Research Article

Determination of Platycodin D and Platycodin D3 in Rat Plasma Using Liquid Chromatography-Tandem Mass Spectrometry

Table 2

Intrabatch and interbatch precision and accuracy of PD in rat plasma.

Nominal concentration
(ng/mL)
Measured concentration
(ng/mL)
CV
(%)
RE
(%)

Intrabatcha
 50 3.21.8
 150 3.7−1.3
 1000 4.40.7
 8000 5.2−1.1
 10000 2.73.1
Interbatchb
 50 4.00.7
 150 6.2−8.7
 1000 8.0−3.2
 8000 8.8−1.5
 10000 1.03.5

CV: coefficient of variation; RE: relative error.
Comparisons of five replicates ( ) observations at each concentration.
Comparisons of 25 replicates ( ) observations over five different analytical runs.