Journal of Ophthalmology
Volume 2017, Article ID 5763172, 13 pages
https://doi.org/10.1155/2017/5763172
Research Article

Characterization of Soft Contact Lens Fitting Using Ultra-Long Scan Depth Optical Coherence Tomography

1School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, China
2Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami, Miami, FL, USA
3Electrical and Computer Engineering Department, University of Miami, Miami, FL, USA

Correspondence should be addressed to Jianhua Wang; ude.imaim.dem@3gnawj

Received 12 February 2017; Accepted 12 April 2017; Published 1 June 2017

Academic Editor: Antonio Queiros

Copyright © 2017 Lele Cui et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Objectives. To evaluate the centration and movement of soft contact lenses and to verify the repeatability of two repeated measurements of the lens centration and movement using ultra-long scan depth optical coherence tomography (UL-OCT). Methods. A 1-day Acuvue® Define™ lens was tested on both eyes of 10 subjects (5 males and 5 females; mean age, 31.6 years). The centration and blink-induced movement of the contact lens were measured using UL-OCT at 5 min and 30 min after insertion. The measurements were repeated once at each checkpoint. Results. Good repeatability was found in the lens centration and movement between the two repeated measurements at either checkpoint. The values of the lens movement were 0.457 ± 0.248 mm and 0.402 ± 0.229 mm at 5 min and decreased to 0.197 ± 0.065 mm and 0.211 ± 0.110 mm at 30 min after insertion for the right and left eyes, respectively (). Conclusions. The custom-built UL-OCT presented good repeatability of centration and movement in Define lenses at 5 min and 30 min after insertion. Most of the lenses were centered temporal and inferior to the cornea during the first 30 min wearing period. Compared with 5 min after insertion, the lens was centered better and exhibited less movement at 30 min.