Research Article

Aerodynamic Analysis of Cup Anemometers Performance: The Stationary Harmonic Response

Table 3

Calibration coefficients, and , measured for the rotors tested with Climatronics 100075 anemometer (see also Table 2 and Figure 2). The coefficient of determination, , of the curve fittings, and the slope of the transfer function based on the rotation frequency, (see expressions (1) and (2)), have also been included.

Conical cups
Rotor [m/pulse] [m/rev] [m/s]

c-20/300.020210.606180.370960.99995
c-20/400.031770.953180.233280.99999
c-20/500.041861.255740.311480.99999
c-20/600.051801.554010.400910.99999
c-25/400.029630.888860.267600.99999
c-25/500.039321.179700.207860.99999
c-25/600.049611.488290.242450.99999
c-25/800.069642.089280.394380.99999
c-25/1000.089522.685520.536850.99999
c-30/400.028610.858180.095510.99995
c-30/500.038611.158440.196060.99999
c-30/600.048501.455050.148230.99998
c-30/800.068362.050890.254031.00000
c-30/1000.086972.609000.382530.99998
c-30/1200.107383.221450.490530.99999
c-35/500.037201.116030.190810.99999
c-35/600.047191.415820.153540.99999
c-35/800.068272.048180.181320.99998
c-35/1000.087372.621110.221680.99999
c-35/1200.107423.222590.341710.99999
c-40/500.035391.061660.137140.99999
c-40/600.045861.375870.166520.99999
c-40/800.066631.998760.169320.99998
c-40/1000.086682.600260.232920.99996
c-40/1200.106383.191400.288050.99996
c-40/1400.126393.791640.341620.99995

Elliptical cups
Rotor [m/pulse] [m/rev] [m/s]

a-27/600.052211.566170.199320.99999
a-30/600.053061.591670.232240.99998
a-35/600.054121.623610.215930.99998

Porous cups
Rotor [m/pulse] [m/rev] [m/s]

h-09/600.054451.633520.314140.99999
h-19/600.065791.973760.545390.99998
h-24/600.117633.528810.315630.99998