Journal of Ophthalmology / 2012 / Article / Tab 1

Clinical Study

Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster

Table 1

Summary of ACD measurements and mean interdevice difference between Galilei and IOLMaster.

DeviceMean (range) ACDSD

Galilei3.37 (2.62–4.13)0.36
IOLMaster3.25 (2.34–3.92)0.38
Galilei-IOLMaster0.12* 0.11**

* 𝑃 < 0 . 0 0 0 1
**95% LoA −0.09 to 0.34
ACD: anterior chamber depth; SD: standard deviation; LoA: limits of agreement. All measurements in millimeters.
Mean ACD, SD, and range of values are shown above for the Galilei and IOLMaster. Mean interdevice difference in ACD is statistically significant.

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.