Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients

Table 2

Intraobserver repeatability outcomes for biometric measurements obtained using IOLMaster 700 swept-source optical coherence tomography in normal subjects.

ParameterObserverMean ± SDSwTRTCoV (%)ICC (95% CI)

AL (mm)1st25.1 ± 1.850.010.020.021.000 (1.000 to 1.000)
2nd25.1 ± 1.850.010.020.031.000 (1.000 to 1.000)
CCT (μm)1st542.84 ± 31.962.216.110.410.995 (0.993 to 0.997)
2nd542.54 ± 32.132.206.100.410.995 (0.993 to 0.997)
ACD (mm)1st3.56 ± 0.270.010.040.400.997 (0.996 to 0.998)
2nd3.56 ± 0.280.020.040.430.997 (0.996 to 0.998)
LT (mm)1st3.75 ± 0.270.020.040.410.997 (0.996 to 0.998)
2nd3.75 ± 0.270.020.050.450.996 (0.995 to 0.997)
Km (D)1st43.72 ± 1.360.100.280.230.994 (0.992 to 0.996)
2nd43.73 ± 1.350.090.250.210.996 (0.994 to 0.997)
WTW (mm)1st12.04 ± 0.350.090.240.720.940 (0.917 to 0.958)
2nd12.04 ± 0.370.090.250.740.943 (0.920 to 0.960)

AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, K = keratometry, WTW = white to white, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.