Research Article

Validation of Two Activity Monitors in Slow and Fast Walking Hospitalized Patients

Table 7

Relative and absolute reliability of steps, walking, and sedentary time assessed with ActivPAL3™, SENS motion®, and direct observations in older medical patients ().

ComparisonMean (SD) of both assessments (sec)Mean (SD) difference (sec) value for differenceICC2.1 (95% CI)SEM (SEM%)MDC95 (MDC%)

Steps
ActivPAL vs. observed229.1 (45.8)-43.6 (29.6)<0.0010.582 (-0.103-0.882)29.6 (13)82.0 (36)
SENS vs. observed235.2 (41.1)-31.5 (23.0)0.0010.678 (-0.076-0.916)23.3 (10)64.6 (27)
ActivPAL vs. SENS213.4 (47.9)-12.1 (24.8)0.10.858 (0.579-0.957)18.1 (8)59.2 (24)
Walking time
ActivPAL vs. observed160.0 (26.4)-25.8 (16.3)<0.0010.582 (-0.102-0.884)17.1 (11)47.4 (30)
SENS vs. observed165.8 (25.1)-14.2 (9.0)<0.0010.816 (-0.038-0.960)10.8 (7)29.9 (18)
ActivPAL vs. SENS152.9 (25.5)-11.5 (12.7)0.0090.813 (0.243-0.950)11.0 (7)30.5 (20)
Sedentary time
ActivPAL vs. observed184.8 (9.8)-9.3 (10.7)0.0120.416 (-0.092-0.778)7.5 (4)20.8 (11)
SENS vs. observed183.4 (9.2)-12.0 (11.9)0.0050.264 (-0.141-0.671)7.9 (4)21.9 (12)
ActivPAL vs. SENS178.8 (12.3)2.8 (13.6)0.50.544 (-0.015-0.844)8.3 (5)23.0 (13)

ICC: intraclass correlation coefficient; SEM: standard error of measurement (); MDC: minimal detectable change (, ). SEM and MDC indicate measurement error (in sec) at group and individual levels, respectively. ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.