Simultaneous Corneal Topography and Epithelial Thickness Mapping from a Single Measurement Using Optical Coherence Tomography

Table 2

Interobserver reproducibility results for ETMs using REVO NX based on the first readings from each session taken by 2 operators in a sample of 40 eyes.

SectorMean ± SDTRTSwCoV (%)ICC

Superior (μm)51.40 ± 3.312.630.951.850.92
Superior-temporal (μm)51.38 ± 3.113.071.112.160.87
Temporal (μm)51.38 ± 3.102.700.971.900.90
Inferior-temporal (μm)51.38 ± 3.082.480.901.740.91
Inferior (μm)51.30 ± 3.032.690.971.890.90
Inferior-nasal (μm)51.32 ± 2.972.951.062.070.87
Nasal (μm)51.32 ± 3.032.460.891.730.91
Superior-nasal (μm)51.45 ± 2.942.971.072.080.86
Central (μm)51.47 ± 3.252.000.721.400.95
Superior_Out (μm)51.40 ± 2.813.781.362.650.76
Superior-temporal_Out (μm)51.50 ± 2.703.591.302.520.77
Temporal_Out (μm)51.38 ± 2.664.061.462.850.69