Research Article

Novel HPLC Analysis of Hydrocortisone in Conventional and Controlled-Release Pharmaceutical Preparations

Table 9

Statistical comparison between the developed method and the USP method.

USP method (column B) versus developed method (column A)

-test
value0.9742
value summaryNs
Significantly different? ()No
One- or two-tailed value?Two-tailed
Mean ± SEM of column A97.63 ± 1.415 = 15
Mean ± SEM of column B97.56 ± 1.440 = 15
Difference between means−0.06600 ± 2.019
95% confidence interval−4.202 to 4.070