Abstract

As the harmful effects of low-level exposure to hazardous organic air pollutants become more evident, there is constant pressure to improve the detection limits of indoor and ambient air monitoring methods, for example, by collecting larger air volumes and by optimising the sensitivity of the analytical detector. However, at the other end of the scale, rapid industrialisation in the developing world and growing pressure to reclaim derelict industrial land for house building is driving the need for air monitoring methods that can reliably accommodate very-high-concentration samples in potentially aggressive matrices. This paper investigates the potential of a combination of two powerful gas chromatography—based analytical enhancements—sample preconcentration/thermal desorption and time-of-flight mass spectrometry—to improve quantitative and qualitative measurement of very-low-(ppt) level organic chemicals, even in the most complex air samples. It also describes new, practical monitoring options for addressing equally challenging high-concentration industrial samples.

1. INTRODUCTION

Samples of vapour-phase volatile organic chemicals (VOCs) in ambient air are collected in a number of different ways, of which sorbent tubes, canisters, and bags are the most popular [1, 2]. The representative sample of pollutants that results is then typically analysed by GC-MS.

However, in order to transfer a sample to the GC-MS without injecting large volumes of gas, a preconcentration device is needed. Ideally, the concentration device should allow selective retention of target VOCs while potential interferences such as water and CO2 are purged to vent. The state-of-the-art pre-concentration devices (focusing traps) built into modern thermal desorbers (TDs) are electrically cooled and configured such that retained analytes are desorbed/injected into the GC-MS with carrier gas flowing in the reverse direction to that used during the trapping/concentration process. Such “backflush” desorption means that the focusing trap can be packed with multiple sorbents in series, allowing a wider range of analytes to be sampled simultaneously (Figures 1 and 2).

The choice of air sampling method depends on the volatility and polarity range of the VOCs of interest [1, 2] and also varies from country to country, primarily because of historical investment. The USA, China and Japan, for example, have traditionally preferred canisters for time-weighted monitoring or grab sampling of the lighter VOCs [35], whereas methods across Europe and the rest of Asia have more typically favoured active or passive sampling using sorbent tubes [68].

Online/real-time air monitoring is also used extensively for kinetic studies and for compliance with specific regulations that require real-time monitoring data—for example, C2 to C10 hydrocarbon ozone precursors [9], odorous sulfur compounds [10, 11], and some trace perfluorinated compounds which are now known to be potent and long-lived greenhouse gases [12].

Each air sampling/monitoring technique has its inherent advantages and disadvantages, and these have been explored in detail in previous work [1, 2].

2. NEW ADVANCES FOR AIR SAMPLING USING CONTAINERS SUCH AS CANISTERS

2.1. High-Level Samples

Small canisters (around 400 mL volume) have been growing in popularity for grab-sampling high-concentration volatiles (with vapour pressures greater than those of n-nonane). Pre-concentration/trapping is still required prior to analysis in order to selectively eliminate the bulk constituents of air, especially oxygen, which would otherwise adversely affect the performance of the GC column and detector. However, sample volumes must be minimised to prevent overloading and/or contaminating the analytical system. Care is also required for small volumes of air to be quantitatively transferred without introducing uncertainty, and this is usually achieved using gas sample loops, capable of accurately metering sample volumes in the order of a few hundred microlitres.

A recent development has enabled gas loop introduction to be combined with the latest TD trapping technology, postdesorption sample splitting and enhanced line purging, such that high-concentration samples can be automatically processed without compromising measurement accuracy and with sufficiently low carryover to enable much lower-level samples to be reliably measured on the same analytical system (Figure 3). Incorporating these techniques into a single system setup allows flexibility in the analysis of unknown concentration samples with the added capability of running trace and high-concentration samples in the same analytical sequence without compromising results. Prior to this development, samples would have to be prescreened on a separate system and once assessed dilution of high-concentration samples using a diluent gas would have to be carried out [3, 13], this was a time consuming and labour intensive setup, and by diluting the samples errors and more importantly the reporting limits will be elevated [13].

Linearity tests were also carried out to evaluate this novel approach for analysing concentrated canister air against more conventional TD analysis procedures for lower-level (ambient) air samples (Figure 4).

The excellent linearity obtained across this range of air sample volumes, despite using two different approaches to metering the volume introduced to the focusing trap, shows good correlation of the data from samples introduced via gas sample loop and those metered using mass flow control. It is also noteworthy that these two methods were employed on the same platform. Combined with the low carryover illustrated in Figure 3, this indicates that it should be possible to extend the concentration range of automatic canister air analysis, without fear of system contamination. One benefit of this is that it should allow prescreening of unknown samples in order to optimise parameters for subsequent quantitative analysis.

Further tests were also carried out to evaluate whether or not this new capability could be used to quantify trace- and high-level contaminants in a single, heated canister air sample (Figure 5). While canisters are not the ideal sampling method for higher-boiling contaminants, they are occasionally applied for middle-distillate fuels by heating the canister during subsequent analysis (Note that the use of canisters for higher-boiling contaminants is not covered by standard air monitoring methods and is not normally recommended. Given the static (unpurged) nature of canisters and the challenge of heating an entire canister assembly uniformly, recoveries of higher-boiling analytes invariably remain compromised.). In this case, vapours from a light diesel fraction were analysed first with a low-volume sample (2 mL, black trace) to quantify high-concentration components and then a larger sample volume (100 mL, blue trace) to analyse trace-level components.

2.2. Optimising Analysis of Low-Level Canister Air Samples

For trace-level target compounds, a large canister (6 L) is typically used to collect the sample, with a large volume (1 L) being introduced to the analytical instrument in order to achieve good limits of detection. Furthermore, for trace-level monitoring, it is important that the entire preconcentrated sample is subsequently injected into the GC(MS) for maximum sensitivity. In this case, and as has been reported previously, a narrow-bore focusing trap, offering high linear gas velocity through the trap during the desorption/transfer to the GC column (Figure 1), helps ensure good peak shape for early-eluting compounds [1, 2], even under splitless conditions.

Historically, the detection of very low-level compounds in air (≪1 ppb) was possible using sensitive detectors specific for the compounds of interest—for example, flame photometric detection (FPD) provides significantly improved sensitivity for sulfur-containing compounds [14]. Alternatively, the latest TD trapping technology in combination with quadrupole MS in selected ion monitoring (SIM) mode can provide very low detection levels [12] (Figure 6). However, in each of these cases, compound identification relies on a limited number of characteristic ions and stable retention times. In this mode, to improve sensitivity, the vast majority of the spectral data are lost, so full characterisation of the sample is rarely possible in a single analysis.

Time-of-flight (TOF) MS detectors for GC overcome this limitation by monitoring all ions simultaneously across the mass range. This makes them significantly more sensitive than scanning mass spectrometers such as quadrupoles for collection of full spectral information. The sensitivity of TOF technology typically allows SIM-type detection limits to be achieved for very-low-level samples without sacrificing spectral information. Conversely, the sensitivity of TOF can also be harnessed to offer more flexibility for “normal” ambient air monitoring applications. By allowing the collection of smaller sample volumes, method detection limits are not compromised relative to quadrupole mass spectrometers used in full-scan mode. There are several advantages to analysing smaller canister air volumes. Smaller canisters are easier/cheaper to transport, and using smaller volumes allows more repeat analyses to be carried out on a single sample, if required. The use of smaller sample volumes also means that less water is introduced into the system during analysis of humid samples.

Figure 7 shows that Freon 113 (present in the atmosphere at ca. 80 ppt) can be confidently identified in as little as 10 mL of semi-rural air using TD-GC-TOF MS. NB Freon 113 provides a useful atmospheric “internal standard,” as its lifetime in the atmosphere is about 90 years and it has a uniform distribution over the globe.

Depending on analyte breakthrough volumes during focusing, another approach to improving detection limits may be to use a larger sample size. However, this would require larger canisters for sample collection, with consequent implications for the cost of shipping and storage, and the ease of deploying them in the field.

Recently, there has been a shift in the USA away from canisters, driven by growing interest in compounds that cannot be quantitatively or reliably recovered from canisters under real-world conditions (e.g., naphthalene). Sorbent tubes are now increasingly deployed, as they allow a much wider range of compounds to be analysed [15]. Sorbent tubes also provide greater sample stability as polar compounds are prone to adsorption onto the canister walls [16]. This can be avoided if the correct type of canister is used that is, fully passivated and not electropolished [17]. Studies have also shown that humidification of the sample provides longer holding times for the polar species [17].

3. NEW ADVANCES FOR AIR MONITORING USING SORBENT TUBES

3.1. Extending the VOC Sampling Range

Sorbent tubes offer numerous advantages for air monitoring, as they are easy to transport, compatible with a wider range of compounds, and are inherently compatible with a much wider range of air sample volumes. They are also compatible with analytes over a much wider polarity and volatility range. As with focusing traps (e.g., Figure 1), the sorbent combination can be varied according to the compounds of interest, with a series of multiple sorbents being used to retain/release the widest range of compounds.

The choice of sorbent(s) principally depends upon the volatility (specifically the vapour pressure) of the analyte concerned. Appropriate sorbent selection and combination have been discussed elsewhere in greater detail [1, 2]. However, for information, typical sorbents and their applicable analyte volatility range are shown in Table 6.

By choosing an appropriate sorbent combination, a whole suite of compounds can be analysed using only one sorbent tube. One such example is the analysis of 62 air toxic compounds ranging in volatility from freons to hexachloro-1,3-butadiene (Figure 8) collected onto a dual-bed sorbent tube.

3.2. Grab-Sampling for Sorbent Tubes

Sorbent tubes are typically used for time-weighted average air monitoring using diffusive (passive) or active sampling methods [1, 2]. Active sampling traditionally requires some form of calibrated pump—either constant-pressure- or constant-flow-type pump. However, simpler active sampling options are now available for “grab sampling” of air/gas volumes up to a few hundred millilitres. Key applications for grab sampling on to sorbent tubes include some industrial emissions (flue gas) measurements, landfill gases [18], and monitoring exhaled breath (e.g., for biological exposure assessment [19]).

Appropriate devices include piston-type pumps or even large gas syringes, which can be coupled to the nonsampling end of tubes, allowing air to be pulled through the sorbent tube as the plunger is withdrawn. Piston pumps are capable of sampling either large sample volumes for trace-level compounds or small sample volumes for high-concentration compounds. The sample size taken depends on the concentration of the compound(s) of interest and the detection method. Figure 9 compares the analysis of two identical laboratory air samples, one “grab sampled” using a piston pump and one collected using a standard constant-flow pump both samples were collected on multi-bed sorbent tubes.

3.3. Benefits of Small Sample Volumes for Sorbent Tubes Using Piston Pump Sampling

Piston pumps facilitate rapid collection of small volumes of air or gas and offer advantages to many air monitoring applications. For example, low sample volumes reduce the risk of volatile components breaking through the sorbent bed during sampling. They also minimise water retention. Sensitive analytical detectors such as TOF MS can be used to maintain required detection limits.

Figure 10 shows analysis of a small volume of air collected near the exhaust of a diesel car using a multi-bed sorbent tube and a piston pump. Note that the type of sorbent tube used in this case is usually rated for C4/5 to C30, but compounds as light as propene were quantitatively retained.

Ambient air monitoring procedures using TD-GC with conventional quadrupole MS technology in scan mode have typically required splitless analysis of sample volumes in the order of 1 L to reach the detection limits required by the standard methods [3, 20]. However, unless care is taken to selectively eliminate water [21, 22], particularly when monitoring humid atmospheres using tubes containing strong sorbents, such large air sample volumes could lead to the transfer of relatively large quantities of water (>500 μg) into the chromatographic system. Potential adverse analytical effects could then include split discrimination, shifting retention times, variable detector response, or masking of high-volatility compounds. A combination of small sample volumes and high-sensitivity TOF MS detection overcomes these issues.

Figure 11 shows the analysis of 500 mL of a 4 ppb-level standard of volatile hydrocarbons. The sample was analysed using TD-GC with three different detection techniques—a quadrupole mass spectrometer in full-scan and SIM mode, and a TOF mass spectrometer. The split flow was recollected onto a clean sorbent tube during each run to facilitate repeat analysis of the same sample by each of the detection methods. Three different detection techniques were chosen to show the enhancement in sensitivity that occurs when using TOF detection. Published specifications of quadrupole analysers have signal-to-noise ratios for octafluoronaphthalene (OFN) between 160 : 1–450 : 1 for full-scan mode, and however, this depends on the mass range that was chosen when the signal-to-noise check was performed. A general signal-to-noise value for quadrupoles tends to be 400 : 1 for a mass range 50 to 400 m/z. Just as quadrupoles differ in sensitivity so do TOF instruments and published specifications for OFN signal to noise range from 10 : 1 to 800 : 1. The instrument used in this paper has a published specification for OFN signal to noise of 800 : 1 over 1–1000 m/z.

RMS signal-to-noise ratios (S/N) for each of the detection methods are shown for isoprene using the ion at m/z 67. The full-scan quadrupole method is near the detection limit, assuming at lower limit of 3 : 1 S/N, while both the SIM quadrupole and the TOF methods provide S/N ratios above the limit of detection. However, using the quadrupole in SIM mode means that spectral information is lost, compromising analysis of unknowns and increasing the possibility of false positives. With the TOF detector, full spectral information is retained while detection limits remain at or better than those obtained from a quadrupole in SIM mode.

In a further example, a 200 mL sample of outdoor air from a predominantly rural environment was taken using a piston pump—Figure 12 shows extracted ion m/z 117 for carbon tetrachloride (present in the atmosphere at ca. 100 ppt) for each of the detection methods, along with respective S/N ratios. Like Freon 113, carbon tetrachloride is also often used as an atmospheric “internal standard”, as its residence time in the atmosphere is ca. 50 years and it has a uniform distribution across the globe.

4. CONCLUDING REMARKS

These examples have shown how recent advances in air monitoring technology have extended the range and robustness of GC-based analytical procedures for air samples collected using either canisters or sorbent tubes. Some of the potential applications and advantages that have been explored here include automated screening of uncharacterised canister samples, reliable detection of very-low-concentration species and technologies to facilitate transition from canisters to sorbent tubes, with the employment of simple grab sampling devices.

With the advancement of air sampling techniques and technology, excellent linearity and detection methods are available to ensure that laboratory inaccuracies are minimised. The greatest area for inaccuracy when making ambient air measurements now lies within the sample collection itself. This paper has discussed the two main sampling types for VOC collection, canister sampling and absorbent tube sampling, and it is important to ensure that when taking the sample the correct collection method is employed.

These new technologies lend themselves to sampling smaller air volumes onto sorbent tubes, which can dispel user worries about breakthrough volumes and humidity effects. Taking small sample volumes and combining this with powerful detection techniques, such as time-of-flight mass spectrometry, allows detection of trace compounds to be achieved either at the same level or at much lower limits of detection when compared to less sensitive detection techniques.

However, more work is required to identify the limits of new time-of-flight mass spectrometer technologies when used in combination with TD-GC systems for real-world air monitoring.

Acknowledgment

Freon is the registered trademark for several different chlorofluorocarbons or CFCs—trademarked by Dupont. The authors do not have a direct financial relation with the commercial identity.