Research Article
Intraocular Telescopic System Design: Optical and Visual Simulation in a Human Eye Model
Table 2
Optical and visual results for both telescopic systems (near target distance).
| | Decentration (mm) | Optimized system | Nonoptimized system | | RMS (wavelengths) | VSOTF | RMS (wavelengths) | VSOTF |
| ITS 1 | 0.0 | 0.00050 | 0.99997 | 0.08162 | 0.51337 | 0.2 | 0.03162 | 0.75714 | 0.13343 | 0.20787 | 0.4 | 0.07285 | 0.42921 | 0.25653 | 0.07948 | 0.6 | 0.13094 | 0.29042 | 0.44222 | 0.05522 | 0.8 | 0.21107 | 0.23356 | 0.69713 | 0.07219 | 1.0 | 0.31791 | 0.21931 | 1.02795 | 0.09214 |
| ITS 2 | 0.0 | 0.00058 | 0.99997 | 0.06363 | 0.64374 | 0.2 | 0.03885 | 0.71124 | 0.12556 | 0.17792 | 0.4 | 0.10720 | 0.44347 | 0.30485 | 0.07998 | 0.6 | 0.22132 | 0.39109 | 0.60986 | 0.09826 | 0.8 | 0.39469 | 0.37503 | 1.05769 | 0.14287 | 1.0 | 0.64508 | 0.17847 | 1.67465 | 0.00579 |
|
|