ABSTRACT

Introduction. Osteotomy or fracture models can be used to evaluate mechanical properties of fixation techniques of the hand skeleton in vitro. Although many studies make use of osteotomy models, fracture models simulate the clinical situation more realistically. This study investigates monocortical and bicortical plate fixation on metacarpal bones considering both aforementioned models to decide which method is best suited to test fixation techniques. Methods. Porcine metacarpal bones ( 𝑛 = 4 0 ) were randomized into 4 groups. In groups I and II bones were fractured with a modified 3-point bending test. The intact bones represented a further control group to which the other groups after fixation were compared. In groups III and IV a standard osteotomy was carried out. Bones were fixated with plates monocortically (group I, III) and bicortically (group II, IV) and tested for failure. Results. Bones fractured at a mean maximum load of 482.8 N  Β±  104.8 N with a relative standard deviation (RSD) of 21.7%, mean stiffness was 122.3  Β±  35 N/mm. In the fracture model, there was a significant difference ( 𝑃 = 0 . 0 1 ) for maximum load of monocortically and bicortically fixed bones in contrast to the osteotomy model ( 𝑃 = 0 . 9 ). Discussion. In the fracture model, because one can use the same bone for both measurements in the intact state and the bone-plate construct states, the impact of inter-individual differences is reduced. In contrast to the osteotomy model there are differences between monocortical and bicortical fixations in the fracture model. Thus simulation of the in vivo situation is better and seems to be suitable for the evaluation of mechanical properties of fixation techniques on metacarpals.

1. INTRODUCTION

A technological progression with an increase in the number and design of implants β€œlike angular stable plates” to assure fracture healing and functional recovery has succeeded in hand surgery [1–4]. Variations in fracture models and biomechanical testing methods make it difficult to compare results among different studies. Most investigators used an osteotomy model and a four- or three-point bending protocol. A homogeneous fracture type is beneficial for the osteotomy model. However, the SD of these models is very variable [5–7].

By contrast, a fracture model could simulate the clinical situation more realistically. The purpose of this study was to create a consistent metacarpal fracture model in vitro using porcine metacarpals. This study investigates monocortical and bicortical plate fixation comparing the osteotomy and fracture models to decide which method is best to test fixation techniques.

2. MATERIAL AND METHODS

Fresh second metacarpal bones from 28-week-old domestic pigs were used ( 𝑛 = 4 0 ). Specimens were dissected from soft tissue; their dimensions were measured (average anterior-posterior diameter 5.26 mm) and stored at βˆ’20Β°C to preserve their mechanical properties close to those of fresh bones [8]. Specimens were embedded in a fixation device (14 mm of the proximal bone) using Palacos (Kulzer GmbH, Wehrheim, Germany). The free length was 25 mm to the center of the metacarpal head. All specimens were allowed to equilibrate to room temperature and were kept moist with saline irrigation during preparation, surgical procedure, and biomechanical testing.

Specimens were randomly divided into 4 groups. In two groups, a midshaft fracture was generated (groups I and II) using a modified 3-point bending method (Zwick/Roell, Z005/TN2A, Ulm, Germany) (Figure 1). Load to failure was applied with a constant speed of 100 mm/min. Data were collected to ascertain the maximum load and the yield load and also to calculate bone stiffness. In the other two groups, a midshaft transverse osteotomy, 15 mm proximal to the embedded bone side, was carried out (groups III and IV).

The fractures were dorsally reduced and fixed using a titanium miniplate (thickness 1.0 mm, 4 holes and 2.3 mm screws) and the midpoint lined up with fracture or osteotomy side (Leibinger-Stryker, Freiburg, Germany). Fixation was monocortical (group I, group III) or bicortical (group II, group IV). All fixations were carried out standardised by one surgeon. Two nonlocking screws were placed on each side of the fracture right-angled to the plate.

Biomechanical testing was performed in the same way the fracture was caused. In the osteotomy groups, a lateral fixation device was used to reduce the rotation of the bones. Load-deflection curves were continuously recorded and the ultimate failure [N] was measured. Stiffness [6] was determined from the linear region of the load-displacement curves. Data was collected using testXpert V10.11 (Zwick/Roell).

Mean value and SD were calculated for each test group and for intact bones using Microsoft Excel (Microsoft, Bellevue, Wash, USA). Failure points and stiffness were compared with nonparametric Kruskal-Wallis test using SPSS 15.0. Statistical significance was set at a value of 𝑃 < 0 . 0 5 .

3. RESULTS

Intact bones showed a highly reproducible pattern of shaft fracture similar to AO A2 with a relative standard deviation of 21.7% (Figure 2, Table 1).

Between maximum load of intact bone and after bicortical ( 𝑅 2 = 0 . 8 1 7 ) and monocortical ( 𝑅 2 = 0 . 6 6 5 ) fixation, there was a linear correlation. The monocortical-fixed bones reached mean 53.7% (SD 10.7%) of the maximum load of native bones ( 𝑃 = 0 . 0 4 9 ). With regard to bicortical fixation the construct reached mean 73.1% (SD 14%) ( 𝑃 = 0 . 0 1 5 ). Regarding stiffness and yield load bicortical-fixed bones reached higher values compared to the native bones (stiffness mean 49.9% SD 25.1%, 𝑃 = 0 . 8 9 3 ; yield load mean 82.8% SD 20.6%, 𝑃 = 0 . 0 6 3 ) than the monocortical-fixed bones (stiffness mean 40.9%, SD 14.5%, 𝑃 = 0 . 9 1 0 , yield load mean 58.8%, SD 18.9%, 𝑃 = 0 . 1 1 4 ) (Figure 3, Table 1).

The comparison of group III (osteotomy, monocortical) and group IV (osteotomy, bicortical) shows no significant differences in terms of maximum load (338.4 N Β± 79.2 N versus 343.9 N Β± 88.5 N, 𝑃 = 0 . 9 1 0 ), yield load (219.3 N Β± 96 N versus 236.4 Β± 69.4 N, 𝑃 = 0 . 5 7 9 ), and stiffness (49.5 N/mm Β± 23.5 N/mm versus 58.9 N/mm Β± 16.3 N/mm, 𝑃 = 0 . 1 6 5 ) (Table 1).

In the fracture model, however, there was a significant difference in maximum load ( 𝑃 = 0 . 0 1 ) between monocortical (mean 250 N SD 56.1 N) and bicortical (mean 359.4 N SD 90.4 N) fixation (Table 1). Yield load (171.5 N Β± 54.7 versus 253.8 N Β± 77.1 N, 𝑃 = 0 . 0 1 9 ) and stiffness (46.5 N/mm Β± 12.6 N/mm versus 56.1 N/mm Β± 21.1 N/mm, 𝑃 = 0 . 3 9 3 ) of these groups demonstrated no significant differences.

Failure was located at the osteotomy with displacement of the apex. In the fracture model there was a bending of the plate with displacement at the fracture and a failure at the screw-bone interface.

4. DISCUSSION

The objective of operative treatment of metacarpal fractures should be anatomical reconstruction and stable fixation for early functional therapy. There are a variety of fracture types including complex fractures with articular involvement to which it is difficult to achieve adequate fragment fixation.

Dorsal plating provides superior biomechanical fixation compared to other techniques. [9–12] The predominant force acting is apex dorsal loading and bending in volar direction due to the action of flexor tendons. A dorsal plate position with at least two screws on each side of the fracture provides the greatest rigidity to an apex applied load [9, 10]. Using a standard osteotomy procedure we could not demonstrate any differences between mono- and bicortical fixations. These results are similar to results from Dona et al. 2004, testing dorsal mono- and bicortical plating in a cadaveric metacarpal osteotomy [6].

The disadvantage using human specimens is that they are not homogenous samples. By contrast, animal specimens have a relatively uniform mechanical strength with similar shape, size, and bone mineral density. Like other investigators we used porcine second metacarpal bones [13, 14]. The diaphyseal region of the nonweight bearing second metacarpal porcine bone is comparable to human metacarpal and proximal phalanx bones [11, 13]. Metacarpal bones of 28-week-old domestic pigs were used which had just reached skeletal maturity. Therefore, the fracture model correlated to fractures in young adults and did not simulate osteoporotic fractures in elderly patients.

Having created a fracture protocol we could demonstrate a significant higher stability of bicortical plate fixation as compared with monocortical plate fixation. This reconfirms our experience from clinical observations. The fracture model shows a relative standard deviation (RSD) for the maximum load of 21.7% for fracturing the native bones. Relative standard deviation (RSD) for maximum load after bicortical and monocortical fixation in the fracture model was 22.4% and 25.1%, for all bones in the fracture model 24.41%. Comparing to the osteotomy model RSD for maximum load ( 𝐹 m a x ) in the fracture model was even lower (24.41% versus 35%). The RSD of osteotomy models in the literature is very variable with 30.75% in load to failure of porcine metacarpal bones [6]. The RSD of human metacarpal osteotomy model testing bioabsorbable plates was 27.6% [5] and on a sawbone osteotomy model 5.6% [7].

Another benefit of the fracture protocol is the possibility to measure a maximum load of the bones to cause a fracture as well as to test the maximum load of the plate-bone construct on the same bone. This procedure considers the interindividual variations of the bones and it allows the determination of a β€œrelative maximum load” for each specimen (= normalized values compared to native properties). Furthermore it could simulate the clinical situation in a better way than an osteotomy.

Influences of the soft tissue especially tendons were not considered in this model. The standard procedure of fracture stabilization with plates is a bicortical position of the screws including drilling through both dorsal and volar cortices, which places the flexor tendons and neurovascular bundles at risk. Moreover, the correct screw size is critical and the screw tip can interfere with the flexor tendons and cause tendon rupture [15, 16]. Statistically there was no significant difference between mono- and bicortical fixation of cadaveric metacarpal bones using dorsally conventional plating [6]. This is similar to our findings but it does not reflect our clinical observations. Using the fracture model, there was a significant difference between mono- and bicortical plate fixation comparable to clinical investigations. Open reduction in the fracture model shows an interlocking of the cortical bone at the fracture site. This prevents rotational dislocation which could be a possible mechanism of failure in the osteotomy setting.

In several biomechanical studies, metallic and bioabsorbable fixation devices of metacarpal and phalangeal osteotomies have been investigated [13, 14, 17]. Metallic plate fixation with an additional interfragmentary screw is biomechanically superior to other fixation methods [9]. In our study we only tested plate fixation without interfragmentary screws.

The type of failure in our study was deformation or cortex breaking at the fracture site with cutting out of the screws at the cortex. Bioabsorbable plates failed through plate or bone cortex breaking, showing the same modulus of elasticity of the plate and the cortex [14].

The present paper has a number of limitations. We tested static unidirectional load only. This testing method provides valuable measurements but does not reflect the clinical situation of fixation devices in hand surgery. A cyclical testing protocol is currently under review. Because of the predominantly dorsal apex loading we tested this load direction only. The influence of torsion or sudden loading cannot be estimated. In vivo loading will be more complex and certain clinical settings may include a torsional component.

Using our model, fractures are reproducible with a low variation. Because one can use the same bone for both measurements, in the intact state and the bone-plate construct state, the impact of interindividual differences is reduced. In contrast to the osteotomy model it shows differences between monocortical and bicortical fixations comparable to clinical investigations. It simulates the in vivo situation realistically and seems to be suitable for the evaluation of mechanical properties of fixation techniques on metacarpals.

AUTHOR CONTRIBUTION

First and second author contributed in equal parts.

ACKNOWLEDGMENTS

The authors would like to thank Steffen Schanz, M.S. from the Biomechanical Laboratory, Department of Trauma, Hand and Reconstructive Surgery, University of Muenster for the excellent support regarding the concept of the study and the acquisition of data. Furthermore, they would like to thank Wolfgang Walch, Transmedical, Ireland for the comprehensive review of the paper. Financial support for fixation devices and plates has been received from Leibinger/Stryker, Freiburg.