Research Article

Mineral Element Contents in Commercially Valuable Fish Species in Spain

Table 2

Instrumental analytical conditions for graphite furnace atomic spectrometry of investigated elements.

CuFeMnCrSnAlCdPbNi

Wavelength (nm)324.8248.3279.5357.9224.6309.3228.8217.0232.0
Slit width (nm)0.50.20.20.50.50.50.50.50.1
Lamp current (%)80100801008080509080
Sample volume (uL)202020202020202020
Matrix modifier/volume (uL)Mg(NO3)2/0.5Mg(NO3)2/0.5Pd(NO3)2/0.5NH4H2PO4/0.5Mg(NO3)2/0.5
Argon flow (mL/min)0.20.20.20.20.20.20.20.20.2
Dryinga150/40/10150/30/10150/30/10120/40/12100/30/10120/30/5
150/20/15
120/30/10120/30/10150/30/10
Ashinga400/20/10450/20/10900/20/150750/20/630800/20/150800/20/5450/20/150400/10/30450/20/10
Pre-atomizationa850/20/1501100/20/1501200/30/4501500/15/20750/20/150800/20/1501000/20/150
Atomizationa2100/3/02100/3/02100/3/02500/3/02300/3/02450/3/01500/3/01450/3/02500/3/0
Cleaninga2700/3/02700/3/02700/3/02700/3/02700/3/02700/3/02700/3/02700/3/02700/3/0

aTemperature (°C)/time (s)/ramp (°C/s).