Research Article

FPGA Implementation of a Single Step MFCV Estimator Based on EMG in Diabetic Neuropathy

Table 4

State of the art solutions comparison table.

Parameters\workThis work[7][8][9]

PlatformEMG
FPGA
EMG
ASIC
EMG
μC
EMG
force controlled chair
PC
ApplicabilityOL1
Clin.2
EMG stimulationVoluntaryVoluntaryElectricalVoluntary
Involved limbLegArmArmArm
Num. of electrodes42864 (array)
Positioning Assistance
Computing methodSingle XNOR cross-correlationOn-going cross-correlationCross-correlationDelay-locked loop (DLL)
Timing3Real-timeReal-timeOnline computedOnline computed
resolution (ms@fclk)40.5 ms at 2 kHz0.5 ms at 2 kHz50 ms at 1 kHz1 ms at 1 kHz
ApplicationFatigue
OL MFCV monitoring
FatigueMFCV monitoringMFCV monitoring

1OL: ordinary life applicability. 2Clinical applicability. 3Real-time requests for acquisition + computing time < 400 ms. 4The resolution is upper limited by 2 kHz (present work sampling frequency).