Research Article

The Use of Silver Solid Amalgam Electrodes for Voltammetric and Amperometric Determination of Nitrated Polyaromatic Compounds Used as Markers of Incomplete Combustion

Table 4

Repeatability ( 𝑛 = 7 , relative standard deviation, RSD) of peak height 𝐼 𝑝 for DC voltammetric and DP voltammetric determination of 1-NP, 2-NF, 3-NFt, and 5-NQ at m-AgSAE. Conditions for each analyte are listed in Table 3.

AnalyteDCVDPV
RSD of 𝐼 𝑝 for
c = 1 1 0 4  mol L−1
RSD of 𝐼 𝑝 for
( 𝐿 𝑄 -close 𝑐 in mol L−1)
RSD of 𝐼 𝑝 for
c = 1 1 0 4  mol L−1
RSD of 𝐼 𝑝 for
( 𝐿 𝑄 -close 𝑐 in mol L−1)

1-NP1.747.89 ( 8 1 0 7 )1.529.12 ( 1 1 0 7 )
2-NF1.326.88 ( 1 1 0 6 )0.963.15 ( 1 1 0 7 )
3-NFt1.558.14 ( 4 1 0 7 )1.107.16 ( 4 1 0 7 )
5-NQa1.839.31 ( 4 1 0 7 )1.988.89 ( 2 1 0 7 )

a from [10].