Clinical Performance of the Spot Vision Photo Screener before and after Induction of Cycloplegia in Children

Table 4

Comparison of previous studies to current study of Spot Screener in detecting ARFs accordingly to 2013 AAOPS criteria.

StudyDeviceAgeSensitivitySpecificityPPVNPVCompared with

PeterseimSpot (v2.0.16)11–22184.870.978.179.2CR
AranaSpot (v2.0.16)20–11992.690.658.198.9CR
ForcinaSpot (v2.0.16)6–3589.870.458.993.6CR
MarzolfSpot (2.1.4)26.4–110.484625886CR
MuSpot (version missing)48–8494.7985CR
QianSpot (v 2.1.4)48–729480CR
Current studySpot (v3.0.04.06)36–12060.994.975.790.4CA
Spot cycloplegic36–12085.387.463.695.8CA

Age: months; PPV: positive predictive value; NPV: negative predictive value; CR: cycloplegic retinoscopy; CA: cycloplegic autorefractometer.