Diagnostic Phase of Calcium Scoring Scan Applied as the Center of Acquisition Window of Coronary Computed Tomography Angiography Improves Image Quality in Minimal Acquisition Window Scan (Target CTA Mode) Using the Second Generation 320-Row CT

1Department of Radiology, Graduate School of Medicine, University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8655, Japan
2Imaging Center, The University of Tokyo Hospital, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8655, Japan
3Department of Radiology, New Tokyo Hospital, 1271 Wanagaya, Matsudo, Chiba 270-2232, Japan
4Department of Radiology, NTT Medical Center Tokyo, 5-9-22 Higashi-Gotanda, Shinagawa-ku, Tokyo 141-8625, Japan
5Toshiba Medical Systems Corporation, Tokyo Metropolitan Regional Office, 1-6 Tsukuda 2-Chome, Chuo-ku, Tokyo 104-0051, Japan

Received 31 August 2015; Revised 4 January 2016; Accepted 13 January 2016

Academic Editor: Jinyuan Zhou

Copyright © 2016 Eriko Maeda et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Objective. To compare the image quality of coronary computed tomography angiography (CCTA) acquired under two conditions: 75% fixed as the acquisition window center (Group 75%) and the diagnostic phase for calcium scoring scan as the center (CS; Group CS). Methods. 320-row cardiac CT with a minimal acquisition window (scanned using “Target CTA” mode) was performed on 81 patients. In Group 75% (n = 40), CS was obtained and reconstructed at 75% and the center of the CCTA acquisition window was set at 75%. In Group CS (n = 41), CS was obtained at 75% and the diagnostic phase showing minimal artifacts was applied as the center of the CCTA acquisition window. Image quality was evaluated using a four-point scale (4-excellent) and the mean scores were compared between groups. Results. The CCTA scan diagnostic phase occurred significantly earlier in CS (75.7 ± 3.2% vs. 73.6 ± 4.5% for Groups 75% and CS, resp., p = 0.013). The mean Group CS image quality score (3.58 ± 0.63) was also higher than that for Group 75% (3.19 ± 0.66, p < 0.0001). Conclusions. The image quality of CCTA in Target CTA mode was significantly better when the center of acquisition window is adjusted using CS.