Statistically Coherent Calibration of X-Ray Fluorescence Spectrometry for Major Elements in Rocks and Minerals

Table 3

Application of the XRF calibrations for the determination of major elements (mean concentration in % m/m; mass/mass unit expressed in percentage, habitually called wt.% and 99% uncertainty or simply ) in four rock samples (data reported as rounded value following flexible criteria proposed by [18]) and their comparison with the literature data.

RockMethod (ref.)SiO2TiO2Al2O3Fe2O3tMnOMgOCaONa2OK2OP2O5

ATT1OLR61.76.10.49610.017310.000.494.2460.2920.023940.001885.3730.3772.1390.0860.0830.0940.77770.03780.80740.0134
UWLR62.302.230.49510.01439.6660.3214.4020.2350.024160.002696.170.522.1440.0760.0580.0650.77810.03660.78280.0125
[76]59.60.80.490.0159.500.293.310.060.0250.0189.140.061.870.030.10.030.860.060.760.01

CSB1OLR52.75.50.14480.015724.600.713.4930.2890.014080.001881.5230.2701.1170.0852.1260.1190.41690.03740.03830.0060
UWLR52.762.040.13430.013024.400.483.5640.2330.013980.002681.6040.3981.1050.0752.1230.0830.43370.03610.03570.0057
[76]55.30.80.160.0321.750.183.880.030.0200.0031.940.081.280.062.070.200.540.03

GHOLR73.36.50.09680.015713.980.531.3360.2850.050840.001900.0740.2620.8430.0843.6180.1515.3020.0590.01190.0059
UWLR73.622.380.08610.012913.7140.3491.4660.2290.050640.002710.0380.3900.8280.0743.6250.1055.2160.0570.00910.0057
[77]75.80.390.080.0712.500.161.340.20.050.0090.030.180.690.133.850.114.760.050.080.03

TLM1OLR57.75.70.84970.020017.380.587.6370.3060.116140.001993.9720.3196.8070.1002.6750.1301.75850.04030.13780.0063
UWLR57.672.110.84020.016517.0460.3867.6820.2470.115920.002814.2810.4476.8210.0882.6670.0901.7420.03900.13360.0060
[78]58.850.350.8200.01317.480.257.670.150.1050.0133.30.086.670.082.980.071.670.080.150.013