Ranges of B, Cd, Cr, Cu, Fe, Pb, Sr, Tl, and Zn Concentrations and Isotope Ratios in Environmental Matrices from an Urban Area

Table 2

MC-ICP-MS operating parameters and measurement conditions for isotope ratio measurementsb. Table modified from Rodushkin et al. [2].

ElementsaConfiguration of introduction systemResolution modeIntegration time (s)Sample uptake rate (mL·min−1)Dynamic mode configurationCup configurationδ-zero” reference standardsReported ratios
L4L3L2L1CH1H2H3H4

BPumped micromist nebulizer, minicyclonic spray chamber, X-type skimmer coneLow0.5240.40–0.5010B11BNIST SRM 951a boric acidδ11B

CdAridus/Apex desolvating systems, self-aspirating microconcentric PFA nebulizer, X-type skimmer coneLow0.5240.04–0.06107Ag108Cd109Ag110Cd111Cd112Cd114Cd116Cd117SnNIST SRM 3108 Cd solution Lot 130116δ114Cd
108Pd110Pd112Sn114Sn116Sn

Zn/CuPumped micromist nebulizer, double spray chamber, H-type skimmer coneMedium0.2620.20–0.2563Cu64Zn65Cu66Zn67Zn68Zn70ZnNIST SRM 976 Cu standard solutionδ56Cu
(64Ni)IRMM-3702 Zn solutionδ66Zn

Cr/NiPumped micromist nebulizer, double spray chamber, H-type skimmer cone/Apex desolvating systems, self-aspirating microconcentric PFA nebulizer, X-type skimmer coneHigh1.2620.05–0.25Main52Cr53Cr54Cr56FeNBS SRM 979 Cr standardδ53Cr
54Fe
Sub configuration60Ni61Ni62Ni

Fe/NiPumped micromist nebulizer, double spray chamber, H-type skimmer coneMedium0.2620.20–0.2554Fe56Fe57Fe60Ni61Ni62NiIRMM-014 Fe metalδ56Fe
54Cr

Pb/TlAridus/Apex desolvating systems, self-aspirating microconcentric PFA nebulizer, X-type skimmer coneLow0.5240.04–0.06202Hg203Tl204Pb205Tl206Pb207Pb208PbNIST SRM 981 common Pb206Pb/207Pb;208Pb/207Pb
Tl: 1000 mg·L−1 elemental standard. Ultra scientific, North Kingstown, RI, USA; Lot L00709δ205Tl
204Hg

SrPumped micromist  nebulizer, double spray chamber, H-type skimmer coneLow0.2620.20–0.2582Kr83Kr84Sr85Rb86Sr87Sr88SrSr carbonate NIST SRM 987 (NBS-987) Sr carbonate87Sr/86Sr
84Kr86Kr87Rb

aOne element is used as the internal standard for the second element, except for Ni and Ag which were only used as the internal standards. bRF power: 1400–1450 W. Coolant gas flow: 15 L·min−1. Auxiliary gas flow: 1.4·L min−1. Sample gas flow: 0.9–1.25 L·min−1. Additional gas flow (N2, Aridus and Apex) 0.01–0.02 L·min−1. Ion lens settings: adjusted daily to obtain maximum sensitivity and signal stability. Zoom optic settings: adjusted daily to obtain maximum resolution. Number of blocks: 9. Number of cycles per block: 5. Number of integrations: 3–5. Amplifier rotation: left.