Research Article
Development and Application of Wireless Power Transmission Systems for Wireless ECG Sensors
Table 1
Power transmission efficiency.
| Distance (mm) | Vp (V) | VR_Tx (V) | Is (A) | Ps (W) | PL (W) | Power efficiency (%) |
| 5 | 7.3 | 1.46 | 0.146 | 0.73 | 0.026645 | 3.65 | 10 | 6.7 | 1.48 | 0.148 | 0.74 | 0.022445 | 3.033108108 | 15 | 5.9 | 1.52 | 0.152 | 0.76 | 0.017405 | 2.290131579 | 20 | 4.8 | 1.56 | 0.156 | 0.78 | 0.01152 | 1.476923077 | 25 | 3.28 | 1.6 | 0.16 | 0.8 | 0.005379 | 0.6724 | 30 | 2.8 | 1.62 | 0.162 | 0.81 | 0.00392 | 0.483950617 | 35 | 2.3 | 1.62 | 0.162 | 0.81 | 0.002645 | 0.32654321 | 40 | 1.84 | 1.6 | 0.16 | 0.8 | 0.001693 | 0.2116 | 45 | 0.82 | 1.62 | 0.162 | 0.81 | 0.000336 | 0.041506173 | 50 | 0.85 | 1.62 | 0.162 | 0.81 | 0.000361 | 0.044598765 |
|
|