Accuracy Verification of Magnetic Resonance Imaging (MRI) Technology for Lower-Limb Prosthetic Research: Utilising Animal Soft Tissue Specimen and Common Socket Casting Materials

Table 1

The minimum, maximum, and mean difference and absolute difference, Pearson Correlation Coefficient, the results of paired samples t-test significance, and 95% confidence interval of the difference.

Measurement differenceMaterialnAbsolute differenceDifferenceCorrelation coefficient ( 𝑃 ) 𝑃 (95% CI of the difference)
Min (%)Max (%)Mean (%)SDMin (%)Max (%)Mean (%)SD

Calliper—photo diameter measurement (mm)NA240.04 (0.03)1.67 (2.29)0.77 (0.85)0.49−1.31 (−1.29)1.67 (1.65)−0.18 (−0.17)0.090.999 (0.000)0.326
(−0.56, 0.19)

MRI—photo diameter measurement (mm)POP120.11 (0.08)1.25 (1.53)0.66 (0.73)0.361.25 (−1.53)0.61 (0.81)−0.31 (−0.33)0.711.000 (0.000)0.16
(−0.14, 0.76)
Silicone120.24 (0.19)1.20 (1.59)0.73 (0.79)0.26−1.20 (−1.59)0.74 (1.17)−0.42 (−0.41)0.671.000 (0.000)0.053
(−0.01, 0.84)

MRI—photo Surface area measurement (mm2)POP66.33 (0.05)51.03 (0.78)24.31 (0.35)18.54−15.06 (−0.10)51.03 (0.78)19.29 (0.31)24.621.000 (0.000)0.113
(−6.55, 45.13)
Silicone615.83 (0.12)68.87 (1.04)37.72 (0.51)21.25−15.83 (−0.12)68.87 (1.04)32.44 (0.47)29.931.000 (0.000)0.045
(1.03, 63.85)

MRI—water suspension volume measurement (mm3)POP38855 (1.31)9443 (4.31)9110 (2.72)301−8855 (−1.31)9443 (4.31)3207 (1.85)104481.000 (0.000)0.648
(−17689, 24103)
Silicone3831 (0.13)10243 (3.00)5521 (1.79)4706−10243 (−3.00)−831 (−0.13)−5521 (−1.79)47061.000 (0.009)0.179
(−14933, 3890)

*Correlation is significant at alpha = 0.01.