The Canine Erythrocyte Sedimentation Rate (ESR): Evaluation of a Point-of-Care Testing Device (MINIPET DIESSE)

Table 1

Statistical analyses of two ESR methods (ESR-MP and ESR-W) compared each other for the three groups of blood samples.

Group 1Group 2Group 3

Number of samples1097633
Sensitivity0.800.820.77
Specificity0.960.951.00
PPV0.910.821.00
NPV0.890.950.69
r0.810.870.76
K0.770.770.69

Note: group 1, all blood samples collected; group 2, nonanemic blood samples; group 3, anemic blood samples; PPV, positive predictive values; NPV, negative predictive values; r, Pearson correlation coefficient; K, Cohen’s Kappa test.