Research Article

A Limited Sampling, Simple, and Useful Method for Determination of Glomerular Filtration Rate in Cats by Using a New Accurate HPLC Method to Measure Iohexol Plasmatic Concentrations

Table 1

Recovery and intra- and interday precision results for the assay; CV: coefficient of variation; intraday in one day; interday for 7 consecutive days.

SampleNominal IOX concentrationsRecovery (%)Intraday precision (% CV)Interday precision (% CV)

Low25 μg/mL 1.53.8
Medium125 μg/mL 2.53.2
High500 μg/mL 4.06.5