Research Article

Assessment of the Impact of Turbo Factor on Image Quality and Tissue Volumetrics in Brain Magnetic Resonance Imaging Using the Three-Dimensional T1-Weighted (3D T1W) Sequence

Table 2

Parameters that were automatically adjusted as the turbo factor was varied.

Turbo factorTime (min: sec)Minimum TI delay (ms)Scan percentage (%)Relative SNR

5020 : 25219.488.11.04
10010 : 13404.688.11.04
15005 : 08589.466.11.20
20005 : 08775.188.11.04
22505 : 08867.999.10.98