Abstract

In modal testing, the measured frequency response functions (FRFs) are often affected by transducer mass effects. Especially when multiple sensors are employed, transducers mass effects are more significant. This paper deals with removing sensors mass effects from the measured FRFs in hammer impact testing. The proposed method offers some advantages over the available techniques in that extra FRFs measurements with different configurations are not required during the correction process. Firstly, correction method for single accelerometer case is discussed. Further, general correction formulation for multiple accelerometers case is derived. Then, validations of the proposed method are demonstrated using simulated and experimental data. It shows good effectiveness in experimental validation for single accelerometer case. Numerical simulation also demonstrates a good performance for multiple accelerometers case. However, further simulations with noisy-polluted data show that the accuracy of correction results will be affected to some extent by the noise. It is suggested that the measured FRFs be preprocessed using the curve-fitting procedure or noise reduction processing before applying the proposed method.

1. Introduction

Frequency response functions (FRFs) are widely utilized in structural dynamics analysis and modification. However, accuracies of the measured FRFs are usually affected due to a variety of factors, one of which is transducer mass effect [1, 2]. For large-scale test structures, mass effect of the sensor is relatively small and it is usually artificially ignored in the test. But for lightweight or small-scale structure, mass effect introduced by the sensor is notable [3, 4]. Especially when multiple sensors are employed, transducers masses loading effects are significant. The measured FRFs may contain larger errors. It is thus necessary to correct the measured FRFs before they are utilized for further analysis.

In early, a mass effects correction method for driving point FRF was introduced by Ewins [1]. Afterwards, some publications focused on removing such effects from transfer FRFs. Decker and Witfeid [5] presented a method to correct the transducer mass effects from transfer FRFs based on substructuring techniques. Ashory [6] proposed a method in which transfer FRFs can be corrected if the measurement is repeated with an accelerometer of different mass. Cakar and Sanliturk [7] presented a correction method by employing dummy mass. Ren et al. [8, 9] proposed a correction method for eliminating accelerometer and force transducer mass loading effects from measured FRFs in shaker modal testing. These researches mainly focus on removing mass effects of individual sensor. And extra FRFs measurements with different configurations are needed in the process of correcting the measured transfer FRFs, such as moving the sensor to different response locations, employing accelerometers of different masses or adding extra accelerometer (or dummy mass) on the structure. Zamani et al. [10] presented a strategy for correcting accelerometer mass effect using sensitivity analysis and the effect of accelerometer’s mass can be neglected using this strategy. Yin [11] presented a method for eliminating transducer additional mass effects in time-domain responses.

In practical modal test, it usually required to install multiple sensors for measuring some certain FRFs. The current authors presented a method for correcting multiple transducers mass effects from the measured FRFs in an earlier work [12]. This article expands on that work with differences on the following: correction method for both single accelerometer and multiple accelerometers cases are discussed; experiments validation, analysis, and discussions are supplemented except for numerical case study. Firstly, correction method for single accelerometer case is discussed. Further, general correction formulation for multiple accelerometers case is derived. Then, validations of the proposed method are demonstrated using simulated as well as experimental data. The proposed method offers some advantages over the available techniques in that extra FRFs measurements with different configurations are not required in the process of correcting the measured FRFs. “Different configurations” here means that test structure is changed due to movement of sensors to different response locations or addition of extra sensors (or dummy masses) on the structure.

2. Correction of Mass Effects of Single Accelerometer from the Measured FRFs

Hammer impact test of a cantilever beam is shown in Figure 1. The beam is excited at point by a hammer. And an accelerometer with a mass of is mounted at point to measure the response. Then dynamics of the beam are changed due to accelerometer mass effects.

To facilitate the analysis, accelerometer is separated from the original system, as shown in Figure 2. If excitation force and reaction force of the added masses at point are and , displacement of point can be expressed asDisplacement of accelerometer can also be established by using the receptance and reaction force :Constraint equations of forces and displacements at areBy substituting (2) and (3) into (1), ones yieldswhere is measured FRF (with mass effects). And and are exact FRFs (without mass effects). FRFs in (4) are receptances. In practical test, however, accelerances are utilized more frequently. This poses no difficulty since receptances and accelerances are just two different forms of presenting the same FRF. Multiplying both sides by of (4), one yieldswhere denotes accelerance. If excitation point is moved from to , accelerance of point can be easily obtained asBy rearranging (5) and (6), correction formulas of measured accelerances can be obtained as follows:

It can be seen from (7) and (8) that the measured accelerances can be corrected if both point and transfer accelerances are measured.

3. Theoretical Development of the Method for Multiple Accelerometers Case

An dof mass-spring system is presented in Figure 3(a). To measure the responses, accelerometers with masses are installed on point , respectively. Then the system is changed due to additional masses, as it can be seen in Figure 3(b).

To facilitate the analysis, accelerometers are separated from the original system, as shown in Figure 4. The system is excited at point and the excitation force is . Then displacements of point can be expressed aswhere are displacements of points 1, 2, and , respectively. is receptance (relating to excitation point and response point ) of original system. And , , and are reaction forces of the added masses at point , respectively. The constraint equations areThe governing equations for the added masses arewhere , , and are receptances of the added masses. Then the reaction forces of the added masses areSubstituting (12) into (9) yieldsDividing by both sides of (13) yields where s are receptances of original system and s are receptances of modified system (with transducer masses attached). Rearranging (14) by separating s and s yields

It shows that the measured FRFs can be obtained according to (15) if exact FRFs are known. However, it is the opposite in the problem of correcting the transducer mass effects that exact FRFs are to be obtained and FRFs are known (by measurement). Obviously, exact FRF cannot be directly obtained according to (15). This paper proposed a method by adding virtual negative masses to solve this problem, as it can be seen in Figure 5. Modification of the structure by adding extra masses is shown in Figure 5(a) which corresponds with (15). In Figure 5(b), however, the modified system (a3) is considered as “original system” (b1). And then the “modified system” can be generated by adding virtual negative masses () to point , respectively. The new obtained “modified system” (b3) is equivalent to original system (a1).

The process of modification of the system in Figure 5(a) can be described by employing (15). To describe Figure 5(b), it only requires to replace , , and with , , and () in (15). The result is shown in the following:It can be seen from (16) that exact FRF () of original system can be obtained given that transducer masses () and FRFs of modified system are known. By extending (8), the genera formula for correcting FRFs (, ) can be easily deduced inBy rearranging (17), exact FRF (, ) can be solved by

Discussion above refers to receptances . In practical test, accelerances are utilized more frequently. Multiplying both sides by of (18), one yieldswhere s are accelerances of original system and s are accelerances of modified system (with mass effects). It is not surprising that (19) will be coincident with (7) and (8) when only one accelerometer is used.

As it can be seen from (16)~(19), all the measured FRFs associated with the response points are required for FRFs corrections. In the case of response points, a symmetric matrix consisting of FRFs needs to be measured. Since FRF equals , only FRFs are required to be measured in fact.

4. Verification of the Method

4.1. Experimental Study for Single Accelerometer Case: Impact Testing of a Rectangle Plate

Free-free impact modal testing of a rectangle plate shown in Figure 6 was chosen for the verification of the proposed method. The rectangular plate is suspended by elastic rope, to simulate free-free boundary conditions. The size and weight of the plate are  mm3 and 0.61 Kg, respectively. The plate is equipped with 15 measuring points; the acceleration sensor (approximately 8 g) is installed at point 2. Hammer impact is moving sequentially from point 1 to point 15 for measuring accelerances . Obviously all the measured FRFs are affected by acceleration sensor mass loading effects. However, only a point FRF and a transfer FRF are chosen to be corrected here for the sake of brevity.

The process of removing accelerometer mass is shown in Figure 7(a). Effectiveness of the proposed method can be assessed by comparing the corrected FRF with the exact FRF (Figure 7(a3)). However, exact value is unavailable which makes it difficult to assess. To solve this difficulty, a comparative experiment is designed, as shown in Figure 7(b). An additional mass is attached point when accelerance is measured (Figure 7(b1)). And corrected accelerance can be obtained by removing additional mass from the measured value. Then, the method can be assessed by comparing corrected FRF with target FRF (Figure 7(b3)) which can be measured directly.

According to the idea of comparative experiments, Target FRFs and (without additional mass ) are firstly measured. Then and (with additional mass 37 g) are measured. Then corrected FRFs and can be calculated by employing (7) and (8). Effectiveness of the correction method can thus be assessed by comparing and with and , respectively.

As shown in Figure 8, resonance frequencies of the measured accelerance are lower than those of the target accelerance due to extra mass effects. After correction of extra mass , as expected, the frequencies of corrected accelerance as a whole are in quite good agreement with those of the target accelerance . However, it is not very satisfying that amplitudes at resonance peaks of are slightly inconsistent with those of , for example, the first resonant peak shown in Figure 9. This is because additional damping effects are unavoidably involved when extra mass is installed on the structure and the proposed method only considers removing the mass effects rather than damping effects. Similar observation can also be found in correction of transfer FRF and point FRF ; see Figures 10 and 11, respectively. It can be seen that the proposed method shows a good effectiveness in resonance frequencies correction. Comparison of frequencies of exact, “measured,” and corrected FRFs is given in Table 1. It is shown that frequencies errors of 1st and 2nd order are obviously reduced after correction, from 5.9%, 7.1% to 0.2%, and 0.6%, respectively, while frequencies errors of 3rd and 4th order are slightly reduced after correction. In particular, for 3rd order frequency, the error reduction is merely 0.1% (from 0.8% to 0.7%) after correction. This is due to the fact that the additional mass effects on different orders of modal are not the same. In this test, accelerometer installation position (point 2) coincides with modal node of 3rd order. Then additional mass has little effect on frequency of this order.

4.2. Numerical Simulation for Multiple Accelerometers Case: Impact Testing of Cantilever Beam

Simulated hammer impact test of a cantilever beam model is shown in Figure 12 and the physical parameters are given in Table 2. Six points are evenly arranged along the longitudinal direction of the beam. Acceleration sensors with masses of 0.38 Kg, 0.42 Kg, and 0.4 Kg are installed at points 2, 4, and 5, respectively. And the excitation point is chosen to be point 2. Obviously, the measured FRFs , , and are affected by the mass effects of accelerometers (, , and ) and need to be corrected.

According to (19), additional FRFs , , and are required to be “measured” for correction of FRFs , , and (note that equals ), as it can be seen in (20). The “measured” accelerances are numerically generated in this simulation. For comparison purpose, exact FRFs , , and corresponding with original structure (without attaching transducers) are also numerical calculated.

The correction results and comparisons are shown in Figures 13, 14, and 15. As shown in these figures, resonance frequencies of the “measured” FRFs are lower than those of exact FRFs due to the mass effects of transducers mass effects. And the magnitudes of resonance frequency reduction of each order are different. It demonstrates that transducers bring different degrees of mass effects on different orders [1315]. In this study, decreases of low-order resonances frequencies are much smaller than those of high-order resonances frequencies. As a result, transducers mass effects can be ignored if only the first order is concerned. However, elimination of transducers mass effects should be taken into consideration if high-order (e.g., the third order) is of interest. After correction, as expected, corrected FRFs , , and are in quite good agreement with exact FRFs , , and which demonstrates a good performance of the proposed method. Although the method has been confirmed to be quite promising in a simulated test, some aspects of the method need further consideration before it can readily be used in practical modal test. The measured FRFs in a practical test are inevitably polluted by noise to some extent. Further simulations presented here examine the performance of the method in the case of noisy data. The same system given in Figure 12 is utilized again. The difference is that , , , , , and are numerically incorporated with commonly used white Gaussian noise (zero mean and variance ) [15]. The noise level is defined aswhere is the mean absolute value of in the frequency range of interest. The noise levels in present study are set to be 0.5% and 1%, respectively. Since FRFs are complex values, the random noise should be added to real and imaginary parts of the FRFs, respectively.

Comparisons of measured, exact, and corrected FRFs are shown in Figures 813. It can be seen from Figures 16, 17, 19, and 20 (corrections of and ) that frequencies bands (15–25 Hz, 45–70 Hz) near second and third resonant peak have been effectively corrected, while others including frequencies bands near first resonant peak exhibit poor correction results. In Figures 18 and 21 (correction of ), however, the frequencies bands near all the resonant peaks show desirable correction results. This is mainly due to the fact that the frequencies bands of the “measured” FRFs with relatively low amplitudes are greatly polluted by the noise, thus affecting the correction results. Therefore, the proposed correction method may be sensitive to a certain extent to the noise in practical test. In this case, it is suggested that the measured FRFs be preprocessed using the curve-fitting procedure before the proposed correction method is applied. Another possible solution to this problem is to eliminate the noise from the measured FRFs before they are utilized for calculations.

5. Conclusion

In modal testing, the measured FRFs are often affected by transducer mass effects. Especially when multiple sensors are employed, transducers mass effects are more significant. This paper deals with removing accelerometers mass effects from the measured FRFs in hammer impacting testing. The research indicates the following:(1)When accelerometers are attached at points , respectively, at least FRFs related to all the response points are required to measure for eliminating accelerometers mass effects from the measured FRFs (subscript denotes excitation point).(2)Advantage of this method is that additional configurations of the structure for FRFs measurements (used for correction) are not required, such as moving the sensor to different response locations, adding extra sensors or dummy masses on the structure.

The proposed method shows good effectiveness in experimental validation for single accelerometer case. A simple numerical example presented also demonstrates a good performance in the numerical simulation for multiple accelerometers case. However, further simulations with noisy-polluted data show that the accuracy of correction results will be affected to some extent by the noise. It is suggested that the measured FRFs be preprocessed using the curve-fitting procedure or noise reduction processing before applying this method. Since the FRFs measurements required for corrections refer to different excitation points, moving the exciter to various locations is required. This is usually easy to achieve in hammer impact testing. However, moving the exciter is usually inconvenient (if not impossible) in shaker modal testing because the excitation point is usually fixed in such case. Therefore, the proposed method is more applicable for hammer impact test. Further work is necessary to extend this proposed method so as to make it also applicable for shaker modal test case. This does not mean “to make the shaker to be movable” but rather attempts to obtain the FRFs with different excitation points by employing the FRFs with same excitation points. Thus, all the FRFs required for correction can then be obtained by using the FRFs with same excitation points which avoids troublesome movements of the shaker. However, it will probably involve additional FRFs measurements with extra dummy masses or accelerometers attached.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This work was supported by Hubei Provincial Natural Science Foundation of China (2017CFB128) and Hubei Provincial Department of Education Scientific Research Project (Q20171405).