Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster

Table 1

Summary of ACD measurements and mean interdevice difference between Galilei and IOLMaster.

DeviceMean (range) ACDSD

Galilei3.37 (2.62–4.13)0.36
IOLMaster3.25 (2.34–3.92)0.38
Galilei-IOLMaster0.12* 0.11**

* 𝑃 < 0 . 0 0 0 1
**95% LoA −0.09 to 0.34
ACD: anterior chamber depth; SD: standard deviation; LoA: limits of agreement. All measurements in millimeters.
Mean ACD, SD, and range of values are shown above for the Galilei and IOLMaster. Mean interdevice difference in ACD is statistically significant.