Research Article

Portable Gas Analyzer Based on Fourier Transform Infrared Spectrometer for Patrolling and Examining Gas Exhaust

Table 3

Testing results with standard gas mixture.

ItemConcentration of analyte/φ(⋅) × 10−6
CH4C2H6C3H8iso-C4H10n-C4H10C2H4C3H6C2H2COCO2

1Exp97619.80000000760
Anal98918.75.24.52.201.30.91.8786
2Exp999729981990000000
Anal9882310212122.19.31.83.41.13.62.1
3Exp9981949819950.050.050.050.050.010.0230
Anal10130047521744.356.644.655.348.89.8221
4Exp1998849819910010049820050.415.01010
Anal2018048321894.680.146521446.315.31034
5Exp1997909996298729932497298999899960010020
Anal2014201025131293043261328231145100262310132
6Exp1000149710050.030.000020490
Anal996247391.654.622.63.93.12.123504
7Exp1001150110050.010019.89.702002520
Anal1030347483.554.911717.711.31.81872612
8Exp99892202110015006025012001003002300
Anal1003211963963492591513213963152387
9Exp501020200124987499350174989299829985002200020
Anal512120193224845507650944941308230494971203512
10Exp450042000100050040020.110.110.210.22200
Anal448412035107248443121.612.39.677.12215
11Exp19897297825530220049.840.019.8199700
Anal19674296527331622353.641.320.5221694
12Exp19818.115.010.000000504
Anal21215.618.616.15.92.34.80.02.5513
13Exp50210.15.000000020.2
Anal5127.10.20.15.20.54.13.1018.3
14Exp39995198710040.030.020.019.90400998
Anal40534183692.435.224.816.318.60.1364973
15Exp198998.219.010.0010.00500
Anal194295.215.510.83.212.80.63.56.10