Results of the intraday variability study on the proposed HPLC method based on three different concentrations of APAP-Cys added to distilled water and to blank rat liver matrix protein dispersiona.

APAP-Cys injected (μg)Peak area ratio
(mean ± SD, 𝑛 = 3 )
RSD (%)

Samples in distilled water

0.1560.005 ± 0.0020.87
0.6250.020 ± 0.0010.60
1.2500.042 ± 0.0010.61

Samples in rat liver matrix protein dispersion

0.1560.004 ± 0.0011.15
0.6250.017 ± 0.0011.32
1.2500.039 ± 0.0031.05

aEach sample was analyzed once a day for three consecutive days.