Abstract

Antibiotics are nowadays used and abused worldwide in common veterinary practice to treat diseases, prevent infections, and promote animal’s growth. Drug resistance occurrence is a relevant phenomenon that can inactivate the antibiotic. Moreover, antibiotics used for animals are also found in milk and it poses serious health risks to humans. Another negative effect is related to milk processors since antibiotics cause detrimental effects on cheese and yogurt starter bacteria. Given its consumption as both beverage and derivatives, milk is one of the most regulated products in food industry. Nowadays several commercial tests are available to investigate antibiotics in milk, but they generally provide a qualitative result, require bulky procedure, and are time consuming. In this paper, we investigate the use of a chromogenic cephalosporin to quantify the concentration of cloxacillin—a lactam difficult to be detected by using “cowside” screenings which is the drug of choice in the method of mastitis control. The proposed measurement method and prototype have been demonstrated to be able to detect cloxacillin in milk at concentrations lower than the MRL set by the European Commission. Moreover, they are also able to detect the illegal practice of using lactamase to degrade lactams in milk.

1. Introduction

According to World Health Organization, in 2010, 600 million people suffered foodborne illnesses and 420,000 died [1]. As a result, there is a continuing search for new measurement techniques able to investigate food quality ranging from staple foods such as cereals and cereal products [24] to more niche products such as wines [5, 6].

Given its consumption as both beverage and derivatives such as cheeses, butter, and yogurt, milk is one of the most regulated products in food industry.

Nowadays almost all livestock receive medication during their lives and antibiotic residues in milk are of great concern to consumers, regulatory agencies, dairy farmers, and milk processors. Indeed, antibiotic residues in milk can [7] (i) seriously spread drug-resistant microorganisms, (ii) cause human health problems such as allergic reaction and imbalance of intestinal microflora, and (iii) badly affect cheese and yogurt starter bacteria. Therefore, there is great interest in methods capable of a fast, reliable, and multiantibiotic sensing.

Among veterinary antibiotics, lactams are one of the most widely used drug classes for the treatment of animals in Europe [8] and all over the world; hence, maximal residue limits (MRLs) have been set by major regulatory agencies, for example, the European Commission Regulation 2377/90. Thus, in the last decades, many measurement methods have been developed to detect lactams in food and feed samples [8, 9].

The oldest and simpler methods are microbial inhibition tests consisting in adding milk samples to organisms susceptible to the target antibiotics, incubating them, and finally monitoring eventual fails to grow.

Then, many other semiquantitative and quantitative methods such as enzymatic colorimetric assays, microbial receptor assays, receptor binding assays, enzymatic assays, chromatographic analysis, and immunoassays have been developed [79]. However, suspected samples detected by qualitative method generally requires being quantified and confirmed by using techniques such as high pressure liquid chromatography (HPLC) and mass spectrometry (MS) [8, 9] which are expensive and impractical for field testing.

Also commercial screening tests such as Charm Farm Cowside by Charm Sciences Inc. (USA), IDEXX SNAP by Thermo Fisher Scientific (USA), Delvotest SP by DSM (Holland), and BetaXpress, TwinsensorBT, and TriSensor by Unisensor Diagnostic Engineering (Belgium) are available to investigate lactams presence in milk samples. Nevertheless, they provide qualitative results only, require bulky procedure, and are time consuming. As an example, the TwinsensorBT requires the following: (i) add l of milk into one reagent microwell and mix, (ii) incubate it for 3 minutes at 40°C, (iii) dip one Dipstick into each microwell, (iv) continue incubating for 3 min at 40°C, and, finally, (v) read the color intensities.

Driven by the continued search for new measurement techniques, several other methods have been proposed (i.e., [1015]). Nonetheless, most of such methods are laborious and time consuming and require modifications of the antibody used in the assay.

Moreover, sometimes dairy industries illegally add lactamase in milk to degrade lactams antibiotics [14, 16, 17]. Such illegal practice may reduce lactams concentration, but it exposes consumers to risks to the health since the safety of lactamase has not been fully evaluated [17].

In this paper, we propose a measurement method based on a relatively low cost chromogenic lactamases substrate (CENTA) and we developed and tested a simple and inexpensive prototype measuring instrument potentially able to detect and quantify lactam antibiotics in milk. In particular, this preliminary study has been focused on the possibility of measuring cloxacillin. Among lactam antibiotics, cloxacillin has been chosen since (i) it is from a long time the drug of choice in the method of mastitis control [7, 18], (ii) it is difficult to be detected by using multiresidue tests [7], and (iii) its detection limit for some of the most commonly used “cowside” screening tests is above the legal MRLs, resulting in false negatives [7].

Moreover, as it will be discussed in detail, the developed measurement method and system are also intrinsically sensitive to the eventual presence of lactamase illegally added to the milk.

Since cattle produce about % of world milk production (followed by buffaloes with ≈11%, goats with ≈2%, and sheep with ≈1%) [19], such preliminary study has been focused on cow’s milk.

In the following, Section 2.1 briefly reviews competitive assays at the basis of the proposed measurement method; such method is then described in Section 2.2.

The developed prototype measuring system (proof of concept) and measurement procedure are described in Section 2.3 and the test performed to verify the method and to calibrate and test the measuring system is reported in Section 2.4.

The obtained results are reported and discussed in Section 3, and conclusions are drawn in Section 4.

Finally, detailed descriptions of both the performed data analysis and lactamases detection are reported in Appendices A and B, respectively.

2. Materials and Methods

2.1. Introduction to Two-Step Competitive Assays and Enzyme Kinetic

As described in detail in Section 2.2, the proposed measurement method is a two-step competitive assay.

Differing from direct assays (noncompetitive), in competitive assays, the concentration/presence of the analyte of interest—the measurand—is measured by adding a known amount of another analyte, competitive with the analyte of interest, then exploiting its ability to compete with analyte added to the sample.

Obviously, the concentration of the competitive analyte added to the solution has to be properly designed to match the required measuring interval for the concentration of the analyte of interest [20, 21].

In particular, given the use of a two-step competitive format, the reagent concentration has to be in excess in comparison to the concentration of the measurand.

Then, as it will be described in Section 2.2, the “substrate” is first added to the specimen containing the measurand and, in the second step, the enzyme competitive with the analyte of interest—lactams antibiotics—is added.

For a fully competitive enzyme reaction, the time evolution of the concentration of the enzyme-substrate intermediate complex is [22]where is the time, is the free enzyme concentration at time , is the substrate concentration, is the substrate concentration at time , is the inhibitor concentration, and, , , and are constants.

2.2. The Proposed Measurement Method

The proposed measurement method is based on a modification of the competitive binding assay reported by Bebrone et al. [23].

In particular, the method is based on the use of CENTA—a commercially available cephalosporin that can be prepared from the cephalothin [24]—and the lactamase enzyme (AmpC). Indeed, CENTA is not an antibiotic itself; therefore, it can be useful for analytical purpose when the lactam antibiotics have to be detected in a sample matrix by using a competitive assay.

It is known that CENTA is readily hydrolyzed by lactamases (AmpC); thus, according to Jones et al. [25], CENTA reaction with lactamase gives rise to changes in the absorption spectra of the CENTA moving its peak wavelength from about 340 nm to about 405 nm concomitantly with hydrolysis of the lactam ring.

According to the results of the inhibition tests reported in Section 3, cloxacillin and other lactams antibiotics are inhibitors of lactamase (AmpC) and reduce their effectiveness. The strength of such inhibition is described by the inhibition constant () whose value depends on the specific antibiotic and substrate. Therefore, the proposed measurement method is based on the hypothesis that if some lactam antibiotic inhibitor of AmpC is present in the milk sample, it will compete with the CENTA for the active sites of the lactamase enzymes (AmpC). As a result, we propose a competitive binding assay in two steps in which known amounts of CENTA ( in (1)) and AmpC ( in (1)) are subsequently added to the sample under test, thus analyzing the variation in the absorption spectrum of the sample itself.

According to (1), if no lactams antibiotics are present in the milk sample (), a fixed and known variation in the absorption spectrum is expected. However, as schematically shown in Figure 1, in the presence of inhibitors, the AmpC is expected to produce a reduced variation of the CENTA spectrum. Therefore, once the concentration of the AmpC () is fixed, the higher the variation of the absorption spectrum of the CENTA, the lower the concentration of the inhibitor ( in (1)) in the tested milk sample.

2.3. The Developed Prototype Measuring System and Measurement Procedure

According to the enzymes kinetic described in Sections 2.1 and 2.2, the developed measuring system analyzes the time evolution of the absorbance of the sample as a consequence of the additions of known amounts of CENTA and AmpC.

In particular, as shown in Figure 2 during measurement the milk sample is placed in a cuvette with an optical path of 1 mm; hence, it is illuminated by using a LED with a peak wavelength of 405 nm (model, LED-405-33V, Roithner LaserTechnik—nominal full width at half maximum at 20 mA equal to 15 nm) and the absorbance at such wavelength is estimated by using the photodiode PD (model HFD 1100, EG&G).

The system is then controlled by using an Arduino UNO board which (i) switches on/off the LED, (ii) analyzes the signal provided by the photodiode, and (iii) communicates with the PC through a USB port. Indeed, the control electronic is basically composed of a constant current generator, whereas the front-end electronic is composed of a transimpedance amplifier and an instrumentation amplifier. The output voltage of the front-end electronic is digitalized by using a 10-bit ADC included in the Arduino UNO board.

The so acquired data are transmitted to the PC and elaborated according to the measurement model described in the following.

In particular, the system analyzes the acquired digitized samples of signal according to the timing diagram shown in Figure 3 and described in Table 1. Supposing that the measurement starts at time and then adding the CENTA and the AmpC at times and , respectively, and values are the mean and values recorded during the time intervals and , respectively:Then, as described in Appendix A, the antibiotic concentration has been estimated by analyzing the final or steady-state pseudo-absorbance :where and are the constants of the calibration function and, as described in Appendix A, was estimated by interpolating the function defined aswith the functionwhere was the independent variable, was the dependent variable, and , and were the parameters.

As described in Appendix A, value in (4) is the mean value recorded before starting the measurement and replacing the cuvette with a beam blocker. Such value was exploited to compensate offsets.

Similarly, value defined in (2) was exploited to check for anomalies due, as an example, to variation in the power emitted by the LED and/or unnatural absorbance of the milk sample. Moreover, as described in Appendix B,  also allows potentially detecting the illegal practice of adding lactamase to milk.

A picture of the developed prototype measuring system is shown in Figure 4.

2.4. The Performed Experimental Activities

As previously introduced, the performed experimental activities have been aimed at investigating both the feasibility of the proposed measurement method and the performance of the developed prototype measuring system.

Thus, four different activities were performed: (i) inhibition test, (ii) feasibility verification, (iii) system calibration, and (iv) system verification.

All experiments have been performed at room temperature and without controlling/modifying the temperature of the milk samples.

All the used milk was whole-milk for human consumption and it was bought on the market; thus, it was supposed to be free of antibiotics. Measurements have been then obtained without diluting the milk samples.

Finally, an Escherichia coli K12 AmpC -lactamase [26] was used for all experimental activities. The concentration of AmpC was determined spectrophotometrically in concentrated stock solutions made from lyophilized powder and subsequently diluted; this enzyme had been previously expressed and purified, as described by Powers et al. [27].

2.4.1. Inhibition Test

Since the basic idea of the proposed method is that -lactams could behave as inhibitors of AmpC-type -lactamases, the strength of such inhibition has been evaluated in terms of inhibition constant () towards AmpC lactamase in buffer and CENTA substrate. In particular, besides cloxacillin, the test investigated cefalexin, cefoperazone, cefalonium, oxacillin, and dicloxacillin.

2.4.2. Feasibility Verification

Feasibility tests have been aimed at investigating the stability of CENTA and AmpC in the milk matrix. Indeed, in case of reagents instability, the overall signal would be decreased for the concentration decrease and not for the enzymatic reaction.

All the feasibility tests have been performed by using an UV/Vis Spectrophotometers (Lambda 25, Perkin Elmer—scan-rate set equal to 60 nm/min) and a glass cuvette with an optical path of 1 mm. Then, all measurements have been performed at room temperature and without incubating the samples. As resumed in Table 2, the feasibility test “1” (FT1) has been aimed at acquiring the “standard” spectra of CENTA, thus firstly measuring the absorbance of buffer and then the absorbance of buffer and CENTA and finally measuring time evolution of the absorbance of the sample composed of buffer, CENTA, and AmpC after the addition of AmpC (4 spectra, minutes). The second feasibility test (FT2) has been aimed at investigating the effect of the milk on the absorption spectra of the CENTA. Hence, FT2 has been performed by firstly measuring the absorbance of the milk, then obtaining the absorbance by adding CENTA, and finally measuring the time evolution of absorbance resulting from the addition of the AmpC (7 spectra, minutes). The third feasibility test (FT3) has been aimed at investigating the effects of the competition between lactam antibiotics and CENTA on the optical absorbance of milk samples. Therefore, as reported in Table 2, the test FT3 investigated the spectra obtained from 4 samples: (i) milk (zero concentration of antibiotic), (ii) milk (zero concentration of antibiotic), CENTA, and AmpC, (iii) milk and antibiotic (cloxacillin, concentration 5 nM), CENTA,and AmpC, and (iv) milk and antibiotic (cloxacillin, concentration 10 nM), CENTA, and AmpC.

Note that the maximal residue limit (MRL) defined by the European Medicines Agency for cloxacillin in milk is about 70 nM [7]; therefore, about an order of magnitude is greater than the selected concentration of cloxacillin in our sample (see Table 2) (cloxacillin molecular weight is 435.88 g/mol). As a result, the FT3 tests were aimed also at a rough estimate of the potential “detection limit.”

2.4.3. Calibration

Calibration activities have been aimed at investigating the relation between the antibiotic concentration () and the pseudo-absorbance . In particular, as reported in Table 3, the calibration of the measuring system had been performed by realizing 8 measurement standards based on milk and antibiotic at known concentrations.

Then, each standard was analyzed according to the measurement procedure described in Section 2.3.

2.4.4. Verification

Verification has been performed by analyzing two (nominally equal) sets of milk samples with known antibiotic concentration, thus comparing the known concentration with results provided by the calibrated system. The composition of the used milk samples is reported in Table 3.

3. Results and Discussion

3.1. Inhibition Test

Inhibition test is resumed in Table 4. According to the reported results, the best inhibitors and thus the easier detectable antibiotics are dicloxacillin, cloxacillin, and oxacillin.

3.2. Feasibility Verification

The results of feasibility tests are shown in Figures 58.

As expected and shown in Figure 5, the CENTA spectrum presents an absorption peak at 345 nm. In the presence of the AmpC, the CENTA reacts with the AmpC active site and the absorption peak is shifted to 405 nm.

As shown in Figure 6, the presence of milk predictably modifies the absorption spectra with respect to the spectra recorded in buffer solution (mainly the signal to noise ratio). Nevertheless, the AmpC is still able to modify the spectra. Note that the spikes in Figure 6 at about 380 nm are reasonably measurement artifacts due to the lamp change performed by the spectrometer at such wavelength.

Finally, the settled absorbance spectra obtained by samples with and without antibiotics (FT3) are shown in Figure 7. As expected, the antibiotic competes with the CENTA for the AmpC protein active site; thus, the same amount of AmpC gives rise to a reduced variation of the absorption spectrum (see Figure 1). Such reduction in the effect of AmpC is visible also in Figure 8 where the time evolution of the samples absorbance at 405 nm is shown. Once again, the spikes at about 380 nm in Figure 7 are reasonably measurement artifacts due to the lamp change performed by the spectrometer at such wavelength.

3.3. Calibration

Calibration has been performed according to the procedure described in Section 2.4. An example of signal recorded using the developed measuring system is shown Figure 9. The results obtained by the analysis of the standards described in Table 3 are shown in Figure 10.

The calibration function has been obtained by least square fitting.

3.4. Verification

As described in Section 2.3, verification has been performed by analyzing two sets of milk samples with known antibiotic concentration, thus comparing the known concentration with results provided by the calibrated system. The obtained results are shown in Figure 11 and the absolute values of the percentage errors are shown in Figure 12.

4. Conclusions

In this paper, we propose a measurement method and we developed and test a “proof of concept” measuring system that have been demonstrated to be able to detect cloxacillin in milk at concentrations lower than the maximal residue limit (MRL) set by the European Commission. Furthermore, according to the results of the inhibition test reported in Section 3, the method could be suitable for the detection of other lactam antibiotics. However, at the current stage, the method can be used only to verify and quantify the concentration of cloxacillin and the presence of other antibiotics and inhibitors of AmpC can lead to measurement artifacts. Thus, it could be useful for both cattle breeders who want to rapidly test the quality of the milk of animals that have been treated with cloxacillin and inspectors (veterinaries, regulatory agencies, and milk processors) who want to test quickly and with high resolution the presence and concentration of cloxacillin in the milk produced by animals suspected of being treated with such antibiotic, for example, animals suffering mastitis.

Moreover, as discussed in Section 3.3 and Appendix B, the developed measurement method and system are also intrinsically sensitive to the eventual presence of lactamase illegally added to the milk to set to zero the lactams concentration. Therefore, they are potentially able to detect both lactam antibiotics and lactamase enzyme (AmpC) illegally added to milk. Indeed, as discussed in Appendix A and shown in Figure 7, an AmpC (Escherichia coli K12 AmpC -lactamase) concentration equal to some M can be easily detected.

CENTA, the used substrate, is not an expensive cephalosporin that can be prepared from the cephalothin [24]. Also the proposed measuring system is simple, inexpensive, and easy to use and transport to the field.

Much more testing is required and at the present time, there is no guarantee that this method could be applied for the detection of any other lactam antibiotics. Furthermore, the applicability of the method on different milk sample matrices (e.g., goat’s and sheep’s milk) has not been verified. Moreover, the measure of absorbance around 2.5 and 3.5 is difficult considering the turbidity of milk (as an example, consider the variation of the spectra from Figure 5—buffer—to Figure 6—milk). However, the reported results suggest that the proposed measurement method and measuring system may be able to provide a quantitative measure of lactam antibiotics concentration in milk.

As shown in Figures 7 and 8, although absorbance variation is very small and the values are very high, the proposed measurement method is substantially able to detect cloxacillin at concentration at about 5 nM. Indeed, according to (3) and Figure 10, the method is very sensitive at low concentration.

The developed prototype measuring system has been tested and demonstrated at concentration as low as 35 nM (half the MRL concentration set by the European Commission).

The performance of the measuring system can be probably further improved by (i) increasing the signal to noise ratio of the block PD/front-end electronic/ADC (see Figure 9) by redesigning the electronic and/or exploiting best performing components, (ii) tuning the bandwidth of the source, and (iii) exploiting the isosbestic point of the CENTA spectra, thus measuring the absorption of the sample also at 370 nm (see Figure 5).

Indeed, the used LED has a nominal full width at half maximum at 20 mA equal to 15 nm; thus, system performance can be reasonably improved by reducing the bandwidth of the source. Such can be accomplished by using optical filters or laser diodes (LD).

Coherent sources such as lasers can lead to problem related to interference (e.g., speckles). However, an automated power control of the laser diode [28] and the analysis of the resulting self-mixing signal can allow us to obtain further information such as the size of milk particles [29], hence potentially providing further information about the milk quality [30].

Moreover, although the absorption peak of the hydrolyzed CENTA is at about 405 nm (see Figures 5 and 6), as shown in Figure 7, the spectral region around 420 nm reveals a great variation of absorbance; thus, we are currently investigating the possibility of performing the measure in such spectral region.

Concluding, the proposed measurement method and the measuring system are simple, quantitative, sensitive, and easy to use on the field. Thus, they are useful in delivering rapid results, where access to laboratory analysis is impractical and uneconomic or delays are commercially unacceptable.

Appendix

A. Derivation of the Measurement Model

The proposed measurement method is based on the hypothesis that, once the concentration of the AmpC is fixed ( in (1)), the lower the concentration of the lactams antibiotics in the tested milk sample ( in (1)), the higher the hydrolyzation of CENTA by the AmpC, and thus the higher the variation of the absorption spectrum of the sample. Hence, the method investigates the concentration of the hydrolyzed CENTA .

According to Beer-Lambert’s law, the attenuation of a monochromatic and collimated light beam of irradiance traversing a semi-infinite slab of purely absorbing medium composed of chromophores—the light-absorbing species—is [31]where is the optical absorbance (quantity of dimension one historically referred to as OD (Optical Density)), is the irradiance of the beam exiting the slab (transmitted), is the (overall) absorption coefficient, is the thickness of the slab (optical path length), is the wavelength, and and are the concentration and the extinction coefficient of the chromophore. To take into account scattering, (A.1) is usually modified as follows [32]:where the differential pathlength factor () takes into account the extended pathlength due to scattering and represents the losses due to scattering.

In our system, supposing that is the (steady-state) overall optical absorbance of the milk sample after the addition of CENTA—time interval in Figure 3—from (A.1) and (A.2) the overall optical absorbance of the milk sample after the addition of AmpC can be written as follows:where is the extinction coefficient of the hydrolyzed CENTA and takes into account the extinction coefficients of the reactants consumed to produce hydrolyzed CENTA.

The digitized output voltage of the front-end electronic is proportional to the current photogenerated by PD. Hence, omitting noises, the signal depends on the optical power impinging on the photodiode active area:where is the product of the sensitivities of the photodiode and front-end electronic, takes into account the offsets, and is determined by the sensitivity and the geometry of the system. Indeed, the irradiance is related to the power emitted by the LED and the power is related to the irradiance . Then, to take into account the offset voltage , before starting the measurement, the cuvette was replaced with a beam blocker, thus obtainingHence, supposing that , , and the geometry of the measuring system do not change during the time interval , according to (A.1) and (A.3), the function, defined asis proportional to the hydrolyzed CENTA concentration . Then, since the final settled hydrolyzed CENTA concentration is determined by the lactams antibiotic concentration (), according to (1) the proposed measurement model estimates the final or steady-state pseudo-absorbance by interpolating the function withwhere, as described in Section 2.3, is the independent variable, is the dependent variable, and , , , and are the parameters.

Note that in our method the CENTA concentration is about three orders of magnitude higher than the AmpC concentration (see Table 3); thus, the substrate concentration in (1) is almost constant ().

Finally, from (1) the final or steady-state concentration of the hydrolyzed CENTA can be written as follows:Hence, from (A.6), the antibiotic concentration can be estimated from the final or steady-state pseudo-absorbance :where and are the constants of the calibration function that will be obtained experimentally.

Note that, according to (A.9), the proposed method is based on , thus substantially resulting in an end-point method.

B. β-Lactamase Detection

As described in Section 2.2, the proposed measurement method is based on the additions of known amounts of CENTA ( in eq. (1)) and, subsequently, AmpC ().

According to eq. (A.1) and as shown in Figure 6, the first addition of CENTA is expected to give rise to a known and fixed variation of the absorbance of the sample under test (curves and ). In other words, according to eq. (A.1):where is the absorption of the sample before the addition of CENTA, is the expected—no lactamases—absorption of the sample after the addition of CENTA and is the extinction coefficient of the CENTA.

Thus, if at the addition of CENTA some AmpC is already present in the sample, the absorption variation in the time interval ,will be different from (i.e., curves, , , , , , in Figure 6). In particular, according to (A.3), the differenceis proportional to the concentration of the (eventually) hydrolyzed CENTA and thus to the lactamase (eventually) illegally added to the milk. Hence, supposing that isrecorded during calibration (in the absence of lactamases in milk), in analogy to (A.6),is proportional to the concentration of the hydrolyzed CENTA . As a result, the system is able to reveal eventual lactamases illegally added to milk. The higher the concentration of the lactamases illegally added to milk, the higher the variation of the absorption spectra , resulting in a quantitative method. The achievable detection limit of lactamases reasonably depends on several parameters. However, as shown in Figure 7, few M of AmpC can be reasonably detected. Indeed, according to Figure 7, the addition of CENTA is expected to produce a negligible variation of the absorption at 420 nm (from curve , to curve ). On the contrary, an AmpC (Escherichia coli K12 AmpC -lactamase) concentration equal to 13.85 M gives rise to a steady-state variation greater than 0.4 OD (from curve , to curve ).

Note also that in our method the AmpC concentration is in excess in comparison to the concentration of the lactam; thus, the analysis of is expected to be able to reveal the presence of lactams even if some lactamases have been illegally added to milk. Nevertheless, the presence of lactamases illegally added to the milk sample results in a different (see (1)); hence, and constants of the calibration function are in general no longer valid; thus, the analysis provided by the developed measuring system will be qualitative only.

Conflicts of Interest

The authors declare that there are no conflicts of interest regarding the publication of this paper.