Effect of Different Breast Densities and Average Glandular Dose on Contrast to Noise Ratios in Full-Field Digital Mammography: Simulation and Phantom Study

Table 7

Results of multivariate linear regression analysis using CNR as the objective variable and phantom type and phantom thickness as the explanatory variables.

Regression resultsAnalysis of variance
VariableLevelRegression coefficientLower limit of 95% confidence intervalUpper limit of 95% confidence intervalWald test p valueContribution rateChi-squared test p value

Constant term-24.62923.22826.0291.038.E-17--
Phantom typeB-1.3-2.597-0.0034.946.E-023.63%9.247.E-03
Phantom typeC-2.157-3.454-0.8612.684.E-03--
Phantom thickness--0.223-0.249-0.1962.108.E-1291.43%2.108.E-12

AIC 71.01658, CNR: contrast to noise ratio, and AIC: Akaike's Information Criterion.