Determination of Ivermectin in Medicated Feeds by Liquid Chromatography with Fluorescence Detection

Table 1

Validation results.

Day 1Day 2Day 3

Standard calibration curve ,
Matrix calibration curve ,
Mean concentration (mg kg−1)
(repeatability)
2.001.972.04
Standard deviation (mg kg−1)
(repeatability)
0.06
CV (%)
(repeatability)
3.0
HorRat (repeatability)0.38
Mean recovery (%)
(repeatability)
100.198.4101.8
Mean concentration (mg kg−1)
(within-laboratory reproducibility)
2.00
Standard deviation (mg kg−1)
(within-laboratory reproducibility)
0.08
CV (%)
(within-laboratory reproducibility)
4.0
Mean recovery (%)
(within-laboratory reproducibility)
100.1
LOD (mg kg−1)0.03
LOQ (mg kg−1)0.50
value (ANOVA)0.20