Comparison of Two Different OCT Systems: Retina Layer Segmentation and Impact on Structure-Function Analysis in Glaucoma

Table 3

Total macula and ganglion cell layer mean thickness, in micrometers (µm), among groups and p values in the central 10°.

POAGControls valueAUC

MT (mean ± SD)
Cirrus
 MT297.6 ± 13.5307.9 ± 12.10.0360.789
Spectralis
 MT319.0 ± 13.0329.8 ± 11.70.0270.801
 cMT319.0 ± 12.9329.8 ± 11.70.0270.805

GCIPL (mean ± SD)
Cirrus
 GCIPL68.5 ± 8.780.7 ± 4.7<0.010.879
 mGCIPL61.2 ± 10.879.5 ± 4.7<0.010.930
Spectralis
 GCIPL75.7 ± 13.893.3 ± 4.6<0.010.886
 cGCIPL76.1 ± 13.793.4 ± 4.6<0.010.886

MT: full macular thickness, GCIPL: ganglion cell-inner plexiform layer thickness, cMT and cGCIPL: average values after manual correction of layer segmentation in Spectralis OCT, mGCIPL: minimum GCIPL value calculated by Cirrus software. AUC: area under the ROC curve. SD: standard deviation. values were obtained with paired -test and posterior adjustment with FDR (false discovery rate).