Research Article

A Pilot Study of Detecting Individual Sleep Apnea Events Using Noncontact Radar Technology, Pulse Oximetry, and Machine Learning

Table 3

Agreement for only pulse oximetry, only radar, and the combination of radar and pulse oximetry (Somnofy+SpO2) against HSAT.

5 class κ3 class κ2 class κREI 95% LoA

Only pulse oximetry0.280.340.71-13.96, 11.18
Only radar0.260.340.78-14.48, 9.98
Somnofy+SpO20.360.430.81-8.25, 7.47

κ = Cohen’s kappa for the whole dataset on seconds resolution, 5 class = no event/hypopnea/obstructive apnea/central apnea/mixed apnea, 3 class = no event/hypopnea/apnea, 2 class = no event/event, and REI 95% LoA = Bland-Altman 95% limits of agreement for REI.