Journal of Ophthalmology / 2012 / Article / Fig 2

Clinical Study

Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster

Figure 2

Bland-Altman plot of ACD: Galilei versus IOLMaster. Bland-Altman plot reveals that, on average, the Galilei measured longer anterior chamber depth (ACD) values than the IOLMaster by 0.12 mm (solid line), with a 95% limits of agreement (LoA) from −0.09 to 0.34 mm (dotted lines). Nearly all data lie within the 95% LoA and are evenly distributed, indicating no relationship between average ACD and interdevice difference.
430249.fig.002

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.