Accuracy of Robot-Assisted Percutaneous Pedicle Screw Placement under Regional Anesthesia: A Retrospective Cohort Study

Table 3

Comparison of pedicle screw placement accuracy, FJV, and fluoroscopy time.

CharacteristicsGroup EP-ROGroup GE-FLUTotalvalue

No. of screws7494168
Accuracy grade
(n%)0.102
Grade A66 (89.2%)75 (79.8%)141 (83.9%)
Grade B4 (5.4%)11 (11.7%)15 (8.9%)
Grade C4 (5.4%)5 (5.3%)9 (5.4%)
Grade D2 (2.1%)2 (1.2%)
Grade E1 (1.1%)1 (0.6%)
Clinically acceptable
(Grade A + B)70 (94.6%)86 (91.5%)156 (92.8%)0.44
No. of screws for FJV comparison364682
FJV grade
(n%)0.009
Grade 034 (94.4%)33 (71.7%)67 (81.7%)
Grade 11 (2.8%)8 (17.4%)9 (11.0%)
Grade 21 (2.8%)2 (4.3%)3 (3.7%)
Grade 33 (6.5%)3 (3.7%)
Violating screws
(Grade 1 + 2+3)2(5.6%)13(28.3%)15(18.3%)0.01
Fluoroscopy time per screw (secs)5.4 ± 1.96.8 ± 2.06.2 ± 2.00.03
Postoperative pain rating (VAS)
4 hours after surgery4.7 ± 2.55.5 ± 2.00.26
Postoperative day 14.7 ± 1.95.1 ± 1.50.44
Postoperative day 23.9 ± 1.13.9 ± 1.30.94
Postoperative day 32.7 ± 1.02.7 ± 1.10.81

Values are the number or the number (%) of patients. Values are presented as the mean ± SD. FJV indicates facet joint violation. VAS indicates visual analogue score.