Research Article

Comparing School-Aged Refraction Measurements Using the 2WIN-S Portable Refractor in Relation to Cycloplegic Retinoscopy: A Cross-Sectional Study

Table 3

Reliability and repeatability of 2WIN-S.

ParametersICC95% CI valueSw
LowerUpper

Spherical equivalent (D)0.9000.8160.939<0.010.61
J00.6660.5570.748<0.010.30
J450.6390.520.728<0.010.31

value <0.05. CI, confidence interval; D, diopter; ICC, intraclass correlation coefficient; J0, (cylinder/2) cos (2 axis); J45, (cylinder/2) sin (2 axis); Sw,within-subject standard deviation.