Simulation Study of Sensitivity Performance of MEMS Capacitive Bending Strain Sensor for Spinal Fusion Monitoring

Table 5

Analysis outcome for different anchor length.

Anchor length, mm0.51.01.5
Nominal capacitance (at 0  ), pF100.1602117.2815134.4027
Capacitance at 1000  , pF37.307461.357684.7248
Sensitivity (0–1000  ), pF/ 0.06280.05590.0497
Maximum percentage difference, %11.915.004.35
Average percentage difference, %6.762.821.75

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.