Research Article

Validation of Two Activity Monitors in Slow and Fast Walking Hospitalized Patients

Table 8

Relative and absolute reliability of steps, walking, and sedentary time assessed with ActivPAL3™, SENS motion®, and direct observations in older patients with hip fracture ().

ComparisonMean (SD) of both assessments (sec)Mean (SD) difference (sec) value for differenceICC2.1 (95% CI)SEM (SEM%)MDC95 (MDC%)

Steps
ActivPAL vs. observed124.6 (83.7)-95.8 (48.8)<0.0010.530 (-0.089-0.868)57.4 (46)159.1 (128)
SENS vs. observed139.3 (83.4)-66.3 (48.4)0.0010.660 (-0.081-0.910)48.2 (35)133.6 (96)
ActivPAL vs. SENS91.4 (82.5)-29.5 (44.4)0.0420.825 (0.431-9.949)14.4 (16)39.9 (44)
Walking time
ActivPAL vs. observed140.3 (47.5)-126.5 (78.5)<0.0010.062 (-0.097-0.375)46.0 (33)127.5 (91)
SENS vs. observed184.5 (54.8)-38.1 (36.8)0.0040.664 (-0.003-0.903)31.8 (17)88.1 (48)
ActivPAL vs. SENS121.3 (51.5)-88.4 (63.2)0.0010.223 (-0.121-0.632)45.4 (37)125.8 (103)
Sedentary time
ActivPAL vs. observed196.5 (12.1)1.1 (2.9)0.20.970 (0.903-0.991)2.1 (1)5.8 (3)
SENS vs. observed192.9 (11.3)-6.0 (6.6)0.0100.756 0.157-0.932()5.6 (3)15.5 (8)
ActivPAL vs. SENS193.5 (12.1)7.0 (6.7)0.0040.716 (0.030-0.933)6.4 (3)17.7 (9)

ICC: intraclass correlation coefficient; SEM: standard error of measurement (); MDC: minimal detectable change (, ). SEM and MDC indicate measurement error (in sec) at group and individual levels, respectively. ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.