Research Article

Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types

Table 2

Mean visible lumen diameters and intraluminal attenuation of the investigated stents using four different reconstruction kernels. Diameters of the visible stent lumen are given as mean (minimum/maximum) ± standard deviation in % of the actual lumen. Attenuation and noise values are given as mean (minimum/maximum) ± standard deviation in Hounsfield units (HUs).

Stent nameB20fB30fB40fB50f
Diameter
(%)
Density
(HU)
Diameter
(%)
Density
(HU)
Diameter
(%)
Density
(HU)
Diameter
(%)
Density
(HU)

SAXX Small66,6262,969,4263,871,4236,474,6176,2
SelfX Xpert67,4263,171,4257,373,4252,774,6270,9
Palmaz Genesis65,4279,969,4273,770,0281,268,6194,4
Absolute65,3234,567,1233,171,9229,874,7269,4
AccuLink Carotid60,4260,065,7278,367,6258,371,9254,6
Express Vascular LD71,4244,473,9216,976,7243,477,1230,6
OmniLink 0,01867,6265,671,9252,675,7251,076,1253,0
OmniLink 0,03569,6249,772,9232,976,1264,376,1235,5
Palmaz Corinthian IQ71,4265,773,3255,774,3252,874,3244,1
Renal10952,4432,656,7436,959,6436,563,9586,8
Renal13749,0327,152,4345,953,3414,657,1643,3
Sentinol70,4234,576,1233,576,7239,180,0212,2
Symphony68,1254,271,9288,673,3243,875,3212,2
Vascuflex SE68,1244,672,4219,575,3234,776,7245,1
Wallstent Uni68,1273,569,6269,171,0262,771,0241,1
Zilver69,0263,771,4234,774,7219,976,7226,3
Evo76,3256,776,6256,877,9247,580,4284,9
RxCarotid66,3268,468,4250,369,1257,972,1293,4
SAXX Large72,9243,976,6240,676,6216,676,6272,2
Evo Target77,3228,381,7245,883,7239,883,3311,8
Wallgraft Endoprothesis72,3271,673,0276,575,3285,176,0396,8
CP Stent72,3222,276,4379,377,7457,481,3538,7

Mean67,6265,870,8270,172,8273,974,5299,7
Min49,0222,252,4216,953,3216,657,1176,2
Max77,3432,681,7436,983,7457,483,3643,3
SD6,643,16,453,26,468,45,7127,2