NaI(Tl) Detector Efficiency Computation Using Radioactive Parallelepiped Sources Based on Efficiency Transfer Principle

Table 8

Comparison between the measured and calculated full-energy peak efficiencies for parallelepiped sources, and , measured at position .

Nuclide Energy
keV
CalculatedMeasuredUnc.Δ%CalculatedMeasuredUnc.Δ%

Eu-152121.782.25E − 022.22E − 021.80E − 042.08E − 022.07E − 021.67E − 04
244.691.75E − 021.70E − 021.38E − 041.64E − 021.64E − 021.35E − 04
344.281.41E − 021.41E − 021.08E − 041.33E − 021.34E − 021.10E − 04
778.906.47E − 036.54E − 035.45E − 056.23E − 036.24E − 034.95E − 05
964.135.06E − 035.09E − 034.26E − 054.89E − 034.90E − 034.05E − 05
1408.013.49E − 033.56E − 032.87E − 053.39E − 033.45E − 032.64E − 05

Nuclide Energy
keV
CalculatedMeasuredUnc.Δ%CalculatedMeasuredUnc.Δ%

Eu-152121.782.28E − 022.22E − 021.80E − 042.11E − 022.07E − 021.67E − 04
244.691.74E − 021.70E − 021.38E − 041.64E − 021.64E − 021.35E − 04
344.281.40E − 021.41E − 021.08E − 041.32E − 021.34E − 021.10E − 04
778.906.34E − 036.54E − 035.45E − 056.11E − 036.24E − 034.95E − 05
964.134.92E − 035.09E − 034.26E − 054.76E − 034.90E − 034.05E − 05
1408.013.39E − 033.56E − 032.87E − 053.30E − 033.45E − 032.64E − 05

Nuclide Energy
keV
CalculatedMeasuredUnc.Δ%CalculatedMeasuredUnc.Δ%

Eu-152121.782.30E − 022.22E − 021.80E − 042.13E − 022.07E − 021.67E − 04
244.691.74E − 021.70E − 021.38E − 041.64E − 021.64E − 021.35E − 04
344.281.40E − 021.41E − 021.08E − 041.33E − 021.34E − 021.10E − 04
778.906.36E − 036.54E − 035.45E − 056.13E − 036.24E − 034.95E − 05
964.134.96E − 035.09E − 034.26E − 054.80E − 034.90E − 034.05E − 05
1408.013.39E − 033.56E − 032.87E − 053.30E − 033.45E − 032.64E − 05