Abstract

During the operation of tapping mode atomic force microscope (TM-AFM), the gap between the cantilever and sample surface is very small (several nanometers to micrometers). Owing to the small gap distance and high vibration frequency, squeeze film force should be considered in TM-AFM. To explore the mechanism of squeeze film damping in TM-AFM, three theoretical microcantilever simplified models are discussed innovatively herein: tip probe, ball probe, and tipless probe. Experiments and simulations are performed to validate the theoretical models. It is of great significance to improve the image quality of atomic force microscope.

1. Introduction

Tapping mode atomic force microscope (TM-AFM) is widely used to observe samples at true atomic resolution [1, 2]. AFM phase images provide more significant information than topographic images, such as information regarding adhesion, elasticity, viscoelasticity, stiffness, and chemical composition [3, 4]. Phase images and the quality factor are directly related to energy dissipation between the tip and sample [5, 6]. Many energy dissipation mechanisms, such as adhesion hysteresis, capillary interactions, air damping, and plastic deformation, have been discussed in previous studies [713]. Increasing the cantilever quality factor improves imaging sensitivity and helps to provide high-resolution phase images [14, 15]. In addition, the accurate identification of each dissipation mechanism is helpful for improving the imaging accuracy.

At present, the research on squeeze film damping is mainly focused on microelectromechanical systems (MEMS) [16], and there are few studies on squeeze film damping in TM-AFM. In related experimental studies, Garcia et al. [17, 18] used tune resonance curves (frequency sweep from low to high) to investigate the change of resonance amplitude with the decrease of tip-sample distance. In their study, the amplitude was considered as constant when the tip-sample distance was large enough. However, in fact, the amplitude exhibited a slight change before the truncated appeared, which was not presented in their papers. This experiment agrees with the study of Hoummady [19], stating that these phenomena can be explained by squeeze film damping. The experiment was carried out by Yang et al. [20] with a common probe (tip probe) in the TM-AFM, which has the advantage of more accurate tip-sample distance. In related theoretical studies, many scholars solved the Reynolds equation to obtain the influence of squeeze film damping on the dynamics of the cantilever beam, but they did not consider the effect of different tip shapes on the squeeze film damping [2124]. Lévêque et al. [25] proposed that the squeeze film damping force is related to the geometry of the tip cantilever system and a related expression can be obtained. In addition, it was found that their conclusions were in quantitative agreement with the experimental data, which clearly showed the effect of viscosity in all distances.

In this study, we demonstrate the squeeze film damping mechanism of different probes in TM-AFM and analyze the effects of squeeze film damping on amplitude and the quality factor . The results of the theoretical models are compared with those obtained via the experiment using three types of AFM probes (tip probe, ball probe, and tipless probe).

2. Experiment Procedure

2.1. Cantilever Types

To verify the squeeze film damping model, we designed an experiment to test amplitude change when the tip-sample distance was in the micron range. Tip probe (AppNano AN-NSC10-2 thickness: 4 μm, length: 125 μm, width: 30 μm, resonance frequency: 200-400 kHz, force constant: 40 N/m, coating: Al, material: Si, tip radius: 10 nm, tip height: 15 μm), ball probe (attach SiO2 spheres to the tipless cantilever, cantilever type: Nanosensors TL-NCH-10, particle shape: ball particle, diameter: 10 μm, particle material SiO2), and tipless probe (Nanosensors TL-NCH-10, thickness: 4 μm, length: 125 μm, width: 30 μm, resonance frequency: 204-479 kHz, force constant: 40 N/m, coating: none, material: Si) were used as shown in Figure 1 to analyze the effects of squeeze film damping. A Bruker Dimension Icon AFM was used in these experiments.

2.2. Experiment Procedure

In the TM-AFM autotune function, the tip offset was used to engage the cantilever on the sample stepwise. The engaged distance per step was controlled within a few micrometers or nanometers. Experimental resonance curves for different tip offset are shown in Figure 2(a). The dependence of the cantilever oscillation amplitude on the tip offset is shown in Figure 2(b).

In Figure 2(b), the tuning curve can be divided into three stages. When the tip-sample distance was relatively high, this area can be defined as the free stage, the amplitude of the tip kept constant. As the tip engaged on the sample, the probe gradually entered the squeeze film damping stage; in this stage, the tip offset led the amplitude to decline slowly. The last stage is the contact stage; in this stage, the tip touched the sample on one side. In the contact stage, the tip offset led to the rapid change in amplitude. This stage can be simply described using a contact model, as the tip-sample contact caused the truncation in Figure 2(b).

3. Experimental Results and Theoretical Calculation

3.1. Van der Waals Forces

It is well known that van der Waals forces should be notices when the distances between objects are small. To determine the effects of the van der Waals forces on the cantilever, force curves can be used to verify the action distance of these forces. The “jump into contact” phenomenon occurs at a tip-sample distance of several nanometers. Van der Waals forces between the tip and sample can be ignored when the tip-sample distance is of the order of microns. The squeeze film force is much larger than the van der Waals forces for micron-scale tip-sample distances [21]. Furthermore, it is necessarily pointed that the samples should be placed for a long time before the experiment to eliminate the effect of electrostatic forces.

3.2. Ball Probe

For the ball probe, a micron size sphere was affixed to the end of the cantilever. Because the ball was much larger than the amplitude observed in the experiment, the tip-sample distance could be considered as a fixed value. The squeeze film force was considered as concentrating on the sphere [25]. For the convenience of the theoretical calculation, the squeeze film damping between the ball probe and the sample was simplified to a one-dimensional oscillator model.

The tip-sample action force can be written as: where is the viscosity of the air, is the radius of the sphere (~10 um), is the velocity of the sphere, and is the sphere sample distance [25].

The distribution of the squeeze film force mainly depends on the instantaneous velocity of the sphere and the distance z. The amplitude of the vibration system is near 50 nm, and the frequency is 200 kHz. In the ball probe experiment, the truncation did not appear, and the small ball did not touch the sample. The squeeze film force reduced the amplitude until the system could no longer present a stable image.

The one-dimensional damping system can be expressed as follows:

In Eq. (2), , , , and are the equivalent mass, equivalent damping coefficient, equivalent stiffness, and equivalent excitation of the cantilever, respectively. Especially, for a cantilever beam with a concentrated mass ball at the free end, , , where is the mass of the microball, indicates the mass of the cantilever, represents the elastic modulus, indicates the cross-sectional moment of inertia, and is the cantilever length. Linear damping consists of squeeze film damping and constant damping (air damping and internal damping). The squeeze film damping is , and the constant damping is ; then, [22]. The system quality factor can be obtained from .

Eq. (3) can be used to calculate the theoretical quality factor of the ball probe with microspheres and is a function of the tip-sample distance . The tuning experiment results are shown in Figure 3(a), the largest amplitude of each resonance curve was also recorded, the theoretical calculations were completed using Eq. (3), and the experimental results were obtained by experimental sweep curves. For different tip-sample distances, the experimental quality factors and the theoretical curves calculated by Eq. (3) were normalized, shown in Figure 3(b). In the ball probe tuning experiment, the tip-sample distance was approximately in the micron range and van der Waals forces could be ignored in this condition. It was clearly seen that the normalized experimental variation trend was consistent with the theoretical calculation results. The one-dimensional oscillator model is reasonable for the ball probe tuning experiment.

3.3. Tipless Probes

For a small cantilever sample distance, the squeeze film damping between the cantilever and sample can be expressed by the Reynolds equation [16, 24]: where is the pressure in the film, is the axis parallel to the beam length, and is the axis parallel to the beamwidth.

Combining with the actual size of the cantilever, the width-length ratio is , where is the cantilever width. According to Pandey [23], the Reynolds equation can be simplified to one dimension and the errors can be neglected. As the curvature radius along direction being less than direction, the Reynolds equation can be simplified to

The pressure is obtained by integrating by Eq. (5) with the boundary condition at . As the cantilever sample distance is much larger than the vibration amplitude , the squeeze force per unit length can be simplified to

The equation of motion for the microcantilever with damping (consisting of squeeze film damping and constant damping) can be written as [24]:

where indicates the volume density of the cantilever, indicates the cantilever thickness, represents the constant damping coefficient per unit length. For a inclined cantilever with an angle , the cantilever-sample distance can be expressed as , where is the distance between the free end of the cantilever beam and the sample at equilibrium. By using modal decomposition, i.e., , Eq. (7) can be expressed as following when only first-order is considered. where , . Therefore, the system damping ratio can be expressed as , where is the first-order resonance frequency, and indicates the constant damping ratio. The system can then be obtained by .

Eq. (9) can be used to calculate the theoretical quality factor of tipless probe; the value of is usually ~0.001. The tuning experiment results are shown in Figure 4(a), the theoretical calculations were completed using Eq. (9), and the experimental results were obtained by experimental sweep curves, together with the largest amplitude in each curve. The normalized experimental and theoretical quality factors for different tip-sample distances are also shown in Figure 4(b). It is clear that the trend of the normalized experimental quality factor is consistent with the theoretical results. The simplified model is in good agreement with the tipless probes in the tuning experiment.

For the three types of probes, the theoretical calculation squeeze film damping ratio is shown in Figure 5. Clearly, the squeeze film damping has different effect region on different probes. At the same tip-sample distance, the influence of squeeze film damping on different probes is also very different.

4. Conclusions

In summary, the effect of the squeeze film force is a nonnegligible factor for the damping when the tip-sample distance is in the range of several micrometers, especially in TM-AFM. The experimental results and theoretical model data were in good agreement, and the simplified models are credible and easy to calculate. For the ball probe, the squeeze film force was mainly concentrated on the sphere which was attached to the end of the cantilever. For tipless probes, the main part of the squeeze film force was concentrated between the cantilever and the sample, distributed over the entire cantilever beam. For tip probes, long and thin tips could increase the distance between the cantilever and sample, and high distances reduced the squeeze film force when the tip-sample distance was in the nanometer range. This study is of great significance for understanding the mechanism of squeeze film damping dissipation in TM-AFM and promoting the development of AFM.