Research Article

Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius

Table 1

The biometric parameters measured by IOLMaster.

ParametersPreoperative3-month postoperativeP

ACD (mm)3.69 ± 0.233.49 ± 0.38< 0.001
LT (mm)3.59 ± 0.223.75 ± 0.35< 0.001
AL (mm)26.21 ± 1.2326.20 ± 1.220.233
WTW (mm)11.86 ± 0.3811.88 ± 0.400.100
K1 (D)43.27 ± 1.2443.31 ± 0.140.117
K2 (D)44.71 ± 1.6244.54 ± 1.66< 0.001
CCT (μm)516.09 ± 30.51515.69 ± 31.140.648
IOL (D)12.54 ± 2.6912.61 ± 2.620.013

ACD, anterior chamber depth; LT, lens thickness; AL, axial length; WTW, white to white; K1, corneal flat axis curvature; K2, corneal steep axis curvature; CCT, central corneal thickness; IOL, intraocular lens. .