Research Article

A Review of Doses for Dental Imaging in 2010–2020 and Development of a Web Dose Calculator

Table 7

Minimum, mean, and maximum values of effective and organ (bone marrow, brain, salivary gland, and thyroid) doses from cone-beam computed tomography procedures by field-of-view area (FOV width × height, cm2) reported in the selected publications.

FOV area (cm2)Effective dose (μSv)Organ dose (μGy)
Bone marrowBrainSalivary glandThyroid

Small (<50) (n = 20)Mean96.5262.8144.21984.7312.0
Min17.115.214.4336.338.6
Max220.21052.0760.04352.01252.0

Medium (50–150) (n = 32)Mean113.7246.4184.92137.2780.7
Min28.06.92.832.08.4
Max298.01478.01155.04700.03533.0

Large (>150) (n = 24)Mean151.4259.31126.82887.31267.7
Min31.010.0124.0738.347.0
Max392.21034.04653.07775.08727.0