Research Article

A Review of Doses for Dental Imaging in 2010–2020 and Development of a Web Dose Calculator

Table 2

Dosimetry method, machine model, phantom age, technical parameters, and effective and organ (bone marrow, brain, salivary gland, and thyroid) doses for intraoral imaging procedures. No simulation-based doses are reported.

MethodMachine modelAgeDosimeterPhantomImaging protocolsDAP (mGy × cm2)kVpFOV width (cm)FOV height (cm)Effective dose (uSv)Organ dose (uGy)Reference
Bone marrowBrainSalivary glandThyroid

MeasurementGendex Oralix DC35TLDARTBitewing single image603.54.50.850.500.0027.002.00Granlund et al. 2015 [10]
MeasurementGendex Oralix DC35TLDARTFull-mouth single image603.54.50.830.330.0025.112.94Granlund et al. 2015 [10]
MeasurementProstyle10Gafchromic filmATOM 706CPeriapical lateral7.42664.55.50.601.200.305.500.00Kadesjo et al. 2018 [11]
MeasurementProstyle10Gafchromic filmATOM 706CPeriapical central7.42664.55.50.700.000.000.010.00Kadesjo et al. 2018 [11]
MeasurementFocus35TLDCDP-R1Maxillary premolar left704.53.52.560.556.1245.2524.47Li et al. 2020 [12]
MeasurementFocus35TLDCDP-R1Maxillary premolar left704.53.52.391.197.7633.8818.38Li et al. 2020 [12]

DAP, dose area product; FOV, field of view.