Validation of Two Activity Monitors in Slow and Fast Walking Hospitalized Patients

Table 4

Agreement between standing, sitting, and lying measured by direct observations and by ActivPAL3™ and SENS motion® used as both single- and dual-sensor systems.

StandingLyingSitting

Assessment methodStanding
N = 35
Supine
N = 35
Flexed hips & knees
N = 35
On bed/chair
N = 35
In bed, backrest elevated 60°
N = 35
Direct observation (sec)22152281224947742281
ActivPAL, single sensor (sec)150721372184
SENS, single sensor (sec)139623392161
ActivPAL, dual sensor (sec)15051665211834011580
SENS, dual sensor (sec)13961808214437401478
Percentage agreement between direct observation and ActivPAL/SENS
ActivPAL (single sensor), median % (IQR)73.5 (47.8-90.6)98.4 (89.6-100)100 (95.2-100)
SENS (single sensor), median % (IQR)67.7 (43.4-85.9)99.2 (94.5-100)99.2 (94.5-100)
ActivPAL (dual sensor), median % (IQR)71.4 (47.6-95.2)78.1 (50.0-98.4)100 (93.8-100)79.5 (60.9-90.2)92.3 (46.9-100)
SENS (dual sensor), median % (IQR)67.7 (43.4-85.9)93.1 (55.9-99.2)99.2 (94.5-100)84.6 (69.2-96.0)76.4 (38.9-93.2)

ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.