. Furthermore, the proposed method has been utilized for the simultaneous determination of ascorbic acid, paracetamol, and caffeine in pharmaceutical preparations with promising results.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This work was supported by Project B2017–DHH–38 sponsored by the Ministry of Education and Training, Vietnam.

Supplementary Materials

Table S1 shows the order of experiments according to the Box–Behnken experimental design for AA. The rightmost column provides the values of peak current corresponding to each experiment. Table S2 shows the order of experiments according to the Box–Behnken experimental design for PA. The rightmost column provides the values of peak current corresponding to each experiment. Table S3 shows the order of experiments according to the Box–Behnken experimental design for CA. The rightmost column provides the values of peak current corresponding to each experiment. Table S4 presents estimated effects and coefficients of second-order polynomial equations. All the statistical and mathematical calculations were conducted using Minitab version 16. Table S5 shows that NaNO3 does not interfere with the peak currents of AA, PA, and CA. Table S6 shows that K2CO3 does not interfere with the peak currents of AA, PA, and CA. Table S7 shows that CaCl2 does not interfere with the peak currents of AA, PA, and CA. Table S8 shows that (NH4)2SO4 does not interfere with the peak currents of AA, PA, and CA. Table S9 shows that glucose does not interfere the peak currents of AA, PA and CA. Table S10 shows that glutamic acid does not interfere with the peak currents of AA, PA, and CA. Table S11 shows that benzoic acid does not interfere with the peak currents of AA, PA, and CA. Figure S1 presents CVs in in 1 mM K3[Fe (CN)6]/K4[Fe (CN)6] at various scan rates to calculate the effective surface area of electrodes. Figure S2 presents the plot of the kinetic parameter ψ versus ν1/2 to obtain the heterogeneous electron transfer rate constant. Figure S3 shows DP-ASV curves using the pulse amplitude in the range of 0.04 V to 0.10 V. The pulse amplitude of 0.06 V provides a symmetric peak with low RSD (0.6%). Figure S4 represents the DP-ASV curves of AA, PA, and CA in which each signal was obtained by successive measurements for nine times. (Supplementary Materials)